Midas Rex MR8 MR8-AF01 (UDI 00643169912397)Recalls & reported problems
Medtronic Powered Surgical Solutions · ATT MR8 MR8-AF01 FOOTED F1 B-5 2.4MM · Class 2
All models of MIDAS REX MR8 (137)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00643169912397), received between Jul 1, 2022 and Jun 26, 2026 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent5
- No Apparent Adverse Event3
- Overheating of Device3
- Detachment of Device or Device Component2
- Naturally Worn2
- Electrical /Electronic Property Problem1
- Noise, Audible1
- Vibration1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
- Insufficient Information1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 26, 2026Naturally Worn
IT IS REPORTED THAT THE DEVICE BEARING BROKEN. IT IS REPORTED THAT THERE IS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25634363 ↗MalfunctionJun 19, 2026Overheating of Device
IT WAS REPORTED THAT PRIOR TO USE THE DEVICE HAD TEMPERATURE INCREASE ISSUE. THERE WAS NO PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT CHIPPED DISTAL BEARING WAS FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25561757 ↗MalfunctionJun 18, 2026No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST, AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25546531 ↗MalfunctionJun 10, 2026Overheating of Device
IT WAS REPORTED THAT, PRIOR TO USE, THE DEVICE WAS OBSERVED TO HAVE A TEMPERATURE INCREASE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25456469 ↗MalfunctionJun 1, 2026No Apparent Adverse Event
MEDTRONIC RECEIVED INFORMATION REGARDING A SCHEDULED PREVENTIVE MAINTENANCE, AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DISTAL BEARINGS WERE DETACHED, TIP OF THE TUBE IS OUT OF SHAPE, LASER MARKINGS AND COLOR BAND WERE FOUND FADED DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25355227 ↗MalfunctionMay 29, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE BEARING WERE FELL OUT DURING A PROCEDURE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DISTAL BEARING WAS DETACHED WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25328453 ↗MalfunctionMay 19, 2026Vibration; Detachment of Device or Device Component; Naturally Worn; Noise, Audible
IT WAS REPORTED THAT THE BEARINGS WERE BAD, LOUD AND ¿RUNNING ROUGH¿. NO ADDITIONAL INFORMATION WAS RECEIVED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 25222359 ↗MalfunctionMar 31, 2026Overheating of Device
IT WAS REPORTED THAT PRE-OPERATIVELY THERE WAS TEMPERATURE INCREASE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24744160 ↗MalfunctionAug 6, 2024Material Twisted/Bent; Material Twisted/Bent
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF BENT FOOT. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. REPAIR ESCALATED TO A PRODUCT EVENT BASED ON REASON FOR RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19921459 ↗MalfunctionAug 6, 2024Material Twisted/Bent; Material Twisted/Bent
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF BENT FOOT. NO PATIENT IMPACT REPORTED. REPAIR IS BEING ESCALATED TO PRODUCT EVENT DUE TO REASON FOR THE RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19915663 ↗MalfunctionApr 12, 2023Electrical /Electronic Property Problem; Material Twisted/Bent; Electrical /Electronic Property Problem; Material Twisted/Bent
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF FROZEN. NO PATIENT IMPACT REPORTED. REPAIR ESCALATED TO PRODUCT EVENT ON EVALUATION DUE TO LEG OF THE ATTACHMENT WAS BENT. ON FOLLOW UP IT WAS REPORTED THAT THERE IS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16723082 ↗MalfunctionNov 22, 2022Material Twisted/Bent; No Apparent Adverse Event; Material Twisted/Bent; No Apparent Adverse Event
REPAIR REQUEST INITIATED FOR DEVICE WITH NO REPORTED MALFUNCTION. NO PATIENT IMPACT REPORTED. REPAIR ESCALATED TO PRODUCT EVENT ON EVALUATION DUE TO LEG OF THE ATTACHMENT WAS BENT. ADDITIONAL INFORMATION REGARDING THE PATIENT IMPACT WAS BEING FOLLOWED UP FROM THE FACILITY. DATE OF EVENT NOTIFICATION: ON (B)(6) 2022.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15842670 ↗MalfunctionJul 1, 2022Material Twisted/Bent; Material Twisted/Bent
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF ATTACHMENT DAMAGED OR BENT LEG. NO PATIENT IMPACT WAS REPORTED. REPAIR ESCALATED TO A PRODUCT EVENT BASED ON REASON FOR RETURN. ADDITIONAL INFORMATION REGARDING THE PATIENT IMPACT WAS BEING FOLLOWED UP FROM THE FACILITY. DATE OF EVENT NOTIFICATION: (B)(6) 2022.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14883435 ↗
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page