Solara™ CRT-P MRI SureScan™ W1TR03 (UDI 00643169891449)Recalls & reported problems
Medtronic, Inc. · CRTP W1TR03 SOLARA CRTP MRI US · Class 3 · Implant
All models of Solara CRT-P MRI SureScan (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Solara CRT-P is a heart pacemaker implanted under the skin that keeps the heart's rhythms synchronized and is designed to be scanned by MRI. Most of the 48 reports describe infections at the incision or generator pocket, with redness, pus and wound opening, along with battery running down too early and pacing or sensing problems. Reports mention 1 death and 34 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 50
- Deaths reported
- 1
- Injuries reported
- 36
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 00643169891449), received between Jul 5, 2018 and Sep 25, 2026 — classified as 1 death, 36 injuries, 13 malfunctions.
The most frequently reported specific device problem is “Premature Discharge of Battery” (10 reports).
10 other recalls cover devices cleared under the same FDA submission (P010015); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem19
- Premature Discharge of Battery10
- Appropriate Device Problem Term/Code Not Available6
- Pacing Problem4
- Electromagnetic Interference3
- Mechanical Problem3
- Insufficient Device Problem Information2
- Over-Sensing2
Patient problems
- No Clinical Signs, Symptoms or Conditions19
- Unspecified Infection14
- Bradycardia4
- Erosion3
- Erythema2
- Insufficient Information2
- Medical device site infection2
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.
InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced twiddlers syndrome. The Implantable Pulse Generator (IPG) system was explanted. No further patient complications have been reported as a result of this event.
Patient problems: Twiddlers Syndrome
MAUDE report 50895376 ↗InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient developed a pocket infection. The Cardiac Resynchronization Therapy Pacemaker (CRT-P) was explanted and replaced. No further patient complications have been reported as a result of this event.
Patient problems: Unspecified Infection
MAUDE report 50771333 ↗InjuryAug 20, 2026Device Sensing Problem
It was reported that the Right Ventricular (RV) lead exhibited T-Wave oversensing and the Cardiac Resynchronization Therapy Pacemaker (CRT-P) exhibited a detection issue. The RV lead and CRT-P remain in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50343372 ↗InjuryMay 19, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY THE PATIENT THAT SHE "FAINTED AND WAS TAKEN TO THE EMERGENCY ROOM". THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Syncope/Fainting
MAUDE report 25218421 ↗InjuryFeb 26, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS PICKING AT THE INCISION WITH THEIR FINGERNAILS AND THEY GOT AN INFECTION. THE PATIENT ALSO EXPERIENCED EROSION AND PURULENT DISCHARGE/PUS. THE PATIENT WAS ADMINISTERED ANTIBIOTICS. THE IMPLANTABLE PULSE GENERATOR IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Erosion; Medical device site infection; Purulent Discharge
MAUDE report 24454000 ↗InjuryFeb 16, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SEVERE TRICUSPID VALVE REGURGITATION DUE TO RIGHT VENTRICULAR ENDOCARDIAL LEAD IMPINGEMENT. THE TRICUSPID REGURGITATION RESULTED IN CONGESTIVE HEART FAILURE. THE PATIENT HAD VERY MINIMAL SUBCUTANEOUS FAT IN THEIR CHEST CAUSING THE IMPENDING PERFORATION OF THE DEVICE AT THE COMERS OF THE GENERATOR. THE BI-VENTRICULAR PACEMAKER AND RIGHT VENTRICULAR (RV) LEAD WERE EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Perforation
MAUDE report 24344322 ↗InjuryDec 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS NOTED THAT THE PATIENT EXPERIENCED ENDOCARDITIS.
Patient problems: Endocarditis; Unspecified Infection
MAUDE report 23768429 ↗InjuryNov 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED APPROXIMATELY FIVE MONTHS POST IMPLANT THAT THE PATIENT DEVELOPED A POCKET INFECTION WITH SYMPTOMS OF PAIN, PURULENT DISCHARGE, REDNESS, HEAT, AND WOUND DEHISCENCE. THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Erythema; Medical device site infection; Pain; Purulent Discharge; Wound Dehiscence
MAUDE report 23495344 ↗InjurySep 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. FATIGUE, FEVER AND PUS AT THE POCKET SITE WERE OBSERVED. THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Fatigue; Fever; Unspecified Infection
MAUDE report 22991360 ↗InjuryMay 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY THE PATIENT'S FAMILY MEMBER THAT THEY DEVELOPED AN INFECTION FOLLOWING A CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) CHANGEOUT PROCEDURE. THE CRT-P SYSTEM WAS EXPLANTED AND REPLACED WITH A LEADLESS PACEMAKER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 22110423 ↗MalfunctionMay 1, 2025Signal Artifact/Noise; Electromagnetic Interference; Over-Sensing; Signal Artifact/Noise; Electromagnetic Interference; Over-Sensing
IT WAS FURTHER REPORTED THAT ELECTROMAGNETIC INTERFERENCE WAS CONFIRMED, AND OBSERVED DURING PULMONARY VEIN ISOLATION (PVI), POSTERIOR WALL ISOLATION (PWI), AND ANTERIOR MITRAL LINE ABLATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21941179 ↗MalfunctionMar 20, 2025Pacing Problem; Pacing Problem
IT WAS REPORTED BY THE PATIENT THAT THEIR CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P)'S LOWER RATE LIMIT WAS SUPPOSED TO BE PROGRAMMED TO SEVENTY BEATS PER MINUTE BUT AT NIGHT, THE PATIENT'S HEART RATE WOULD DIP DOWN TO FIFTY BEATS PER MINUTE. THE PATIENT CONTINUED, "DURING THE DAY IT WILL TAKE CARE OF IT, IT WILL GO INTO THE ONE-HUNDREDS". THE PATIENT STATED THAT THEY RECOGNIZED THIS PATTERN AFTER WEARING THEIR SMART WATCH ON AND OFF. THE PATIENT STATED, "ACCORDING TO THE WATCH, IT STAYS THERE MOST OF THE NIGHT AND THEN IT BOUNCES BACK". ADDITIONALLY, THE PATIENT NOTED THAT THEIR PHYSICIAN SAID THAT THEY "THROW A LOT OF PREMATURE VENTRICULAR CONTRACTIONS (PVC)". THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 21654391 ↗InjuryMar 5, 2025Premature Discharge of Battery; Pacing Problem; Premature Discharge of Battery; Pacing Problem
ADDITIONAL INFORMATION RECEIVED NOTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21529806 ↗MalfunctionFeb 20, 2025High impedance; Mechanical Problem; High impedance; Mechanical Problem
IT WAS REPORTED THAT DURING AN IMPLANT PROCEDURE THE WHEN THE RIGHT VENTRICULAR (RV) LEAD WAS INSERTED INTO THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) HEADER AND THE SETSCREW WAS TIGHTENED THERE WAS A HIGH UNDEFINED RV LEAD IMPEDANCE MEASUREMENT OBSERVED. THE SETSCREW WAS LOOSENED AND AN ATTEMPT TO INSERT THE RV LEAD FURTHER INTO THE HEADER WAS ATTEMPTED. THE RV LEAD WAS REMOVED FROM THE CRT-P HEADER AND THE LEAD PIN WAS CLEANED AND REINSERTED INTO THE HEADER. THE CRT-P SETSCREW WAS ENGAGED AND NO CLICKING SOUND WAS HEARD. IT WAS NOTED THAT THE SETSCREW FELT LIKE IT WAS "JUST SPINNING". THE CRT-P WAS REPLACED WITH A NEW DEVICE AND THE RV LEAD REMAINS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21423465 ↗MalfunctionNov 27, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE PATIENT THAT WHEN THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) WAS PROGRAMMED AND TURNED ON PROPERLY THEY STARTED DYING. THE PATIENT NOTED THAT THEY DIDN'T THINK THEY NEEDED THE DEVICE AND THAT FOR TWENTY-ONE YEARS THE DEVICE HAS NOT WORKED FOR THEM. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 20799453 ↗InjuryNov 18, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SYNCOPE AND A LOSS OF CONSCIOUSNESS. IT WAS NOTED THAT THE PATIENT HEART RATE DROPPED TO THE THIRTIES. THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia; Loss of consciousness; Syncope/Fainting
MAUDE report 20712385 ↗InjurySep 18, 2024Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) EXHIBITED PREMATURE BATTERY DEPLETION. THE CRT-P WAS EXPLANTED AND REPLACED. IT WAS ALSO REPORTED THAT THE HIS BUNDLE LEAD EXHIBITED HIGH THRESHOLDS. THE HIS BUNDLE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20252182 ↗InjuryAug 22, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED A FUNGAL INFECTION ON THE IMPLANTED LEADS. VEGETATION WAS OBSERVED ON THE LEADS, AND THE INFECTION SOURCE WAS IDENTIFIED AS A STRAIGHT CATH. THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) SYSTEM WAS EXPLANTED AND REPLACED WITH A LEADLESS IMPLANTABLE PULSE GENERATOR (IPG). NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 20028949 ↗InjuryAug 14, 2024Capturing Problem; Capturing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED LIGHTHEADEDNESS AND SHORTNESS OF BREATH. LOSS OF CAPTURE WAS SEEN ON A HOLTER MONITOR. IT WAS NOTED THIS WAS DUE TO THE DEVICE FEATURE THAT MONITORS THE PACING AMPLITUDE THRESHOLD AND ADJUSTS LEAD OUTPUT RUNNING. THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) REMAINS IN USE. IT WAS FURTHER REPORTED THAT THE REPROGRAMMING HAD BEEN PERFORMED, AND THE DEVICE FEATURE WAS WAS PROGRAMMED OFF. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Dizziness; Dyspnea
