Skip to content

Midas Rex MR8 MR8-AVA10 (UDI 00643169912656)Recalls & reported problems

Medtronic Powered Surgical Solutions · ATT MR8 MR8-AVA10 10CM VAR ANG 2.4MM · Class 2

All models of MIDAS REX MR8 (137)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Midas Rex MR8 attachment is a 10 cm angled surgical drill burr. FDA records include 38 reports on it, with no deaths and no injuries. The reports mostly describe the attachment not working before or during use: bearings or rings coming detached, the tool vibrating, overheating, losing power, or being hard to fit. Most mention no patient harm at all, and no recalls name this device. Reports rose sharply, from 4 in the prior 12 months to 27 in the latest 12.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 00643169912656), received between Nov 15, 2022 and Sep 28, 2026 — classified as 40 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 4 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 9 reportsMay 2026: 1 reportsJun 2026: 5 reportsJul 2026: 10 reportsAug 2026: 1 reportsSep 2026: 3 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Apparent Adverse Event17
  • Detachment of Device or Device Component7
  • Overheating of Device7
  • Difficult to Insert4
  • Loss of Power4
  • Vibration3
  • Naturally Worn2

Patient problems

  • No Clinical Signs, Symptoms or Conditions35
  • Insufficient Information5

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Difficult to Insert

    It was reported that the attachment with the report of difficulty with assembly. It was reported that there was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911122 ↗
  • MalfunctionSep 15, 2026No Apparent Adverse Event

    Medtronic received information regarding a loaner device return with no alleged product deficiencies. No patient involvement in this event. Additional information was received stating that bearing detachment was found during evaluation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50770768 ↗
  • MalfunctionSep 4, 2026No Apparent Adverse Event

    Medtronic received information regarding a planned maintenance with no alleged product deficiency. There was no patient involvement. Additional information was received stating bearing detachment was found during evaluation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50609348 ↗
  • MalfunctionAug 8, 2026Difficult to Insert

    It was reported that during preparing for use attachments was not working. It was reported there was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50187250 ↗
  • MalfunctionJul 6, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE ATTACHMENT WAS NOT WORKING PROPERLY. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHMENT WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25715955 ↗
  • MalfunctionJul 6, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE ATTACHMENT WAS DEFECTIVE. AT THE TIME OF REPORT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION RECEIVED UPON EVALUATION THAT THE DISTAL BEARINGS WERE DETACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25714975 ↗
  • MalfunctionJul 3, 2026Vibration

    IT WAS REPORTED THAT THE ATTACHMENT DRILL WAS VIBRATING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25704164 ↗
  • MalfunctionJul 3, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARING WAS FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25703338 ↗
  • MalfunctionJul 3, 2026Naturally Worn

    IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF BROKEN RING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25701281 ↗
  • MalfunctionJul 2, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARING WAS FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25696289 ↗
  • MalfunctionJul 2, 2026Difficult to Insert

    IT WAS REPORTED THAT THE ATTACHMENT DOES NOT RETAIN. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25687206 ↗
  • MalfunctionJul 1, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST WITH NO ALLEGED PRODUCT DEFICIENCY. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING BROKEN BEARINGS WAS FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25677462 ↗
  • MalfunctionJul 1, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THERE WERE NO ALLEGED ISSUE WITH THE DEVICE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25676438 ↗
  • MalfunctionJul 1, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THERE WAS NO ALLEGED ISSUE AT THE TIME OF INTAKE. NO PATIENT IMPACT REPORTED.  ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: Insufficient Information

    MAUDE report 25676048 ↗
  • MalfunctionJun 30, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. NO PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHED WERE FOUND DURING EVALUATION.

    Patient problems: Insufficient Information

    MAUDE report 25669196 ↗
  • MalfunctionJun 25, 2026Loss of Power

    IT WAS REPORTED THAT PRIOR TO USE THE ATTACHMENT WAS NOT WORKING. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION RECEIVED STATING BEARING DETACHED WAS FOUND DURING EVALUATION MADE IT REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25617733 ↗
  • MalfunctionJun 19, 2026Loss of Power

    IT WAS REPORTED THAT THE DEVICE SEIZED. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25559033 ↗
  • MalfunctionJun 19, 2026Vibration; Detachment of Device or Device Component

    IT WAS REPORTED THAT IN THE ATTACHMENT DRILL, SWINGS ENORMOUSLY. NO PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: Insufficient Information

    MAUDE report 25557121 ↗
  • MalfunctionJun 5, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE DEVICE BEARING IS FALLING PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25404805 ↗
  • MalfunctionMay 6, 2026Loss of Power

    IT WAS REPORTED THAT AN ATTACHMENT WAS NOT WORKING, IT WAS OBSERVED PRE-OPERATIVELY WHILE PREPARING FOR USE. THERE WAS NO PATIENT IN VOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25093087 ↗
  • MalfunctionApr 20, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST, AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24922715 ↗
  • MalfunctionApr 20, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT CUSTOMER HAD DISSATISFACTION WITH THE ATTACHMENT. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. ADDITIONAL INFORMATION RECEIVED AFTER PRODUCT EVALUATION THAT BEARING WAS DETACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24921564 ↗
  • MalfunctionApr 17, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE ATTACHMENT WAS NOT WORKING AND COULD NOT ATTACH PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION RECEIVED AFTER EVALUATION WAS DETACHED BEARING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24907814 ↗
  • MalfunctionApr 14, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT A SERVICE REQUEST WAS CREATED FOR THE ATTACHMENT AND IT WAS DESCRIBED AS DEFECTIVE. IT WAS UNKNOWN IF THERE WAS ANY PATIENT INVOLVEMENT OR COMPLICATIONS AS A RESULT OF THE EVENT.

    Patient problems: Insufficient Information

    MAUDE report 24868269 ↗
  • MalfunctionApr 13, 2026Difficult to Insert

    IT WAS REPORTED THAT THE ATTACHMENT DOESN'T HOLD THE TOOL. IT WAS REPORTED THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24854480 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices