Avalus™ 40025Recalls & reported problems

Medtronic, Inc. · VALVE 40025 AVALUS AOR MMX U30 US · Class 3 · Implant

Quick summary

This Medtronic Avalus heart valve is a tissue valve implanted in the heart. Of its 90 reports, 4 involve deaths and 71 involve injuries. Reports mostly describe patients needing a pacemaker after implant because of heart rhythm problems, valves that narrowed or leaked, and valves being replaced later. There were 9 reports in the last 12 months, down from 16 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
89
Deaths reported
4
Injuries reported
70
Recalls naming it
0

Summary

The FDA's MAUDE database contains 89 adverse event reports that list this exact device (DI 00643169983205), received between Aug 20, 2018 and Aug 8, 2026 — classified as 4 deaths, 70 injuries, 15 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (38 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 3 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 4 reportsOct 2024: 7 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 2 reportsFeb 2025: 3 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 1 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information38
  • Patient Device Interaction Problem23
  • Inadequacy of Device Shape and/or Size19
  • Adverse Event Without Identified Device or Use Problem13
  • Device Stenosis8
  • Intravalvular regurgitation7
  • Gradient Increase5
  • Incomplete Coaptation2

Patient problems

  • Insufficient Information42
  • Aortic Valve Stenosis10
  • Aortic Valve Insufficiency/ Regurgitation8
  • Endocarditis8
  • Heart Block8
  • No Information7
  • No Clinical Signs, Symptoms or Conditions6
  • Non specific EKG/ECG Changes5

Recent adverse event reports

The 25 most recent of 89 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 8, 2026Device Stenosis

    Medtronic received information that 6 years and 10 months post implant of this 25mm Avalus aortic bioprosthetic valve, the patient is being evaluated for a transcatheter valve-in-valve replacement. The reason for evaluation was reported as aortic stenosis (AS). Intervention is required. No additional adverse patient effects were reported.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 50187218 ↗
  • MalfunctionMar 17, 2026Packaging Problem

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE VALVE JAR WAS BROKEN UPON OPENING THE PRODUCT PACKAGING. IT WAS FURTHER NOTED THAT THE VALVE WAS NOT USED. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24619719 ↗
  • InjuryMar 6, 2026Patient Device Interaction Problem

    IT WAS REPORTED THAT 3 YEARS AND 10 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SICK SINUS SYNDROME (SSS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 24533800 ↗
  • InjuryFeb 10, 2026Patient Device Interaction Problem

    IT WAS REPORTED THAT APPROXIMATELY 1 MONTH POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS COMPLETE HEART BLOCK (CHB). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Heart Block

    MAUDE report 24305111 ↗
  • InjuryDec 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT 3 YEARS AND 3 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN DEVICE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 23875438 ↗
  • InjuryDec 22, 2025Gradient Increase; Insufficient Device Problem Information; Device Stenosis; Gradient Increase; Insufficient Device Problem Information; Device Stenosis

    IT WAS REPORTED THAT 5 YEARS AND 3 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A 29MM TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis; Insufficient Information

    MAUDE report 23862061 ↗
  • InjuryDec 4, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 3 YEARS AND 5 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 23MM VALVE OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 23717746 ↗
  • MalfunctionNov 26, 2025Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information; Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 23MM VALVE OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 23659302 ↗
  • InjuryOct 28, 2025Insufficient Device Problem Information; Intravalvular regurgitation; Insufficient Device Problem Information; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 5 MONTHS POST IMPLANT OF THIS 25MM AVALUS AORTIC VALVE, THE PATIENT IS BEING EVALUATED FOR A VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Insufficient Information

    MAUDE report 23400935 ↗
  • InjuryOct 23, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT AN UNKNOWN DURATION POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 23367654 ↗
  • InjuryJul 18, 2025Intravalvular regurgitation; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT POST IMPLANT OF THIS AVALUS AORTIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC REGURGITATION (AR). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 22540037 ↗
  • InjuryJul 17, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A BI-VENTRICULAR DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE DEFIBRILLATOR IMPLANT WAS REPORTED AS COMPLETE HEART BLOCK, CARDIOMYOPATHY, DYSSYNCHRONY AND LOW EJECTION FRACTION (EF). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Cardiomyopathy; Heart Block; Non specific EKG/ECG Changes

    MAUDE report 22530857 ↗
  • InjuryJun 24, 2025Insufficient Device Problem Information; Patient Device Interaction Problem; Insufficient Device Problem Information; Patient Device Interaction Problem

    IT WAS REPORTED THAT POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE DEFIBRILLATOR IMPLANT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Cardiomyopathy; Insufficient Information

    MAUDE report 22318186 ↗
  • InjuryMar 19, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT POST IMPLANT OF THIS 25MM AVALUS AORTIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 21MM AVALUS AORTIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 21642248 ↗
  • InjuryMar 7, 2025Patient Device Interaction Problem; Device Stenosis; Intravalvular regurgitation; Patient Device Interaction Problem; Device Stenosis; Intravalvular regurgitation

    ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE PATIENT HAD BEEN DIAGNOSED WITH SUBACUTE BACTERIAL ENDOCARDITIS. THE ORGANISM CULTURED WAS NOT REPORTED. IT WAS ALSO REPORTED THE PATIENT HAD NON-RHEUMATIC AORTIC INSUFFICIENCY. SUBSEQUENTLY, 3 MONTHS AFTER EVALUATION, THE 25MM AORTIC BIOPROSTHETIC VALVE WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Endocarditis

    MAUDE report 21547730 ↗
  • InjuryMar 5, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 1 DAY POST IMPLANT OF THIS 25 MM AORTIC BIOPROSTHETIC VALVE, IT WAS REPORTED THAT THE PATIENT WAS SLIGHTLY TACHYCARDIC AND WAS ADMINISTERED MEDICATION. IT WAS FURTHER REPORTED THAT 3 DAYS POST IMPLANT OF THE VALVE, THE PATIENT WENT INTO ATRIAL FIBRILLATION (AFIB) AND WAS ADMINISTERED AN ANTIARRHYTHMIC MEDICATION. IT WAS REPORTED THAT 4 DAYS POST IMPLANT, THE PATIENT EXPERIENCED ATRIAL FLUTTER. THE PATIENT WAS STARTED ON ANTICOAGULATION MEDICATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Atrial Fibrillation; Atrial Flutter; Tachycardia

    MAUDE report 21526783 ↗
  • InjuryFeb 20, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING THE IMPLANT OF THIS AVALUS AORTIC VALVE, IT WAS IMMEDIATELY EXPLANTED. NO NEW VALVE WAS IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 21420122 ↗
  • InjuryFeb 18, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 7 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 25MM VALVE OF A DIFFERENT MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 21405272 ↗
  • InjuryFeb 10, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 2 YEARS AND 9 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 25MM VALVE OF A NEW MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 21345574 ↗
  • MalfunctionJan 29, 2025Gradient Increase; Intravalvular regurgitation; Gradient Increase; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 3 YEARS AND 4 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AN INCREASED MEAN GRADIENT OF 33MMHG AND MODERATE AORTIC INSUFFICIENCY. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 21268071 ↗
  • InjuryJan 9, 2025Gradient Increase; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation; Gradient Increase; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation

    IT WAS REPORTED THAT DURING A PROCEDURE, AN OLD AORTIC TISSUE VALVE WAS EXPLANTED AND A NEW AORTIC TISSUE VALVE WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information

    MAUDE report 21110416 ↗
  • InjuryDec 13, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 8 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 25MM VALVE OF A DIFFERENT MODEL. THE REASON FOR THE REPLACEMENT WAS REPORTED AS ENDOCARDITIS INFECTION POSSIBLY DUE TO CLOSTRIDIOSES DIFFICILE. DURING EXPLANT, IT WAS DISCOVERED THAT THERE WAS A THROMBUS ON THE VALVE. IT WAS FURTHER REPORTED THAT ENDOCARDITIS HAS NOT BEEN CONFIRMED AS NO ORGANISMS HAVE BEEN CULTURED, AND THE PHYSICIAN BELIEVES THIS IS DUE TO ANTIBIOTIC TREATMENTS PRIOR TO SURGERY. IT WAS REPORTED THAT THE SOURCE OF THE INFECTION IS BELIEVED TO BE DUE TO A RECENT DENTAL PROCEDURE AND IS NOT RELATED TO THE DEVICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Thrombosis/Thrombus; Unspecified Infection

    MAUDE report 20926425 ↗
  • InjuryDec 11, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT THE SAME DAY POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, POST OPERATIONAL PAROXYSMAL ATRIAL FIBRILLATION (A.FIB) WAS OBSERVED. MEDICATION WAS ADMINISTERED. NO INTERVENTION OR ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Atrial Fibrillation

    MAUDE report 20901754 ↗
  • InjuryNov 19, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 8 MONTHS POST IMPLANT OF THIS 25 MM AORTIC BIOPROSTHETIC VALVE, IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN ISCHEMIC CEREBRAL VASCULAR ACCIDENT (CVA). THE PHYSICIAN BELIEVED THE MAIN CAUSE OF THE CVA WAS A CURRENT INFECTION. IT WAS NOTED THAT A THROMBUS WAS OBSERVED ON THE VALVE. NO INTERVENTION OR ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Stroke/CVA

    MAUDE report 20723683 ↗
  • InjuryOct 21, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION FROM THE PATIENT THAT 2 DAYS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS COMPLETE HEART BLOCK (CHB). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Heart Block

    MAUDE report 20492365 ↗

Other heart-valve, non-allograft tissue devices

All heart-valve, non-allograft tissue devices →