Skip to content

SelectSecure™ MRI SureScan™ 383069 (UDI 00643169857681)Recalls & reported problems

Medtronic, Inc. · LEAD 383069 SELECTSECURE US MRI · Class 3 · Implant

All models of SelectSecure MRI SureScan (9)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic SelectSecure MRI SureScan pacemaker lead connects a pacemaker to the heart. The 1,511 reports most often describe the lead needing more energy to trigger a heartbeat, failing to trigger one, moving out of position, and sensing heartbeats incorrectly. Some reports describe a lead fracturing, and patients mention dizziness and fainting. The reports list 9 deaths and 997 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,515
Deaths reported
9
Injuries reported
998
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,515 adverse event reports that list this exact device (DI 00643169857681), received between Aug 9, 2017 and Sep 29, 2026 — classified as 9 deaths, 998 injuries, 508 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (677 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 7 reportsJan 2024: 8 reportsFeb 2024: 10 reportsMar 2024: 5 reportsApr 2024: 9 reportsMay 2024: 3 reportsJun 2024: 4 reportsJul 2024: 10 reportsAug 2024: 11 reportsSep 2024: 8 reportsOct 2024: 11 reportsNov 2024: 9 reportsDec 2024: 1 reportsJan 2025: 5 reportsFeb 2025: 1 reportsMar 2025: 4 reportsApr 2025: 6 reportsMay 2025: 3 reportsJun 2025: 6 reportsJul 2025: 7 reportsAug 2025: 2 reportsSep 2025: 9 reportsOct 2025: 0 reportsNov 2025: 7 reportsDec 2025: 5 reportsJan 2026: 6 reportsFeb 2026: 3 reportsMar 2026: 5 reportsApr 2026: 5 reportsMay 2026: 6 reportsJun 2026: 4 reportsJul 2026: 1 reportsAug 2026: 4 reportsSep 2026: 6 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Capture Threshold677
  • Failure to Capture246
  • Device Dislodged or Dislocated224
  • Over-Sensing207
  • Under-Sensing168
  • Capturing Problem163
  • Adverse Event Without Identified Device or Use Problem103
  • Intermittent Capture85

Patient problems

  • No Known Impact Or Consequence To Patient695
  • No Clinical Signs, Symptoms or Conditions463
  • Unspecified Infection72
  • No Consequences Or Impact To Patient54
  • Dizziness45
  • Syncope32
  • Chest Pain25
  • Dyspnea23

Recent adverse event reports

The 25 most recent of 1,515 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Over-Sensing; Pacing Problem; Under-Sensing; High Capture Threshold

    It was reported that the Right Atrial (RA) lead exhibited Far Field R-Wave (FFRW) oversensing triggering AT/AF episode misclassification. It was also reported that the Right Ventricular (RV) lead exhibited undersensing leading to overpacing on Atrial Tachycardia/A trial Fibrillation (AT/AF) episodes, oversensing on the ventricular lead resulting in misclassification of episode type and high thresholds. The RA lead and RV lead remain in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50929805 ↗
  • MalfunctionSep 14, 2026Over-Sensing; High Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited a large number of Sensing Integrity Counter (SIC), cross chamber oversensing, and chronic high thresholds post reprogramming. A possible setscrew issue was suspected. The IPG and RV lead remain in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50752814 ↗
  • MalfunctionSep 10, 2026Signal Artifact/Noise; Over-Sensing; High Capture Threshold

    It was reported that the Right Ventricular (RV) lead oversensing , high short V-V interval Sensing Integrity Counter (SIC) count , high thresholds and noise. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50710914 ↗
  • InjurySep 9, 2026Capturing Problem

    It was reported that the His Bundle Right Ventricular (RV) lead exhibited rising thresholds. The His Bundle RV lead was removed and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50690681 ↗
  • MalfunctionSep 8, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50675941 ↗
  • InjurySep 4, 2026Device Dislodged or Dislocated

    It was reported that the patient experienced syncope. It was further reported that the Right Ventricular (RV) lead dislodged. The lead was capped and replaced. No patient complications have been reported as a result of this event.

    Patient problems: Syncope/Fainting

    MAUDE report 50606581 ↗
  • InjuryAug 25, 2026Intermittent Capture; Capturing Problem; High Capture Threshold

    It was reported that the Right Ventricular (RV) HIS bundle lead exhibited rising and high thresholds and intermittent capture. The lead was capped and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50427355 ↗
  • MalfunctionAug 21, 2026Pacing Problem; Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50367396 ↗
  • MalfunctionAug 12, 2026Pacing Problem; Decreased Sensitivity

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50222643 ↗
  • InjuryAug 11, 2026Fracture; Over-Sensing; High Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited high thresholds, increased short v-v intervals, noise and was fractured. The lead was reprogrammed and remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50207970 ↗
  • MalfunctionJul 9, 2026Pacing Problem; Under-Sensing

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED CHEST PAIN. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING RESULTING IN INAPPPROPRIATE PACING. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Chest Pain

