Skip to content

VISIONIST™ X4 CRT-P U228Recalls & reported problems

Boston Scientific Corporation · Cardiac Resynchronization Therapy Pacemaker · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Boston Scientific Visionist X4 CRT-P pacemaker has 788 FDA reports: 375 malfunctions, 409 injuries and 4 deaths. Many reports describe battery and sensing problems, and many injuries involve infections that led to system replacement. There is 1 open recall, started in August 2025, concerning high battery impedance that can trigger a fallback mode during everyday use, and a software update is available. Reports rose from 143 in the previous 12 months to 212 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
819
Deaths reported
4
Injuries reported
413
Recalls naming it
1

Summary

The FDA's MAUDE database contains 819 adverse event reports that list this exact device (DI 00802526559471), received between Jan 24, 2019 and Sep 28, 2026 — classified as 4 deaths, 413 injuries, 402 malfunctions.

The most frequently reported specific device problem is “Premature Discharge of Battery” (161 reports).

1 FDA recall record name this device's identifier, 1 still open.

19 other recalls cover devices cleared under the same FDA submission (P030005); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 10 reportsDec 2023: 6 reportsJan 2024: 10 reportsFeb 2024: 8 reportsMar 2024: 11 reportsApr 2024: 15 reportsMay 2024: 10 reportsJun 2024: 9 reportsJul 2024: 9 reportsAug 2024: 15 reportsSep 2024: 10 reportsOct 2024: 16 reportsNov 2024: 13 reportsDec 2024: 9 reportsJan 2025: 9 reportsFeb 2025: 12 reportsMar 2025: 18 reportsApr 2025: 6 reportsMay 2025: 20 reportsJun 2025: 8 reportsJul 2025: 9 reportsAug 2025: 12 reportsSep 2025: 11 reportsOct 2025: 16 reportsNov 2025: 12 reportsDec 2025: 10 reportsJan 2026: 32 reportsFeb 2026: 26 reportsMar 2026: 27 reportsApr 2026: 24 reportsMay 2026: 29 reportsJun 2026: 23 reportsJul 2026: 20 reportsAug 2026: 17 reportsSep 2026: 23 reportsOct 2026: 0 reports20242025202632
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 20, 2025Class IOpen, Classified

    Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

    Z-0105-2026 · Root cause: Other · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem250
  • Premature Discharge of Battery161
  • Over-Sensing138
  • High impedance95
  • Pacing Problem86
  • Signal Artifact/Noise73
  • Failure to Capture57
  • Incorrect, Inadequate or Imprecise Result or Readings44

Patient problems

  • No Clinical Signs, Symptoms or Conditions443
  • Unspecified Infection212
  • Sepsis38
  • No Consequences Or Impact To Patient24
  • Asystole22
  • Device Overstimulation of Tissue22
  • No Code Available20
  • Pocket Erosion19

Recent adverse event reports

The 25 most recent of 819 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Premature Discharge of Battery

    It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917670 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50917021 ↗
  • MalfunctionSep 28, 2026Premature Discharge of Battery

    It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) recorded a Code 1003, indicating the voltage was too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring until radio frequency (RF) disabled, after which time device replacement was recommended. The device remains in services. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915237 ↗
  • MalfunctionSep 28, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914860 ↗
  • InjurySep 26, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50907233 ↗
  • MalfunctionSep 24, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884943 ↗
  • MalfunctionSep 23, 2026Use of Device Problem; Interrogation Problem

    It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) exhibited a Remote Monitoring Disabled (RMD) alert. Engineering analysis determined that the alert was very likely a false positive radio frequency (RF) disablement caused by magnet application during a loaded battery voltage measurement rather than an actual battery-related condition. Review of prior device data demonstrated normal loaded battery voltage measurements. A software update was recommended to restore telemetry functionality. No adverse patient effects were reported. This CRT-P remains in service.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50869328 ↗
  • MalfunctionSep 18, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50831014 ↗
  • InjurySep 18, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50828982 ↗
  • InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) was part of a system revision due to infection. There was no additional adverse patient effects reported. The CRT-P was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50777189 ↗
  • MalfunctionSep 11, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50741460 ↗
  • MalfunctionSep 10, 2026Premature Discharge of Battery

