Midas Rex MR8 MR8-AA15 (UDI 00643169912359)Recalls & reported problems
Medtronic Powered Surgical Solutions · ATT MR8 MR8-AA15 15CM ANGLED 2.4MM · Class 2
All models of MIDAS REX MR8 (137)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Midas Rex angled drill attachment is a surgical power tool used to drill or burr bone. The reports mostly describe the attachment overheating during use and internal bearings breaking or detaching, sometimes leaving loose parts inside the tool, with sticking or hard-to-insert behavior also mentioned. There are 28 reports, with no deaths and no injuries, and no recalls name this device. Reports rose sharply in the past year, from 4 to 20.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 30
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 00643169912359), received between Nov 21, 2022 and Sep 23, 2026 — classified as 30 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- No Apparent Adverse Event12
- Overheating of Device10
- Detachment of Device or Device Component3
- Difficult to Insert2
- Physical Resistance/Sticking2
- Naturally Worn1
- Noise, Audible1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions26
- Insufficient Information4
Recent adverse event reports
The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026No Apparent Adverse Event
It was reported that the device was defective. There was no patient impact. Additional information received upon evaluation stating that the bearings were detached.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50863666 ↗MalfunctionSep 9, 2026Use of Device Problem
Medtronic received information regarding a repair request and no alleged product deficiencies were reported. There was no patient impact. Additional information was received stating that the bearing detached was found during evaluation.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50684713 ↗MalfunctionJul 2, 2026Overheating of Device
IT WAS REPORTED THAT THE ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHMENT WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25687977 ↗MalfunctionJul 2, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE ATTACHMENT WAS DEFECTIVE. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THERE WERE DETACHED BEARINGS INSIDE THE DEVICE.
Patient problems: Insufficient Information
MAUDE report 25687720 ↗MalfunctionJul 1, 2026No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST WITH NO ALLEGED PRODUCT DEFICIENCY. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THE DISTAL BEARING IS BROKEN WAS FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25677722 ↗MalfunctionJul 1, 2026No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION RECEIVED AFTER EVALUATION STATING THAT DETACHED BEARINGS WAS FOUND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25676199 ↗MalfunctionJun 30, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THERE WERE NO ALLEGED ISSUE WITH THE DEVICE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25668057 ↗MalfunctionJun 30, 2026Difficult to Insert
IT WAS REPORTED THAT THE ATTACHMENT WILL WON'T BE FIXED WITH BUR. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25663659 ↗MalfunctionJun 30, 2026Naturally Worn
IT WAS REPORTED THAT ATTACHMENT HAS BEARING PROBLEM. THERE WAS NO PATIENT INVOLVED.ADDITIONAL INFORMATION RECEIEVD STATING THAT THE BEARINGS INSIDE THE DEVICE WERE BROKEN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25663033 ↗MalfunctionJun 30, 2026No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST, AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO THE PATIENT.ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION
Patient problems: Insufficient Information
MAUDE report 25662952 ↗MalfunctionJun 27, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE ALLEGED ISSUE WAS UNKNOWN DEFECT. AT THE TIME OF REPORT, THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION RECEIVED UPON EVALUATION STATING THAT THE BEARING WAS DETACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25646324 ↗MalfunctionJun 12, 2026Difficult to Insert
IT WAS REPORTED THAT DURING PRE-OPERATION THE ATTACHMENT HAD LOCKING ISSUE. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BROKEN BEARING WAS FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25486826 ↗MalfunctionJun 10, 2026No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST, AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION RECEIVED THAT THE BEARINGS ARE DETACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25461221 ↗MalfunctionJun 10, 2026Overheating of Device
IT WAS REPORTED THAT THE ATTACHMENT GETS HOT. THERE WAS NO PATIENT IMPACT IN THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION OF THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25459465 ↗MalfunctionMay 6, 2026No Apparent Adverse Event
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE DEVICE WITH NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT NO PATIENT WAS IMPACTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25091649 ↗MalfunctionApr 8, 2026Physical Resistance/Sticking
IT WAS REPORTED THAT THE ATTACHMENT HAD LOCK OR ADJUSTMENT RING STICKING. NO KNOWN PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24814296 ↗MalfunctionApr 8, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE HANDPIECE WAS DEFECTIVE. THERE WAS NO PATIENT IMPACT AT THE TIME OF THE REPORT. ADDITIONAL INFORMATION RECEIVED THAT THE BEARINGS WERE DETACHED DURING THE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24812319 ↗MalfunctionApr 3, 2026Overheating of Device
IT WAS REPORTED THAT DURING USE, THE ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24775640 ↗MalfunctionMar 26, 2026No Apparent Adverse Event
IT IS REPORTED THAT THE DEVICE WITH GENERAL DISSATISFACTION. IT IS REPORTED THAT THERE IS NO PATIENT IMPACT . ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24699395 ↗MalfunctionMar 24, 2026Vibration
IT WAS REPORTED THAT THE ATTACHMENT WAS OBSERVED TO SHAKE AND MAKE NOISE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24678256 ↗MalfunctionMar 20, 2026Physical Resistance/Sticking
IT WAS REPORTED THAT PRIOR TO USE ATTACHMENT DOES NOT LOCK THE DRILLS. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24647209 ↗MalfunctionFeb 18, 2026No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. AT THE TIME OF REPORT, THERE WAS NO KNOWN IMPACT TO A PATIENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: Insufficient Information
MAUDE report 24374884 ↗MalfunctionAug 14, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22802456 ↗MalfunctionMay 22, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. NO PATIENT IMPACT REPORTED.
Patient problems: Insufficient Information
MAUDE report 22069265 ↗MalfunctionMar 20, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21651596 ↗
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page