Midas Rex MR8 MR8-AF02 (UDI 00643169912403)Recalls & reported problems
Medtronic Powered Surgical Solutions · ATT MR8 MR8-AF02 FOOTED F2 B-1 2.4MM · Class 2
All models of MIDAS REX MR8 (137)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Midas Rex MR8 is a powered surgical drill used in head and spine surgery. Many of its 188 reports describe bearings or other parts coming loose or detaching, and some describe overheating or worn parts. Most reports describe no effect on patients. There is 1 injury report and no deaths. No recalls name this device. Reports rose from 13 in the year before to 140 in the latest year.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 189
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 189 adverse event reports that list this exact device (DI 00643169912403), received between May 17, 2021 and Sep 28, 2026 — classified as 1 injury, 188 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (69 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component69
- No Apparent Adverse Event46
- Naturally Worn27
- Overheating of Device20
- Difficult to Insert16
- Difficult to Remove14
- Physical Resistance/Sticking8
- Loss of Power4
Patient problems
- No Clinical Signs, Symptoms or Conditions178
- Insufficient Information12
- Unintended Radiation Exposure1
Recent adverse event reports
The 25 most recent of 189 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Overheating of Device
It was reported that the attachment with the report of temperature increase. It was reported that there was no known patient impact. Additional information was received stating that detached bearings were found during evaluation.
Patient problems: Insufficient Information
MAUDE report 50911026 ↗MalfunctionSep 25, 2026Naturally Worn
It was reported that the bearings were broken. It was reported that there was no patient involvement. Additional information was received stating that detached bearings were found during evaluation.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50895041 ↗MalfunctionSep 23, 2026No Apparent Adverse Event
Medtronic received information regarding a repair request, and no alleged product deficiencies were reported. There was no patient impact reported. Additional information was received stating that loose bearings were found during evaluation.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50865712 ↗MalfunctionSep 15, 2026No Apparent Adverse Event
It was reported that the device was defective. There was no patient impact. Additional information received stating that the distal bearing got detached.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50777512 ↗MalfunctionSep 15, 2026Detachment of Device or Device Component
It was reported that ball bearing fell out / missing. No patient involvement and impact was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50769372 ↗MalfunctionSep 15, 2026Detachment of Device or Device Component
It was reported that the bearings were missing from the device. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50768344 ↗MalfunctionSep 15, 2026No Apparent Adverse Event
Medtronic received information regarding a repair request and no alleged product deficiencies were reported. It was reported that there was no patient involvement. Additional information was received stating that bearings were detached found in analysis.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50768259 ↗MalfunctionSep 9, 2026Naturally Worn
It was reported that prior to use, the bearing was broken. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50683655 ↗MalfunctionAug 28, 2026Naturally Worn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50494079 ↗MalfunctionAug 28, 2026Naturally Worn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50493097 ↗MalfunctionAug 17, 2026Difficult to Insert
It was reported that the cannot insert the tool in the attachment. There was no patient impact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50272453 ↗MalfunctionAug 17, 2026Difficult to Insert
It was reported that the attachment no longer connects to the drill. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50268818 ↗MalfunctionJul 28, 2026No Apparent Adverse Event
Medtronic received information regarding a repair request and no alleged product deficiencies were reported. There was no known patient impact reported. Additional information was received stating that distal bearing detached were found during evaluation.
Patient problems: Insufficient Information
MAUDE report 50049028 ↗MalfunctionJul 6, 2026Difficult to Remove; Noise, Audible
IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF REMOVAL DIFFICULTY AND NOISE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25714559 ↗MalfunctionJul 3, 2026Difficult to Insert
IT WAS REPORTED THAT DIFFICULT TO INSERT ON MOTOR. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHMENT WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25702873 ↗MalfunctionJul 3, 2026Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25702160 ↗MalfunctionJul 3, 2026Difficult to Insert
IT WAS REPORTED THAT IMPOSSIBLE TO LOCK ON THE MOTOR. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHMENT WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25702146 ↗MalfunctionJul 3, 2026Naturally Worn
IT WAS REPORTED THAT THE ATTACHMENT HAD BEARINGS BROKEN. THERE IS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25700981 ↗MalfunctionJul 2, 2026No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION
Patient problems: Insufficient Information
MAUDE report 25689857 ↗MalfunctionJul 2, 2026No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION
Patient problems: Insufficient Information
MAUDE report 25689800 ↗MalfunctionJul 2, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE ATTACHMENT HAS GENERAL DISSATISFACTION. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHMENT WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25689280 ↗MalfunctionJul 2, 2026No Apparent Adverse Event
IT WAS REPORTED THAT A ATTACHMENT WAS DEFECTIVE. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BROKEN BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25688585 ↗MalfunctionJul 2, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THERE WERE NO ALLEGED ISSUE WITH THE DEVICE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25688285 ↗MalfunctionJul 2, 2026No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST WITH NO ALLEGED PRODUCT DEFICIENCY. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING BEARINGS ARE DETACHED WAS FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25687565 ↗MalfunctionJul 1, 2026Overheating of Device; Difficult to Remove
IT WAS REPORTED THAT DURING SUPRATENTORIAL HEMISPHERIC TUMOR RESECTION BY OSTEOPLASTIC CRANIOTOMY PROCEDURE THE CUTTER/TOOL WAS STUCK INSIDE THE ATTACHMENT, ALSO THE ATTACHMENT APPEARED HOT FROM THE BEGINNING OF THE DRILLING AND REMAINED HOT AFTER SWITCHING TO THE CRANIOTOMY PROCEDURE. AT THAT TIME, THE ENTIRE CUT THROUGH THE BONY TABLE COULD BE COMPLETED, SO THERE WAS NO IMPACT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25680628 ↗
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
- NA 5407-120-450Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)149 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page