    It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50722579 ↗
  • MalfunctionSep 8, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50674129 ↗
  • InjurySep 7, 2026Signal Artifact/Noise; Over-Sensing; Pacing Problem

    Patient problems: Asystole; Syncope/Fainting

    MAUDE report 50658147 ↗
  • MalfunctionSep 7, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50651631 ↗
  • MalfunctionSep 7, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50650664 ↗
  • MalfunctionSep 6, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50639950 ↗
  • MalfunctionSep 4, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50602445 ↗
  • MalfunctionSep 4, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50602408 ↗
  • MalfunctionSep 4, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50602361 ↗
  • MalfunctionSep 4, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50602295 ↗
  • InjurySep 4, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Implant Pain

    MAUDE report 50600120 ↗
  • InjurySep 3, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50594208 ↗
  • MalfunctionAug 24, 2026Failure to Capture; Low impedance; High Capture Threshold; Interrogation Problem

    Patient problems: Device Overstimulation of Tissue; Discomfort; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50416002 ↗
  • MalfunctionAug 20, 2026Insufficient Device Problem Information

    Patient problems: Irregular Pulse

    MAUDE report 50351084 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030005. One submission often covers many models, so check the product and lot details in each record.

  • Z-1771-2026 · Mar 19, 2026 · Open, Classified
    Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL
  • Z-0102-2026 · Aug 20, 2025 · Open, Classified
    Model Number U128, VALITUDE CRT-P EL MRI
  • Z-0104-2026 · Aug 20, 2025 · Open, Classified
    Model Number U226, VISIONIST CRT-P EL
  • Z-0103-2026 · Aug 20, 2025 · Open, Classified
    Model Number U225 VISIONIST CRT-P EL
  • Z-0101-2026 · Aug 20, 2025 · Open, Classified
    Model Number U125, VALITUDE CRT-P EL
  • Z-0827-2025 · Dec 12, 2024 · Open, Classified
    VISIONIST CRT-P, EL (Model Number U225 and U226) and EL MRI (Model Number U228)
  • Z-0826-2025 · Dec 12, 2024 · Open, Classified
    VALITUDE CRT-P, EL (Model Number U125) and EL MRI (Model Number U128)
  • Z-2019-2021 · Jun 3, 2021 · Open, Classified
    INTUA CRT-P, Model Numbers: a) V272 b) V273 c) W273
  • Z-2018-2021 · Jun 3, 2021 · Open, Classified
    INVIVE CRT-P, Model Numbers: a) V172 b) V173 c) V182 d) V183 e) W172 f) W173
  • Z-2003-2021 · Jun 3, 2021 · Open, Classified
    VISIONIST CRT-P pacemaker, Models: a) U225 b) U226 c) U228
  • Z-2020-2021 · Jun 3, 2021 · Open, Classified
    INLIVEN CRT-P, Model Numbers: a) V284 b) V285 c) W274 d) W275 (added 7/15/2021)
  • Z-2002-2021 · Jun 3, 2021 · Open, Classified
    VALITUDE CRT-P pacemaker, Models: a) U125 b) U128
  • Z-3259-2018 · Sep 10, 2018 · Terminated
    VISIONIST X4 CRT-P (Cardiac Resynchronization Therapy Pacemaker)
  • Z-3258-2018 · Sep 10, 2018 · Terminated
    VALITUDE X4 CRT-P (Cardiac Resynchronization Therapy Pacemaker)
  • Z-0366-2018 · Dec 7, 2017 · Open, Classified
    VISIONIST X4 CRT-P Pacemaker

…and 4 more.

Other Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P) devices

All Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page