Midas Rex MR8 MR8-AF01 (UDI 00643169912397)Recalls & reported problems

Medtronic Powered Surgical Solutions · ATT MR8 MR8-AF01 FOOTED F1 B-5 2.4MM · Class 2

All models of MIDAS REX MR8 (137)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00643169912397), received between Jul 1, 2022 and Jun 26, 2026 — classified as 13 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 5 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent5
  • No Apparent Adverse Event3
  • Overheating of Device3
  • Detachment of Device or Device Component2
  • Naturally Worn2
  • Electrical /Electronic Property Problem1
  • Noise, Audible1
  • Vibration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions13
  • Insufficient Information1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 26, 2026Naturally Worn

    IT IS REPORTED THAT THE DEVICE BEARING BROKEN. IT IS REPORTED THAT THERE IS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25634363 ↗
  • MalfunctionJun 19, 2026Overheating of Device

    IT WAS REPORTED THAT PRIOR TO USE THE DEVICE HAD TEMPERATURE INCREASE ISSUE. THERE WAS NO PATIENT IMPACT REPORTED.  ADDITIONAL INFORMATION WAS RECEIVED STATING THAT CHIPPED DISTAL BEARING WAS FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25561757 ↗
  • MalfunctionJun 18, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST, AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25546531 ↗
  • MalfunctionJun 10, 2026Overheating of Device

    IT WAS REPORTED THAT, PRIOR TO USE, THE DEVICE WAS OBSERVED TO HAVE A TEMPERATURE INCREASE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25456469 ↗
  • MalfunctionJun 1, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A SCHEDULED PREVENTIVE MAINTENANCE, AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DISTAL BEARINGS WERE DETACHED, TIP OF THE TUBE IS OUT OF SHAPE, LASER MARKINGS AND COLOR BAND WERE FOUND FADED DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25355227 ↗
  • MalfunctionMay 29, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE BEARING WERE FELL OUT DURING A PROCEDURE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DISTAL BEARING WAS DETACHED WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25328453 ↗
  • MalfunctionMay 19, 2026Vibration; Detachment of Device or Device Component; Naturally Worn; Noise, Audible

    IT WAS REPORTED THAT THE BEARINGS WERE BAD, LOUD AND ¿RUNNING ROUGH¿. NO ADDITIONAL INFORMATION WAS RECEIVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 25222359 ↗
  • MalfunctionMar 31, 2026Overheating of Device

    IT WAS REPORTED THAT PRE-OPERATIVELY THERE WAS TEMPERATURE INCREASE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24744160 ↗
  • MalfunctionAug 6, 2024Material Twisted/Bent; Material Twisted/Bent

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF BENT FOOT. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. REPAIR ESCALATED TO A PRODUCT EVENT BASED ON REASON FOR RETURN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19921459 ↗
  • MalfunctionAug 6, 2024Material Twisted/Bent; Material Twisted/Bent

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF BENT FOOT. NO PATIENT IMPACT REPORTED. REPAIR IS BEING ESCALATED TO PRODUCT EVENT DUE TO REASON FOR THE RETURN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19915663 ↗
  • MalfunctionApr 12, 2023Electrical /Electronic Property Problem; Material Twisted/Bent; Electrical /Electronic Property Problem; Material Twisted/Bent

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF FROZEN. NO PATIENT IMPACT REPORTED. REPAIR ESCALATED TO PRODUCT EVENT ON EVALUATION DUE TO LEG OF THE ATTACHMENT WAS BENT. ON FOLLOW UP IT WAS REPORTED THAT THERE IS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16723082 ↗
  • MalfunctionNov 22, 2022Material Twisted/Bent; No Apparent Adverse Event; Material Twisted/Bent; No Apparent Adverse Event

    REPAIR REQUEST INITIATED FOR DEVICE WITH NO REPORTED MALFUNCTION. NO PATIENT IMPACT REPORTED. REPAIR ESCALATED TO PRODUCT EVENT ON EVALUATION DUE TO LEG OF THE ATTACHMENT WAS BENT. ADDITIONAL INFORMATION REGARDING THE PATIENT IMPACT WAS BEING FOLLOWED UP FROM THE FACILITY. DATE OF EVENT NOTIFICATION: ON (B)(6) 2022.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15842670 ↗
  • MalfunctionJul 1, 2022Material Twisted/Bent; Material Twisted/Bent

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF ATTACHMENT DAMAGED OR BENT LEG. NO PATIENT IMPACT WAS REPORTED. REPAIR ESCALATED TO A PRODUCT EVENT BASED ON REASON FOR RETURN. ADDITIONAL INFORMATION REGARDING THE PATIENT IMPACT WAS BEING FOLLOWED UP FROM THE FACILITY. DATE OF EVENT NOTIFICATION: (B)(6) 2022.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14883435 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices