Midas Rex MR8 MR8-AVS14 (UDI 00643169912632)Recalls & reported problems
Medtronic Powered Surgical Solutions · ATT MR8 MR8-AVS14 14CM VAR STR 3.2MM · Class 2
All models of MIDAS REX MR8 (137)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00643169912632), received between Aug 25, 2025 and Sep 29, 2026 — classified as 1 injury, 13 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device7
- Detachment of Device or Device Component4
- No Apparent Adverse Event2
- Difficult to Insert1
- Vibration1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Insufficient Information2
- Burn(s)1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026No Apparent Adverse Event
Medtronic received information regarding a repair request with no alleged product deficiency. There was no patient impact. Additional information was received stating bearings misaligned, laser markings faded, and o-ring worn was found during evaluation.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50930541 ↗MalfunctionSep 27, 2026No Apparent Adverse Event
Patient problems: Insufficient Information
MAUDE report 50910282 ↗MalfunctionJul 2, 2026Overheating of Device
IT WAS REPORTED THAT DURING INTRA-OPERATIVE USE OF AN ATTACHMENT, IT BECAME OVERHEATED. IT WAS FURTHER REPORTED THAT THERE WAS NO IMPACT ON THE PATIENT OR STAFF. ADDITIONAL INFORMATION RECEIVED AS BEARINGS MISALIGNED FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25699677 ↗MalfunctionJul 2, 2026Overheating of Device
IT WAS REPORTED THAT DURING INTRA-OPERATIVE USE OF AN ATTACHMENT, IT BECAME OVERHEATED. IT WAS FURTHER REPORTED THAT THERE WAS NO IMPACT ON THE PATIENT OR STAFF. ADDITIONAL INFORMATION RECEIVED AS BEARINGS MISALIGNED FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25699626 ↗MalfunctionJun 26, 2026Overheating of Device
IT WAS REPORTED THAT THE ATTACHMENT WAS OVERHEATING PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25633610 ↗MalfunctionJun 19, 2026Difficult to Insert
IT WAS REPORTED THAT WHILE SERVICING, THE TOOLS WERE NOT LOCKING IN THE ATTACHMENT. THERE WAS NO PATIENT IMPACT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25565981 ↗MalfunctionJun 16, 2026Overheating of Device
IT WAS REPORTED THAT THE ATTACHMENT WAS OVERHEATING OBSERVED PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25514526 ↗MalfunctionJun 8, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE ATTACHMENT STUCK AND WILL NOT EXTEND TO ATTACH. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THERE WERE DETACHED BEARINGS INSIDE THE DEVICE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25427303 ↗MalfunctionJun 5, 2026Overheating of Device; Vibration
IT WAS REPORTED THAT THE DEVICE IS RATTLING AND HEATING UP DURING THE PROCEDURE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHED WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25403975 ↗MalfunctionJun 2, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT A PIECE FELL OUT FROM INSIDE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DISTAL BEARING WAS CHIPPED WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25361793 ↗MalfunctionMay 27, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT ATTACHMENT BEARINGS FELL OUT. AT THE TIME OF REPORT, THERE WAS NO KNOWN IMPACT TO A PATIENT.
Patient problems: Insufficient Information
MAUDE report 25296341 ↗MalfunctionMay 25, 2026Overheating of Device
IT WAS REPORTED THAT THE DEVICE TEMPERATURE WAS INCREASE PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARING WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25270133 ↗MalfunctionMay 25, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT PRIOR TO USE, BEARING FELL OUT FROM THE ATTACHMENT. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25268526 ↗InjuryAug 25, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THAT POTENTIAL SMALL BURN TO THE PATIENTS SKIN.
Patient problems: Burn(s)
MAUDE report 22882085 ↗
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.