Enlite ® MMT-7008BRecalls & reported problems
Medtronic Minimed, Inc. · SENSOR MMT-7008B ENLITE PLUS 2L 1CL PR · Class 3
All models of Enlite (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This is a Medtronic MiniMed Enlite Plus 2 continuous glucose sensor used with an insulin pump that pauses insulin when glucose drops low. The 36 reports mostly describe the sensor giving readings far off from the fingerstick blood glucose value, which triggered a threshold suspend alarm and stopped insulin delivery even when blood sugar was high; some describe sensors with no electrode or a transmitter that would not blink. Three reports involve injuries, and there are no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 36
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 00643169786738), received between Mar 22, 2018 and Sep 10, 2019 — classified as 3 injuries, 33 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (20 reports).
14 other recalls cover devices cleared under the same FDA submission (P120010, P150001, P150019, P150029); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings20
- Unexpected Therapeutic Results6
- Break4
- Incorrect Or Inadequate Test Results3
- Calibration Problem2
- Device Displays Incorrect Message2
- Activation Problem1
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Consequences Or Impact To Patient28
- Hyperglycemia4
- Blood Loss2
- Hypoglycemia2
- Foreign Body In Patient1
Recent adverse event reports
The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 10, 2019Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results
THE CUSTOMER REPORTED VIA PHONE CALL THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 393 MG/DL AND THE SENSOR GLUCOSE WAS 64 MG/DL AT THE TIME OF THE INCIDENT. INSULIN DELIVERY WAS SUSPENDED DUE TO BLOOD GLUCOSE VERSUS SENSOR GLUCOSE LEVEL. THE LOW LIMIT SETTINGS WAS 70. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8980973 ↗MalfunctionAug 19, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED VIA PHONE CALL THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR VALUES. SENSOR GLUCOSE VALUE THAT TRIGGERED SUSPEND ON LOW OR SUSPEND BEFORE LOW EVENT WAS 60 MG/DL. LOW LIMIT IN SENSOR SETTINGS WAS 70. THE BLOOD GLUCOSE VALUE ASSOCIATED WITH THE TRIGGERED SUSPEND EVENT 169 MG/DL. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8905170 ↗MalfunctionJul 26, 2019Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results
IT WAS REPORTED THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 22 MMOL/L. THE SENSOR GLUCOSE WAS 3.2 MG/DL. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8829743 ↗MalfunctionJul 19, 2019Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results
THE CUSTOMER REPORTED VIA PHONE CALL THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE VALUE WAS 112 MG/DL AND THE SENSOR GLUCOSE WAS 43 MG/DL. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR VALUES. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8807195 ↗MalfunctionJul 11, 2019Break; Break
IT WAS REPORTED THAT THE SENSOR HAD NO ELECTRODE. THE CUSTOMER'S BLOOD GLUCOSE VALUE WAS 83 MG/DL AT THE TIME OF INCIDENT. THE SENSOR WAS INSERTED INTO ABDOMEN. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8779558 ↗MalfunctionJun 27, 2019Break; Activation Problem; Break; Activation Problem
IT WAS REPORTED THAT THE SENSOR HAD NO ELECTRODE. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 78 MG/DL. THE CUSTOMER REPORTED THAT THE TRANSMITTER DID NOT BLINK WHEN CONNECTED TO SENSOR. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8742502 ↗MalfunctionJun 10, 2019Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results
IT WAS REPORTED THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 13.3 MMOL/L AND SENSOR VALUE THAT TRIGGERED THE SUSPEND EVENT WAS 5.5 MMOL/L. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR VALUES. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Blood Loss
MAUDE report 8682098 ↗MalfunctionApr 16, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED VIA PHONE CALL THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 222 MG/DL AND THE SENSOR GLUCOSE WAS 66 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8517142 ↗MalfunctionMar 22, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 285 MG/DL AND THE SENSOR GLUCOSE WAS 50 MG/DL. CUSTOMER REPORT OTHER BLOOD GLUCOSE VALUE WAS 153 MG/DL. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR VALUES. SENSOR VALUE THAT TRIGGERED THE SUSPEND EVENT WAS 50 MG/DL. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS 50 MG/DL. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8443171 ↗MalfunctionMar 7, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE VALUE WAS 14.6 MMOL/L AND THE SENSOR GLUCOSE WAS 5.1 MMOL/L. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR VALUES. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. CUSTOMER STATES HIGH BLOOD GLUCOSE TREATED WITH BOLUS. CUSTOMER DECLINED TROUBLESHOOTING FOR HIGH BLOOD GLUCOSE. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 8398433 ↗MalfunctionMar 5, 2019Break; Break
