NA 5407-120-450Recalls & reported problems
Stryker Corporation · Elite Straight Attachment · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker Elite straight attachment is used with powered surgical drills. Its 148 reports describe overheating, flaking, and parts detaching, and a few mention burns. The reports include no injuries and no deaths. Reports fell to 9 in the last 12 months from 22 in the 12 months before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 149
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 149 adverse event reports that list this exact device (DI 07613327096545), received between Jul 24, 2018 and Jul 22, 2026 — classified as 149 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (55 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device55
- Flaked42
- Detachment of Device or Device Component33
- Fracture20
- Fluid/Blood Leak7
- Material Fragmentation5
- Break3
- Corroded2
Patient problems
- No Clinical Signs, Symptoms or Conditions116
- No Patient Involvement26
- No Consequences Or Impact To Patient19
- No Known Impact Or Consequence To Patient7
- Insufficient Information5
- Burn(s)4
- Chemical Exposure1
Recent adverse event reports
The 25 most recent of 149 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 22, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25907628 ↗MalfunctionMay 27, 2026
UNABLE TO LOCK DRILL BIT IN PLACE. DR. ATTEMPTED TO SECURE DRILL BIT OVER STERILE FIELD AND THREE SMALL METAL PIECES FELL OUT OF DRILL ATTACHMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25300673 ↗MalfunctionApr 22, 2026Detachment of Device or Device Component
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1 ¿ MARCH 31 2026. THIS REPORT SUMMARIZES 17 EVENTS FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT. - 10 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 6 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT. - 1 EVENT HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24954385 ↗MalfunctionApr 22, 2026Overheating of Device
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1 ¿ MARCH 31 2026. THIS REPORT SUMMARIZES 99 EVENTS FOR THE FAILURE: OVERHEATING OF DEVICE. - 67 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 24 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT. - 7 EVENTS HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD INSUFFICIENT INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24954375 ↗MalfunctionApr 22, 2026Material Fragmentation
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1 ¿ MARCH 31, 2026. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: MATERIAL FRAGMENTATION. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24954014 ↗MalfunctionApr 22, 2026Overheating of Device; Detachment of Device or Device Component
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1 ¿ MARCH 31 2026. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURES: OVERHEATING OF DEVICE; DETACHMENT OF DEVICE OR DEVICE COMPONENT. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24953820 ↗MalfunctionApr 22, 2026Flaked
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1 ¿ MARCH 31 2026. THIS REPORT SUMMARIZES 3 EVENTS FOR THE FAILURE: FLAKED. 3 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24953593 ↗MalfunctionJan 23, 2026Flaked
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) 2025. THIS REPORT SUMMARIZES 19 EVENTS FOR THE FAILURE: FLAKED. - 19 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24163596 ↗MalfunctionJan 23, 2026Fracture
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24163410 ↗MalfunctionJan 23, 2026Detachment of Device or Device Component
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31 2025. THIS REPORT SUMMARIZES 13 EVENTS FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT. - 12 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24163083 ↗MalfunctionOct 28, 2025Fluid/Blood Leak; Overheating of Device; Fluid/Blood Leak; Overheating of Device
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURES: FLUID/BLOOD LEAK; OVERHEATING OF DEVICE. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23399960 ↗MalfunctionOct 28, 2025Fracture; Fracture
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23399930 ↗MalfunctionOct 28, 2025Overheating of Device; Overheating of Device
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 85 EVENTS FOR THE FAILURE: OVERHEATING OF DEVICE. - 62 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 19 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT. - 4 EVENTS HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23399836 ↗MalfunctionOct 28, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 25 EVENTS FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT, 6 EVENTS HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT, 14 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT, 5 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23399763 ↗MalfunctionOct 23, 2025Flaked; Flaked
MalfunctionOct 23, 2025Overheating of Device; Detachment of Device or Device Component; Overheating of Device; Detachment of Device or Device Component
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURES: OVERHEATING OF DEVICE; DETACHMENT OF DEVICE OR DEVICE COMPONENT. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23368922 ↗MalfunctionJul 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 19 EVENTS FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT. 15 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 3 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT. 1 EVENT HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22647064 ↗MalfunctionJul 29, 2025Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 58 EVENTS FOR THE FAILURE: OVERHEATING OF DEVICE. 42 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 15 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT. 1 EVENT HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22646531 ↗MalfunctionJul 24, 2025Fracture; Fracture
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN APRIL 1 ¿ JUNE 30, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22587170 ↗MalfunctionJul 24, 2025Overheating of Device; Detachment of Device or Device Component; Overheating of Device; Detachment of Device or Device Component
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN APRIL 1 ¿ JUNE 30, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURES: OVERHEATING OF DEVICE; DETACHMENT OF DEVICE OR DEVICE COMPONENT. - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22587128 ↗MalfunctionJul 24, 2025Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FLUID/BLOOD LEAK. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22587111 ↗MalfunctionJul 24, 2025Flaked; Flaked
THIS REPORT SUMMARIZES 31 EVENTS FOR THE FAILURE: FLAKED. - 31 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22586856 ↗MalfunctionApr 21, 2025Fracture; Fracture
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 3 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876860 ↗MalfunctionApr 21, 2025Flaked; Flaked
THIS REPORT SUMMARIZES 17 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD PAINT CHIPS MISSING. - 17 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876834 ↗MalfunctionApr 21, 2025Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 56 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 47 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. - 8 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876668 ↗
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page