Skip to content

Midas Rex MR8 MR8-AVA07 (UDI 00643169912649)Recalls & reported problems

Medtronic Powered Surgical Solutions · ATT MR8 MR8-AVA07 7CM VARI ANG 2.4MM · Class 2

All models of MIDAS REX MR8 (137)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00643169912649), received between Jun 21, 2024 and Aug 17, 2026 — classified as 14 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 1 reportsJun 2026: 3 reportsJul 2026: 4 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Apparent Adverse Event4
  • Overheating of Device4
  • Difficult to Remove2
  • Naturally Worn2
  • Detachment of Device or Device Component1
  • Difficult to Insert1
  • Loss of Power1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 17, 2026No Apparent Adverse Event

    Medtronic received information regarding a repair request and no alleged product deficiencies were reported. Additional information was received stating that detached bearings were found during evaluation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50273849 ↗
  • MalfunctionJul 4, 2026Naturally Worn

    IT WAS REPORTED THAT THE ATTACHMENT IS ROLLING ISSUE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25709821 ↗
  • MalfunctionJul 3, 2026Difficult to Remove

    IT WAS REPORTED THAT THE DEVICE ADJUSTMENT RING BLOCKED. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25706659 ↗
  • MalfunctionJul 3, 2026Overheating of Device

    IT WAS REPORTED THAT THE DEVICE WAS OVERHEATING WHILE ON MOTOR. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHMENT WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25701166 ↗
  • MalfunctionJul 1, 2026Overheating of Device

    IT WAS REPORTED THAT THE ATTACHMENT WAS OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHMENT WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25675260 ↗
  • MalfunctionJun 18, 2026Naturally Worn

    IT WAS REPORTED THAT THE DEVICE WAS BROKEN. THERE IS NO PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25544974 ↗
  • MalfunctionJun 10, 2026Difficult to Remove

    IT WAS REPORTED THAT ATTACHMENT WAS STUCK. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25462913 ↗
  • MalfunctionJun 5, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST, AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. THERE IS NO PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION RECEIVED STATED THAT THERE WAS BEARING DETACHMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25405416 ↗
  • MalfunctionMay 20, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST, AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT INPUT AND OUTPUT DRIVE SHAT WAS CORRODED, BEARING INTERNAL TUBE IS DETACHED, TUBE BEARINGS ARE WORN AND COLOR RING AND LASER MARKINGS WERE FOUND FADED DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25229065 ↗
  • MalfunctionApr 17, 2026Loss of Power

    IT WAS REPORTED THAT THE HANDPIECE SEIZED. THERE WAS NO PATIENT IMPACT AS A RESULT OF THE EVENT. ADDITIONAL INFORMATION RECEIVED AFTER EVALUATION AS DETACHED BEARING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24907070 ↗
  • MalfunctionApr 1, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE DEVICE WAS DEFECTIVE. AT THE TIME OF REPORT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24751718 ↗
  • MalfunctionMar 18, 2025Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. NO PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21630666 ↗
  • MalfunctionSep 18, 2024Difficult to Insert; Detachment of Device or Device Component; Difficult to Insert; Detachment of Device or Device Component

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF DIFFICULTY WITH ASSEMBLY. NO PATIENT IMPACT REPORTED. REPAIR ESCALATED TO PRODUCT EVENT ON EVALUATION DUE TO DETACHED BEARING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20249193 ↗
  • MalfunctionJun 21, 2024Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. NO PATIENT IMPACT REPORTED. REPAIR REQUEST ESCALATED TO PRODUCT EVENT DUE TO REASON FOR RETURN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19582408 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices

All DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices →

Sources and updates