IT WAS REPORTED THAT THE LAST SENSOR HAD NO ELECTRODE WHICH WAS NOTICED BY CUSTOMER¿S MOTHER AFTER SENSOR INSERTION. CUSTOMER'S BLOOD GLUCOSE VALUE WAS UNKNOWN. THE CUSTOMER WAS NOT ASSISTED WITH TROUBLESHOOTING. THE DEVICE WILL NOT BE RETURN FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8389207 ↗MalfunctionFeb 12, 2019Fracture; Fracture
IT WAS REPORTED THAT THE SENSOR NEEDLE WAS BREAK. THE CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN AT THE TIME OF INCIDENT. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Foreign Body In Patient
MAUDE report 8330147 ↗MalfunctionFeb 5, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 16.1 MMOL/L AND THE SENSOR GLUCOSE WAS 4.7 MMOL/L. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR VALUES. SENSOR VALUE THAT TRIGGERED THE SUSPEND EVENT WAS 2.9 MMOL/L. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS 3 MMOL/L. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 8305165 ↗MalfunctionJan 11, 2019Incorrect, Inadequate or Imprecise Result or Readings; Calibration Problem; Incorrect, Inadequate or Imprecise Result or Readings; Calibration Problem
IT WAS REPORTED THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 12 MMOL/L AND THE SENSOR GLUCOSE WAS 2.2 MMOL/L. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR VALUES. SENSOR VALUE THAT TRIGGERED THE SUSPEND EVENT WAS 2.2 MMOL/L. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS 4 MMOL/L. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Blood Loss
MAUDE report 8238753 ↗MalfunctionJan 7, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED VIA PHONE CALL THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 170 MG/DL AND THE SENSOR GLUCOSE WAS 95 MG/DL. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR GLUCOSE VALUES. SENSOR VALUE THAT TRIGGERED THE SUSPEND EVENT WAS 95 MG/DL. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS 80 MG/DL. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8221231 ↗MalfunctionJan 3, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 178 MG/DL AND THE SENSOR GLUCOSE WAS 108 MG/DL. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR GLUCOSE VALUES. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8211313 ↗MalfunctionDec 20, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 11.8 MMOL/L AND THE SENSOR GLUCOSE WAS 2.3 MMOL/L. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR VALUES. SENSOR VALUE THAT TRIGGERED THE SUSPEND EVENT WAS 4.4 MMOL/L. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS 4.8 MMOL/L. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8183566 ↗MalfunctionDec 19, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 10 MMOL/L AND THE SENSOR GLUCOSE WAS 2.2 MMOL/L. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR VALUES. SENSOR VALUE THAT TRIGGERED THE SUSPEND EVENT WAS 2.2 MMOL/L. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS 3.2 MMOL/L. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8177617 ↗MalfunctionDec 10, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 13 MMOL/L AND THE SENSOR GLUCOSE WAS 18 MMOL/L. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR VALUES. SENSOR VALUE THAT TRIGGERED THE SUSPEND EVENT WAS 3.4 MMOL/L. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS 5 MMOL/L. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8149550 ↗MalfunctionNov 29, 2018Break; Break
