VALITUDE™ CRT-P U125Recalls & reported problems

Boston Scientific Corporation · Cardiac Resynchronization Therapy Pacemaker · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Valitude pacemaker from Boston Scientific helps the heart's chambers beat in sync. The FDA database holds 542 reports for it, 311 of them describing injuries and 2 deaths. Many describe early battery drain or high battery resistance, and some mention infections. Two open recalls name this device, over battery resistance that can switch some pacemakers into backup mode. Reports rose from 68 in the previous 12 months to 117 in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
558
Deaths reported
2
Injuries reported
306
Recalls naming it
2

Summary

The FDA's MAUDE database contains 558 adverse event reports that list this exact device (DI 00802526559389), received between Apr 25, 2016 and Sep 29, 2026 — classified as 2 deaths, 306 injuries, 250 malfunctions.

The most frequently reported specific device problem is “Premature Discharge of Battery” (135 reports).

2 FDA recall records name this device's identifier, 2 still open.

18 other recalls cover devices cleared under the same FDA submission (P030005); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 7 reportsDec 2023: 3 reportsJan 2024: 12 reportsFeb 2024: 7 reportsMar 2024: 2 reportsApr 2024: 6 reportsMay 2024: 4 reportsJun 2024: 6 reportsJul 2024: 6 reportsAug 2024: 1 reportsSep 2024: 7 reportsOct 2024: 6 reportsNov 2024: 5 reportsDec 2024: 0 reportsJan 2025: 7 reportsFeb 2025: 7 reportsMar 2025: 10 reportsApr 2025: 5 reportsMay 2025: 6 reportsJun 2025: 4 reportsJul 2025: 4 reportsAug 2025: 3 reportsSep 2025: 4 reportsOct 2025: 13 reportsNov 2025: 5 reportsDec 2025: 7 reportsJan 2026: 20 reportsFeb 2026: 14 reportsMar 2026: 15 reportsApr 2026: 16 reportsMay 2026: 10 reportsJun 2026: 12 reportsJul 2026: 14 reportsAug 2026: 8 reportsSep 2026: 12 reportsOct 2026: 0 reports20242025202620
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 20, 2025Class IOpen, Classified

    Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

    Z-0101-2026 · Root cause: Other · Recall details · FDA record ↗

  2. Dec 12, 2024Class IOpen, Classified

    A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.

    Z-0826-2025 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Premature Discharge of Battery135
  • High impedance114
  • Adverse Event Without Identified Device or Use Problem105
  • Over-Sensing102
  • Incorrect, Inadequate or Imprecise Result or Readings80
  • Signal Artifact/Noise65
  • Pacing Problem59
  • Failure to Capture48

Patient problems

  • No Clinical Signs, Symptoms or Conditions298
  • Unspecified Infection90
  • No Known Impact Or Consequence To Patient70
  • Sepsis22
  • No Code Available19
  • Dizziness14
  • Asystole13
  • Device Overstimulation of Tissue13

Recent adverse event reports

The 25 most recent of 558 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941407 ↗
  • MalfunctionSep 28, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918059 ↗
  • MalfunctionSep 28, 2026Use of Device Problem; Interrogation Problem

    It was reported that the communicator of this cardiac resynchronization therapy pacemaker (CRT-P) displayed a yellow collecting wave. Technical Services (TS) recommended working with the clinic to verify the radio frequency (RF) settings in the device. It was noted the patient was seen in the clinic approximately one week later and a patient-initiated interrogation (PII) was attempted, and the yellow collecting wave was still present. While in the clinic, it was determined that RF had been disabled. It was suspected the patient likely got the newest software update when they were seen in the clinic approximately four months ago, and RF disabled with the first measurement taken by the new algorithm, which is why there was no alert. It was noted the patient would be scheduled for a device change out. No adverse patient effects were reported. This CRT-P remains in service.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914865 ↗
  • MalfunctionSep 25, 2026Premature Discharge of Battery

    It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50898079 ↗
  • InjurySep 20, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50834818 ↗
  • MalfunctionSep 16, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50796760 ↗
  • MalfunctionSep 14, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50756128 ↗
  • InjurySep 8, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50678897 ↗
  • MalfunctionSep 7, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50652338 ↗
  • MalfunctionSep 7, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50650186 ↗
  • MalfunctionSep 4, 2026Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that this pacemaker reverted to Safety Mode. Technical Services (TS) was contacted and after reviewing the data, device replacement was recommended. This pacemaker remains in service. No adverse patient effects were reported. Additional information was received indicating that this device was explanted and replaced. No additional adverse patient effects were reported. This device has been returned for analysis.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50602350 ↗
  • MalfunctionSep 4, 2026Premature Discharge of Battery; Data Problem

    It was reported that this device recorded a Code 1003, indicating the voltage was too low for the projected remaining capacity. Additionally, no data was available. Technical Services (TS) was consulted and stated that data will need to be transmitted from home or through an in office check. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. The device will remain implanted, and the patient will be monitored at this time. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50602276 ↗
  • InjuryAug 30, 2026Signal Artifact/Noise; Failure to Capture

    Patient problems: Unspecified Infection

    MAUDE report 50510331 ↗
  • InjuryAug 24, 2026Extracardiac stimulation; Incorrect, Inadequate or Imprecise Result or Readings; Battery Problem

    It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) reverted to Safety Mode. When attempting a software update, the device went from ten months remaining to seven months remaining then reverted to Safety Mode. "0xa" and "Oscillator mismatch" codes were also recorded. The patient stated they felt uncomfortable and experienced stimulation while in office. Technical Services (TS) recommended device replacement. This CRT-P was explanted, replaced, and returned for analysis. No additional adverse patient effects were reported.