MAUDE report 19981488 ↗InjuryAug 9, 2024Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT POST DEVICE IMPLANT, THE BATTERY VOLTAGE WAS LESS THAN EXPECTED. IT WAS DETERMINED THAT THE DEVICE HAD BEEN EXPOSED TO A MAGNET SOURCE PRIOR TO IMPLANT FOR OVER TWO THOUSAND HOURS, CREATING A FOUR YEAR DECREASE TO THE DEVICE LONGEVITY. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19949746 ↗InjuryAug 6, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH THRESHOLDS AND THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOSS OF CAPTURE DUE TO FRACTURE. THE RV AND LV LEADS WERE CAPPED AND REPLACED. IT WAS FURTHER REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) WAS EXPLANTED AND REPLACED DUE TO AN UNKNOWN PRODUCT PERFORMANCE ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19921522 ↗MalfunctionJul 2, 2024Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV), RIGHT ATRIAL (RA), AND LEFT VENTRICULAR (LV) LEADS ALL HAD ELEVATED OUTPUTS. THE CLINICIAN INQUIRED IF THE HIGH OUTPUTS WERE CAUSING UNEXPECTED DEPLETION OF THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) BATTERY. THE CRT-P AND LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19660175 ↗InjuryJun 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) SYSTEM WAS EXPLANTED AND REPLACED WITH A LEADLESS PACEMAKER DUE TO AN INFECTION. IT WAS NOTED THAT A PIN PLUG ERODED THROUGH THE DEVICE POCKET AT THE INCISION LINE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Erosion; Unspecified Infection
MAUDE report 19605842 ↗MalfunctionJun 12, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT DUE TO BRADYCARDIA. IT WAS NOTED THAT THE PATIENT SUSPECTED THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS NOT WORKING CORRECTLY. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 19521466 ↗MalfunctionApr 2, 2024Electromagnetic Interference; Electromagnetic Interference
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) EXPERIENCE ELECTROMAGNETIC INTERFERENCE THAT PRESENTED AS INTERMITTENT OVERSENSING ON STORED EPISODES. THE CRT-P REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19028702 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P010015. One submission often covers many models, so check the product and lot details in each record.
- Z-2128-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
- Z-2127-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
- Z-2511-2020 · Jun 4, 2020 · Open, ClassifiedPercepta CRT-P MRI, Model Numbers: a) W1TR01, b) W1TR04, c) W4TR01, d) W4TR04
- Z-2512-2020 · Jun 4, 2020 · Open, ClassifiedSerena CRT-P MRI, Model Numbers: a) W1TR02, b) W1TR05, c) W4TR02, d) W4TR05
- Z-2513-2020 · Jun 4, 2020 · Open, ClassifiedSolara CRT-P MRI, Model Numbers: a) W1TR03, b) W1TR06, c) W4TR03, d) W4TR06
- Z-2415-2018 · Jun 21, 2018 · TerminatedPercepta Quad CRT-P MRI SureScan, REF W4TR04 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchronization therapy (CRT-P). These devices monitor
- Z-2413-2018 · Jun 21, 2018 · TerminatedPercepta Quad CRT-P MRI SureScan, REF W4TR01 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchronization therapy (CRT-P). These devices monitor
- Z-2412-2018 · Jun 21, 2018 · TerminatedPercepta CRT-P MRI SureScan, REF W1TR01 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchronization therapy (CRT-P). These devices monitor and
- Z-2414-2018 · Jun 21, 2018 · TerminatedPercepta CRT-P MRI SureScan, REF W1TR04 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchronization therapy (CRT-P). These devices monitor and
- Z-0342-2016 · Nov 9, 2015 · TerminatedMedtronic InSync III, model 8042, 8042B and 8042U. Implantable pulse generator pacemaker.
Other Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P) devices
- VALITUDE™ X4 CRT-P U128Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)2 recalls26 deaths2,073 reports
- Quadra Allure MP™ PM3562St. Jude Medical, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)26 deaths1,180 reports
- Percepta™ Quad CRT-P MRI SureScan™ W4TR01Medtronic, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 recall28 deaths897 reports
- VISIONIST™ X4 CRT-P U228Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 recall4 deaths819 reports
- Percepta™ Quad CRT-P MRI SureScan™ W4TR01Medtronic, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 recall8 deaths703 reports
- INVIVE™ V173Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 death670 reports
- VALITUDE™ CRT-P U125Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)2 recalls2 deaths558 reports
- QUADRA ALLURE MP™ PM3262St. Jude Medical, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)27 deaths518 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page