CUSTOMER REPORTED VIA PHONE CALL THAT WHEN THEY HE INSERTED THE SENSOR THE LITTLE NEEDLE DID NOT COME. CUSTOMER BLOOD GLUCOSE LEVEL WAS 150 MG/DL. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8114752 ↗MalfunctionNov 13, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE SENSOR WAS BROKEN. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE SENSOR WILL NOT BE RETURNED FOR THE ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8064259 ↗MalfunctionNov 12, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE VALUES WAS 6.5 MMOL/L AND THE SENSOR GLUCOSE WAS 2.9 MMOL/L. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR VALUES. SENSOR VALUE THAT TRIGGERED THE SUSPEND EVENT WAS UNKNOWN. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS UNKNOWN. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8061566 ↗MalfunctionNov 7, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 113 MG/DL AND THE SENSOR GLUCOSE WAS 50 MG/DL AT THE TIME OF THE INCIDENT. IT ALSO REPORTED THAT INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR GLUCOSE VALUES. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. IT ALSO REPORTED THAT THE SENSOR GLUCOSE VALUE THAT TRIGGERED THE SUSPEND EVENT WAS 50 MG/DL AND THRESHOLD SUSPEND LIMIT IN SENSOR SETTING WAS 60 MG/DL. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8045306 ↗MalfunctionOct 25, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 142 MG/DL AND THE SENSOR GLUCOSE WAS 57 MG/DL. THE SENSOR GLUCOSE VALUE THAT TRIGGERED THE SUSPEND EVENT WAS 57 MG/DL. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR GLUCOSE VALUES. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8004881 ↗MalfunctionSep 21, 2018Calibration Problem; Calibration Problem
THE CUSTOMER REPORTED VIA PHONE CALL THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 130 MG/DL AND THE SENSOR GLUCOSE WAS 59 MG/DL. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR VALUES. SENSOR VALUE THAT TRIGGERED THE SUSPEND EVENT WAS 59 MG/DL. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS 70 MG/DL. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7896922 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120010, P150001, P150019, P150029. One submission often covers many models, so check the product and lot details in each record.
- Z-1750-2026 · Feb 13, 2026 · Open, ClassifiedParadigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
- Z-1751-2026 · Feb 13, 2026 · Open, ClassifiedMiniMed 530G Insulin Pump (MMT-551, MMT-751)
- Z-1740-2026 · Feb 13, 2026 · Open, ClassifiedMiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
- Z-1259-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 630G (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
- Z-0001-2025 · Jul 31, 2024 · Open, ClassifiedMiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755; MiniMed 700G Insulin Pump, REF: MMT-1800, MMT-1801, MMT-1805, MMT-1850, MMT-1851
- Z-0193-2023 · Sep 20, 2022 · Open, ClassifiedInsulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742
- Z-0855-2022 · Jan 27, 2022 · Open, ClassifiedMiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782; MiniMed 770G: MMT-1880, MMT-18
- Z-0517-2021 · Apr 30, 2020 · Open, ClassifiedMiniMed 630G with SmartGuard, insulin pump, Ref/Model # MMT-1754K, MMT-1714K, MMT-1755K - Product Usage: intended for continuous delivery of basal insulin (at user selectable rate
- Z-0955-2020 · Nov 21, 2019 · Open, ClassifiedMedtronic MiniMed 630G System with SmartGuard, Ref/Model # MMT-1515, MMT-1715, MMT-1714, MMT-1754, and MMT-1755 *US and Canada Version Only
- Z-0683-2019 · Oct 3, 2018 · Open, ClassifiedMedtronic MiniMed 630G 6025805-003_C, Rx Only Medtronic MiniMed 670G 6025805-005_B, Rx Only Medtronic MiniMed 640G (International)
- Z-0670-2018 · Oct 10, 2017 · TerminatedMedtronic CareLink iPro Therapy Management Software, MMT-7340 The CareLink iPro Therapy Management Software (model, MMT-7340) is intended to be used by a healthcare professional (
- Z-2602-2017 · May 8, 2017 · TerminatedMiniMed 620G Insulin Infusion Pump, Catalog No. MMT-1510, MMT-1710, MMT-1550, MMT-1750.
- Z-2603-2017 · May 8, 2017 · TerminatedMiniMed 630G Insulin Infusion Pump, Catalog No. MMT-1715, MMT-1755.
- Z-2544-2014 · Mar 13, 2014 · TerminatedMedtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K
Other Automated insulin dosing , threshold suspend devices
- t:slim X2 Insulin Pump with Basal-IQ 1000886Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall17 deaths108,656 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls98 deaths35,170 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000912Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall2 deaths30,219 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000898Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall6 deaths16,106 reports
- MiniMed® 630G MMT-1715KMMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls50 deaths9,892 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend4 recalls36 deaths7,947 reports
- MiniMed® 530G MMT-751NASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend36 deaths7,479 reports
- MiniMed®630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls17 deaths6,645 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page