    Patient problems: Device Overstimulation of Tissue; Discomfort

    MAUDE report 50419152 ↗
  • MalfunctionAug 24, 2026Use of Device Problem; Interrogation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50416021 ↗
  • MalfunctionAug 19, 2026Premature Discharge of Battery; Battery Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50331330 ↗
  • InjuryAug 14, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient developed an infection of the cardiac resynchronization therapy pacemaker (CRT-P) system and underwent an explant procedure of all components on (b)(6) 2026. The procedure started with extraction of the lead that was placed for left bundle branch area pacing. Minimal laser was required and the lead was successfully removed. Next, the atrial lead was extracted and during removal a loss of capture occurred on the temporary right ventricular (RV) pacing lead due to the lead catheter being pulled on. The patient was noted to be dependent, with no underlying rhythm. The healthcare provider (HCP) moved to adjust the temporary pacing device and inadvertently caught his foot on the laser equipment and pulled on the atrial lead, dislodging it from the heart. At the same time another HCP was able to increase the output on the temporary device. After the atrial lead was extracted the patient's blood pressure began to drop when capture was lost. Medication was administered and the patient underwent a transesophageal echocardiogram (TEE) to examine the heart and no issues were observed.Once the patient's BP was stabilized, and capture with the temporary RV lead

    Patient problems: Unspecified Infection

    MAUDE report 50248282 ↗
  • MalfunctionAug 7, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50176692 ↗
  • InjuryAug 5, 2026Premature Discharge of Battery; Use of Device Problem; Interrogation Problem

    It was reported that this pacemaker disabled radio frequency (RF) telemetry due a single low battery voltage measurement. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Additional information received reported that this pacemaker was explanted and replaced. No additional adverse patient effects were reported. Product return was requested.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50137458 ↗
  • MalfunctionAug 3, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50117794 ↗
  • MalfunctionJul 29, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50071245 ↗
  • MalfunctionJul 28, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50057737 ↗
  • MalfunctionJul 28, 2026Premature Discharge of Battery

    It was reported that this device recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. Additional information was received indicating that this device was explanted and replaced. No additional adverse patient effects were reported, and this device has not been returned.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50057736 ↗
  • InjuryJul 27, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) was part of a system revision due to infection. There were no additional adverse patient effects reported. This CRT-P was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50018626 ↗
  • MalfunctionJul 27, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50008624 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030005. One submission often covers many models, so check the product and lot details in each record.

  • Z-1771-2026 · Mar 19, 2026 · Open, Classified
    Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL
  • Z-0105-2026 · Aug 20, 2025 · Open, Classified
    Model Number U228, VISIONIST CRT-P EL MRI
  • Z-0102-2026 · Aug 20, 2025 · Open, Classified
    Model Number U128, VALITUDE CRT-P EL MRI
  • Z-0104-2026 · Aug 20, 2025 · Open, Classified
    Model Number U226, VISIONIST CRT-P EL
  • Z-0103-2026 · Aug 20, 2025 · Open, Classified
    Model Number U225 VISIONIST CRT-P EL
  • Z-0827-2025 · Dec 12, 2024 · Open, Classified
    VISIONIST CRT-P, EL (Model Number U225 and U226) and EL MRI (Model Number U228)
  • Z-2003-2021 · Jun 3, 2021 · Open, Classified
    VISIONIST CRT-P pacemaker, Models: a) U225 b) U226 c) U228
  • Z-2019-2021 · Jun 3, 2021 · Open, Classified
    INTUA CRT-P, Model Numbers: a) V272 b) V273 c) W273
  • Z-2018-2021 · Jun 3, 2021 · Open, Classified
    INVIVE CRT-P, Model Numbers: a) V172 b) V173 c) V182 d) V183 e) W172 f) W173
  • Z-2002-2021 · Jun 3, 2021 · Open, Classified
    VALITUDE CRT-P pacemaker, Models: a) U125 b) U128
  • Z-2020-2021 · Jun 3, 2021 · Open, Classified
    INLIVEN CRT-P, Model Numbers: a) V284 b) V285 c) W274 d) W275 (added 7/15/2021)
  • Z-3259-2018 · Sep 10, 2018 · Terminated
    VISIONIST X4 CRT-P (Cardiac Resynchronization Therapy Pacemaker)
  • Z-3258-2018 · Sep 10, 2018 · Terminated
    VALITUDE X4 CRT-P (Cardiac Resynchronization Therapy Pacemaker)
  • Z-0366-2018 · Dec 7, 2017 · Open, Classified
    VISIONIST X4 CRT-P Pacemaker
  • Z-0361-2018 · Dec 7, 2017 · Open, Classified
    VALITUDE X4 CRT-P Pacemaker

…and 3 more.

Other Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P) devices

All Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P) devices →