Skip to content

Midas Rex MR8 MR8-AF02 (UDI 00643169912403)Recalls & reported problems

Medtronic Powered Surgical Solutions · ATT MR8 MR8-AF02 FOOTED F2 B-1 2.4MM · Class 2

All models of MIDAS REX MR8 (137)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Midas Rex MR8 is a powered surgical drill used in head and spine surgery. Many of its 188 reports describe bearings or other parts coming loose or detaching, and some describe overheating or worn parts. Most reports describe no effect on patients. There is 1 injury report and no deaths. No recalls name this device. Reports rose from 13 in the year before to 140 in the latest year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
189
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 189 adverse event reports that list this exact device (DI 00643169912403), received between May 17, 2021 and Sep 28, 2026 — classified as 1 injury, 188 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (69 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 4 reportsSep 2024: 23 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 3 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 3 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 5 reportsApr 2026: 14 reportsMay 2026: 8 reportsJun 2026: 82 reportsJul 2026: 19 reportsAug 2026: 4 reportsSep 2026: 8 reportsOct 2026: 0 reports20242025202682
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component69
  • No Apparent Adverse Event46
  • Naturally Worn27
  • Overheating of Device20
  • Difficult to Insert16
  • Difficult to Remove14
  • Physical Resistance/Sticking8
  • Loss of Power4

Patient problems

  • No Clinical Signs, Symptoms or Conditions178
  • Insufficient Information12
  • Unintended Radiation Exposure1

Recent adverse event reports

The 25 most recent of 189 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Overheating of Device

    It was reported that the attachment with the report of temperature increase. It was reported that there was no known patient impact. Additional information was received stating that detached bearings were found during evaluation.

    Patient problems: Insufficient Information

    MAUDE report 50911026 ↗
  • MalfunctionSep 25, 2026Naturally Worn

    It was reported that the bearings were broken. It was reported that there was no patient involvement. Additional information was received stating that detached bearings were found during evaluation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50895041 ↗
  • MalfunctionSep 23, 2026No Apparent Adverse Event

    Medtronic received information regarding a repair request, and no alleged product deficiencies were reported. There was no patient impact reported. Additional information was received stating that loose bearings were found during evaluation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50865712 ↗
  • MalfunctionSep 15, 2026No Apparent Adverse Event

    It was reported that the device was defective. There was no patient impact. Additional information received stating that the distal bearing got detached.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50777512 ↗
  • MalfunctionSep 15, 2026Detachment of Device or Device Component

    It was reported that ball bearing fell out / missing. No patient involvement and impact was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50769372 ↗
  • MalfunctionSep 15, 2026Detachment of Device or Device Component

    It was reported that the bearings were missing from the device. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50768344 ↗
  • MalfunctionSep 15, 2026No Apparent Adverse Event

    Medtronic received information regarding a repair request and no alleged product deficiencies were reported. It was reported that there was no patient involvement. Additional information was received stating that bearings were detached found in analysis.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50768259 ↗
  • MalfunctionSep 9, 2026Naturally Worn

    It was reported that prior to use, the bearing was broken. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50683655 ↗
  • MalfunctionAug 28, 2026Naturally Worn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50494079 ↗
  • MalfunctionAug 28, 2026Naturally Worn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50493097 ↗
  • MalfunctionAug 17, 2026Difficult to Insert

    It was reported that the cannot insert the tool in the attachment. There was no patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50272453 ↗
  • MalfunctionAug 17, 2026Difficult to Insert

    It was reported that the attachment no longer connects to the drill. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50268818 ↗
  • MalfunctionJul 28, 2026No Apparent Adverse Event

    Medtronic received information regarding a repair request and no alleged product deficiencies were reported. There was no known patient impact reported. Additional information was received stating that distal bearing detached were found during evaluation.

    Patient problems: Insufficient Information

    MAUDE report 50049028 ↗
  • MalfunctionJul 6, 2026Difficult to Remove; Noise, Audible

    IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF REMOVAL DIFFICULTY AND NOISE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25714559 ↗
  • MalfunctionJul 3, 2026Difficult to Insert

    IT WAS REPORTED THAT DIFFICULT TO INSERT ON MOTOR. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHMENT WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25702873 ↗
  • MalfunctionJul 3, 2026Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25702160 ↗
  • MalfunctionJul 3, 2026Difficult to Insert

    IT WAS REPORTED THAT IMPOSSIBLE TO LOCK ON THE MOTOR. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHMENT WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25702146 ↗
  • MalfunctionJul 3, 2026Naturally Worn

    IT WAS REPORTED THAT THE ATTACHMENT HAD BEARINGS BROKEN. THERE IS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25700981 ↗
  • MalfunctionJul 2, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION

    Patient problems: Insufficient Information

    MAUDE report 25689857 ↗
  • MalfunctionJul 2, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION

    Patient problems: Insufficient Information

    MAUDE report 25689800 ↗
  • MalfunctionJul 2, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE ATTACHMENT HAS GENERAL DISSATISFACTION. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHMENT WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25689280 ↗
  • MalfunctionJul 2, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT A ATTACHMENT WAS DEFECTIVE. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BROKEN BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25688585 ↗
  • MalfunctionJul 2, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THERE WERE NO ALLEGED ISSUE WITH THE DEVICE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25688285 ↗
  • MalfunctionJul 2, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST WITH NO ALLEGED PRODUCT DEFICIENCY. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING BEARINGS ARE DETACHED WAS FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25687565 ↗
  • MalfunctionJul 1, 2026Overheating of Device; Difficult to Remove

    IT WAS REPORTED THAT DURING SUPRATENTORIAL HEMISPHERIC TUMOR RESECTION BY OSTEOPLASTIC CRANIOTOMY PROCEDURE THE CUTTER/TOOL WAS STUCK INSIDE THE ATTACHMENT, ALSO THE ATTACHMENT APPEARED HOT FROM THE BEGINNING OF THE DRILLING AND REMAINED HOT AFTER SWITCHING TO THE CRANIOTOMY PROCEDURE. AT THAT TIME, THE ENTIRE CUT THROUGH THE BONY TABLE COULD BE COMPLETED, SO THERE WAS NO IMPACT ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25680628 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices