This Medtronic Midas Rex MR8 is a powered surgical drill accessory. Reports mostly describe ball bearings falling out or coming loose, parts detaching, and the attachment not taking the drill, and most list no clinical effect on the patient. There are 905 reports, no deaths and 2 injuries. No recalls name this device. Reports rose from 41 in the prior 12 months to 801 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
952
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 952 adverse event reports that list this exact device (DI 00763000441005), received between May 16, 2023 and Sep 30, 2026 — classified as 2 injuries, 950 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (288 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 4 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 12 reportsSep 2024: 37 reportsOct 2024: 3 reportsNov 2024: 6 reportsDec 2024: 3 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 8 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 3 reportsJul 2025: 5 reportsAug 2025: 5 reportsSep 2025: 4 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 6 reportsJan 2026: 0 reportsFeb 2026: 6 reportsMar 2026: 58 reportsApr 2026: 105 reportsMay 2026: 109 reportsJun 2026: 437 reportsJul 2026: 46 reportsAug 2026: 20 reportsSep 2026: 59 reportsOct 2026: 0 reports202420252026437
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component288
  • Naturally Worn238
  • No Apparent Adverse Event174
  • Difficult to Insert96
  • Overheating of Device65
  • Difficult to Remove56
  • Loss of Power39
  • Physical Resistance/Sticking21

Patient problems

  • No Clinical Signs, Symptoms or Conditions921
  • Insufficient Information30
  • Unintended Radiation Exposure2
  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 25 most recent of 952 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Overheating of Device; Vibration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951953 ↗
  • MalfunctionSep 30, 2026No Apparent Adverse Event

    It was reported that the attachment is faulty. At the time of event, there is no patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950664 ↗
  • MalfunctionSep 30, 2026No Apparent Adverse Event

    It was reported that there was no issue. No additional information was received. There were analysis findings.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946573 ↗
  • MalfunctionSep 30, 2026Difficult to Remove

    It was reported that the tool was stuck inside the attachment. There was no patient impact. Additional information was received upon evaluation stating that the distal bearings were detached.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944452 ↗
  • MalfunctionSep 30, 2026Loss of Power

    It was reported that the device was faulty. There was no impact to the patient at the time of report. Additional information received as distal bearings detached found during evaluation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943100 ↗
  • MalfunctionSep 29, 2026Naturally Worn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934413 ↗
  • MalfunctionSep 29, 2026No Apparent Adverse Event

    It was reported that the customer had general dissatisfaction over the product. There was no patient involvement or impact. Additional information received as bearings misaligned found during evaluation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50930216 ↗
  • MalfunctionSep 29, 2026Naturally Worn

    It was reported that prior to use, there was bearing damage. No patient involvement and impact was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50930065 ↗
  • MalfunctionSep 28, 2026Corroded

    It was reported that rust was observed in the device. At the time of report there was no patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911958 ↗
  • MalfunctionSep 28, 2026No Apparent Adverse Event

    It was reported that prior to use device was defective. At the time of report there was no patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911716 ↗
  • MalfunctionSep 28, 2026Difficult to Insert; Loss of Power

    It was reported that the device was seized and the craniotome does not fit tightly. There was no patient impact in the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911501 ↗
  • MalfunctionSep 28, 2026No Apparent Adverse Event

    It was reported that the device is faulty. It was reported that there was no patient impact. Additional information received stating detached bearing was found during evaluation made it reportable.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911451 ↗
  • MalfunctionSep 28, 2026No Apparent Adverse Event

    It was reported that during procedure the inner part of the attachment is loose, which compromises the normal operation of the tool. There was no patient involved with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911019 ↗
  • MalfunctionSep 28, 2026Naturally Worn

    It was reported that after use, the bearings in the attachment were damaged. There was no patient involved in this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50910939 ↗
  • MalfunctionSep 25, 2026No Apparent Adverse Event

    Medtronic received information regarding a repair request and no alleged product deficiencies were reported. It was reported that there was no patient impact. Additional information was received stating that loose bearings were found during evaluation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50895722 ↗
  • MalfunctionSep 25, 2026Naturally Worn

    It was reported that the device bearing was damaged prior to use. It was reported that there was no patient involvement. Additional information was received stating that detached bearings were found during evaluation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50895332 ↗
  • MalfunctionSep 25, 2026Naturally Worn

    It was reported that during pre-op the bearings were broken . At the time of report no patient was involved . Additional information received that bearings were detached during evaluation. .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50894553 ↗
  • MalfunctionSep 24, 2026Naturally Worn

    It was reported that the bearings were damaged. There was no patient involvement. Additional information received upon evaluation stating that the bearings were detached.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883953 ↗
  • MalfunctionSep 24, 2026No Apparent Adverse Event

    Medtronic received information regarding a repair request, and no alleged product deficiencies were reported. There was no patient impact in this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883084 ↗
  • MalfunctionSep 24, 2026No Apparent Adverse Event

    Medtronic received information regarding a repair request, and no alleged product deficiencies were reported. At the time of report there was no patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50882906 ↗
  • MalfunctionSep 24, 2026No Apparent Adverse Event

    Medtronic received information regarding a repair request and no alleged product deficiencies were reported. It was reported that there was no patient impact. Additional information was received stating that misaligned bearings were found during evaluation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50877990 ↗
  • MalfunctionSep 24, 2026Detachment of Device or Device Component

    It was reported that ball bearing fell out off the attachment. It was reported that there was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50877716 ↗
  • MalfunctionSep 23, 2026Difficult to Insert

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873399 ↗
  • MalfunctionSep 23, 2026Difficult to Remove

    It was reported that the drill bit used to create the bone flap is stuck and cannot be removed. It was reported that there was no patient impact. Additional information was received stating that detached bearings were found during evaluation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50866476 ↗
  • MalfunctionSep 23, 2026Naturally Worn

    It was reported that prior to use, the bearing was broken. It was reported that there was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50865204 ↗