    MAUDE report 25759052 ↗
  • MalfunctionJun 25, 2026Degraded

    IT WAS REPORTED THAT THE LOW VOLTAGE PORTION OF THE RIGHT VENTRICULAR (RV) DEFIBRILLATION LEAD WAS SUSPECTED TO HAVE FRACTURED. THE LEAD WAS EXPLANTED AND REPLACED. IT WAS FURTHER REPORTED THAT THE LEAD WAS RETURNED TO THE MANUFACTURER AFTER BEING PROPHYLACTICALLY REPLACED AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25620440 ↗
  • MalfunctionJun 15, 2026Pacing Problem; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOW AMPLITUDE R WAVES AND UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS AND UNNECESSARY PACING IN THE VENTRICLE. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25503807 ↗
  • InjuryJun 4, 2026High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) HIS BUNDLE LEAD EXHIBITED HIGH THRESHOLDS. THE RV HIS BUNDLE LEAD WAS CAPPED AND WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25391787 ↗
  • MalfunctionJun 2, 2026Over-Sensing; Pacing Problem; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) HIS LEAD EXHIBITED UNDERSENSING RESULTING IN OVER-PACING ALONG WITH SHORT V-V INTERVALS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25367552 ↗
  • MalfunctionMay 27, 2026Failure to Capture; Pacing Problem; Under-Sensing

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED DIZZINESS AND NEAR SYNCOPE. THE RIGHT VENTRICULAR (RV) HIS BUNDLE LEAD EXHIBITED LOSS OF CAPTURE AND UNDERSENSING RESULTING IN INAPPROPRIATE PACING. THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING. THE RV HIS BUNDLE LEAD AND THE RA LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Dizziness; Presyncope

    MAUDE report 25298307 ↗
  • MalfunctionMay 25, 2026Low impedance

    IT WAS REPORTED THAT THE HIS BUNDLE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A LOW IMPEDANCE WARNING. THE HIS BUNDLE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25270170 ↗
  • MalfunctionMay 21, 2026Failure to Capture; Under-Sensing; High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS, LOSS OF CAPTURE AND UNDERSENISNG. THE RIGHT ATRIAL (RA) LEAD ALSO EXHIBITED AN LEAD POSITION CHECK FAIL. BOTH LEADS REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25243354 ↗
  • MalfunctionMay 14, 2026Low impedance; Impedance Problem

    IT WAS FURTHER REPORTED THAT THE LV LEAD EXHIBITED LOW IMPEDANCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25173483 ↗
  • MalfunctionMay 14, 2026Low impedance; High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) HIS BUNDLE LEAD EXHIBITED A LOW IMPEDANCE WARNING AND HIGH THRESHOLDS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25169563 ↗
  • MalfunctionMay 14, 2026Failure to Capture; Under-Sensing

    IT WAS NOTED THAT THE RV LEAD UNDERSENSING WAS FUNCTIONAL AND DUE TO PREMATURE VENTRICULAR CONTRACTIONS (PVC).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25169526 ↗
  • MalfunctionApr 24, 2026Over-Sensing; Pacing Problem; High Capture Threshold

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED NOISE. IT WAS FURTHER REPORTED THAT THE RIGHT VENTRICULAR (RV) HIS BUNDLE LEAD EXHIBITED HIGH THRESHOLDS, SHORT V-V INTERVALS AND OVERSENSING ON THE VENTRICULAR LEAD RESULTING IN INHIBITION OF PACING AND MISCLASSIFICATION OF EPISODE TYPE. THE IPG SYSTEM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24977965 ↗
  • InjuryApr 21, 2026Failure to Capture; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD DISLODGED AND EXHIBITED NO CAPTURE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Presyncope

    MAUDE report 24936880 ↗
  • InjuryApr 14, 2026High Capture Threshold

    IT WAS REPORTED THAT THE LEFT BUNDLE BRANCH (LBB) LEAD EXHIBITED SLIGHTLY ELEVATED THRESHOLDS. THE LBB LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24870029 ↗
  • InjuryApr 8, 2026Device Dislodged or Dislocated; Positioning Problem

    IT WAS REPORTED THAT DURING IMPLANT PROCEDURE WHEN THE RIGHT VENTRICULAR (RV) LEFT BUNDLE BRANCH LEAD WAS ATTEMPTED TO BE PLACED IT DISLODGED EACH TIME. IT IS NOT KNOWN IF THE POCKET WAS CLOSED AT THIS TIME OR NOT. ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24817084 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P030036 on Aug 3, 2005. Submitted by Medtronic Cardiac Rhythm Disease Management as “Medtronic Selectsecure Lead Model 3830”. The FDA review took 699 days. The FDA has since approved 159 supplements (changes to the device or its labeling), most recently on Oct 16, 2026. FDA record ↗

The FDA regulates product code DTB (permanent pacemaker electrode) under 21 CFR 870.3680 as a Class III device. 21 CFR 870.3680 ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Common questions about the SelectSecure™ MRI SureScan™ 383069

Has the SelectSecure™ MRI SureScan™ 383069 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00643169857681), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the SelectSecure™ MRI SureScan™ 383069?

The FDA's MAUDE database lists 1,515 adverse event reports for this device, received between Aug 9, 2017 and Sep 29, 2026. In those reports, the most common specific device problems are “High Capture Threshold” (677), “Failure to Capture” (246) and “Device Dislodged or Dislocated” (224); the most common patient problems are “Unspecified Infection” (72), “Dizziness” (45) and “Syncope” (32). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SelectSecure™ MRI SureScan™ 383069?

9 reports are classified as deaths, 998 as injuries and 508 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SelectSecure™ MRI SureScan™ 383069 FDA approved?

It was approved through premarket approval (PMA) P030036 on Aug 3, 2005.

Is the SelectSecure™ MRI SureScan™ 383069 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the SelectSecure™ MRI SureScan™ 383069, and is it still sold?

The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand SelectSecure™ MRI SureScan™, model 383069). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other permanent pacemaker Electrode devices

All permanent pacemaker Electrode devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page