Abre™ AB9U18150090Recalls & reported problems
Medtronic, Inc. · STENT AB9U18150090 ABRE V01 · Class 3 · Implant
All models of Abre (26)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169796423), received between Apr 27, 2021 and Apr 26, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Material Deformation” (1 report).
3 other recalls cover devices cleared under the same FDA submission (P200026); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Pain1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PHYSICIAN IMPLANTED AN ABRE SELF-EXPANDING STENT IN THE RIGHT LIAC VEIN, RIGHT SUPERFICIAL VEIN, RIGHT FEMORAL VEIN ((B)(6) 2023). THE LESION WAS REPORTED AS FIBROUS, WITH NO TORTUOSITY OR CALCIFICATION. THERE WERE NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. THERE WAS NO DAMAGE NOTED TO PACKAGING (I.E. SHELF CARTON, HOOP/TRAY). THERE WERE NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES IDENTIFIED. EMBOLIC PROTECTION WAS NOT USED. A 9F SHEATH WAS USED. THELESION WAS NOT PRE-DILATED. THE STENT DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. NO EXCESSIVE FORCE WAS USED. THE PROCEDURE WAS COMPLETED WITHOUT ANY ISSUES. IT WAS REPORTED POST-OP WHEN THE PATIENT WENT HOME, AFTER THE STENT HAD BEEN IMPLANTED IN THE RIGHT ILEO-FEMORAL SPACE, THEY STARTED COMPLAINING ABOUT PAIN ON THE CONTRALATERAL (LEFT) SIDE. APPROXIMATELY 6 MONTH POST PROCEDURE THE PATIENT WAS TREATED FOR HEAVINESS AND PAIN ON THE LEFT SIDE ((B)(6) 2024).
Patient problems: Pain
MAUDE report 19189543 ↗MalfunctionApr 27, 2021Material Deformation; Material Deformation
PHYSICIAN USED AN ABRE STENT WITH A 9FR NON-MEDTRONIC SHEATH DURING TREATMENT OF A 145MM SOFT TISSUE LESION IN THE PATIENT¿S PROXIMAL, MID, AND DISTAL RIGHT COMMON ILIAC VEIN/EXTERNAL ILIAC VEIN. SLIGHT VESSEL TORTUOSITY IS REPORTED. ARTERY DIAMETER REPORTED AS 16MM. LESION EXHIBITED 80% STENOSIS. EMBOLIC PROTECTION WAS NOT USED. THERE WAS NO DAMAGE NOTED TO THE PRODUCT PACKAGING. NO ISSUES WERE NOTED WHEN REMOVING THE DEVICE FROM THE PACKAGING. IFU WAS FOLLOWED AND THE DEVICE WAS PREPPED WITHOUT ISSUE. PRE-DILATION WAS PERFORMED USING A 16MM PRE-DILATION DEVICE. THE DEVICE WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT AND NO RESISTANCE WAS NOTED DURING ADVANCEMENT. STENT FORESHORTENING IS REPORTED. A SECOND STENT WAS PLACED TO COVER THE LENGTH NEEDED. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11730594 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P200026. One submission often covers many models, so check the product and lot details in each record.
- Z-1098-2026 · Dec 4, 2025 · Open, ClassifiedAbre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U12060090 AB9U14100090 AB9U14080090 AB9U12080090 AB9U14060090 AB9G1410
- Z-0464-2022 · Nov 19, 2021 · TerminatedInstructions for Use for all model numbers of the Medtronic Abre Venous Self-Expanding Stent System.
- Z-1532-2021 · Mar 16, 2021 · TerminatedMedtronic Abre Venous Self-expanding Sent System, Catalog number REF AB9U16080090 - Product Usage: intended for use in the iliofemoral veins for the treatment of symptomatic venous
Other Stent, iliac vein devices
- Abre™ AB9U14080090Medtronic, Inc. · Stent, iliac vein2 recalls9 reports
- Venovo® VENUM16100Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall55 reports
- Venovo® VENUM14080Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall36 reports
- Venovo® VENUM14120Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall33 reports
- Venovo® VENUM14140Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall31 reports
- Venovo® VENUM14100Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall30 reports
- Venovo® VENUM16120Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall29 reports
- Venovo® VENUM14160Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall27 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page