OASIS 3612-100Recalls & reported problems

Atrium Medical Corporation · Atrium Oasis Chest Drain Infant/Pediatric Collection Dry Suction Water Seal · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
18
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00650862111019), received between Jun 13, 2019 and Apr 28, 2026 — classified as 4 injuries, 14 malfunctions.

The most frequently reported specific device problem is “Product Quality Problem” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 18, 2023Open, Classified

    Instructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains

    Z-0317-2024 · Root cause: Error in labeling · FDA record ↗

Most reported problems

Device problems

  • Product Quality Problem4
  • Adverse Event Without Identified Device or Use Problem2
  • Overfill2
  • Appropriate Device Problem Term/Code Not Available1
  • Break1
  • Calibration Problem1
  • Component Missing1
  • Component or Accessory Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient5
  • Pneumothorax2
  • Aspiration Pneumonitis1
  • Insufficient Information1
  • Low Oxygen Saturation1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 28, 2026Adverse Event Without Identified Device or Use Problem

    THE OASIS DRAIN WAS REPLACED DUE TO A PNEUMOTHORAX THAT WAS NOT RESOLVING AS EXPECTED. FOLLOWING REPLACEMENT WITH ANOTHER OASIS DRAIN, A LACK OF VIGOROUS BUBBLING WAS OBSERVED, HOWEVER THE PATIENT REMAINED CLINICALLY STABLE. REVIEW CONFIRMED THE PRESENCE OF A SMALL AIR LEAK WITH APPROPRIATE DEVICE FUNCTION. A SUBSEQUENT CHEST XRAY DEMONSTRATED IMPROVEMENT IN THE PNEUMOTHORAX. THERE WAS NO HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25014706 ↗
  • MalfunctionApr 23, 2026Adverse Event Without Identified Device or Use Problem

    THE OASIS DRAIN WAS REPLACED DUE TO A PNEUMOTHORAX THAT WAS NOT RESOLVING AS EXPECTED. FOLLOWING REPLACEMENT WITH ANOTHER OASIS DRAIN, A LACK OF VIGOROUS BUBBLING WAS OBSERVED, HOWEVER THE PATIENT REMAINED CLINICALLY STABLE. REVIEW CONFIRMED THE PRESENCE OF A SMALL AIR LEAK WITH APPROPRIATE DEVICE FUNCTION. A SUBSEQUENT CHEST XRAY DEMONSTRATED IMPROVEMENT IN THE PNEUMOTHORAX. THERE WAS NO HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24961597 ↗
  • MalfunctionMay 29, 2025Inadequate Instructions for Healthcare Professional; Component or Accessory Incompatibility; Inadequate Instructions for Healthcare Professional; Component or Accessory Incompatibility

    N/A.

    Patient problems: Insufficient Information

    MAUDE report 22104454 ↗
  • MalfunctionMar 19, 2025Missing Information; Missing Information

    IT WAS REPORTED THAT WHILE USING AN OASIS PEDIATRIC DRAIN ON A PATIENT THE DOCTOR AND NURSING STAFF WERE CONFUSED WITH DRY SUCTION/WET SEAL ON THE DRAIN AS THEY WANTED TO KNOW THE PROCESS TO CONVERT THE DRAIN TO WET SEAL FROM SUCTION AND THE STEPS NEEDED TO BE TAKEN. ONLY INSTRUCTIONS WITH THE DRAIN WAS PICTURES OF AN OASIS DRAIN NO WRITTEN INSTRUCTIONS OR DETAILS FOR DOCTOR AND NURSING STAFF TO REVIEW. DRAIN WAS SET UP PROPERLY. NO HARM OR INJURY REPORTED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21649449 ↗
  • InjuryJan 13, 2025Overfill; Overfill

    PICU RN CALLED FOR ASSISTANCE WITH ATRIUM OASIS PEDIATRIC DRAIN. INFANT HAD BEEN ADMITTED FOR PNEUMOTHORAX AND HAD A CHEST TUBE PLACED. PATIENT THEN DEVELOPED A LARGER PNEUMOTHORAX BELOW THE FIRST, SO A SECOND CHEST TUBE WAS PLACED AFTER NEEDLE ASPIRATION. THE CALLER DESCRIBED THAT THE BLUE WATER IN THE AIR LEAK MONITOR CHAMBER C OF THE SECOND CHEST DRAIN HAD OVERFILLED TO WATER SEAL CHAMBER B, MEANING THAT THERE WAS AN ACUTE RISE IN THE WATER SEAL CHAMBER LEVEL. THE DRAIN WAS CHANGED OUT AND THE SAME THING WAS OBSERVED ON THE REPLACEMENT DRAIN. SUPPORT TEAM DISCUSSED WITH THE CALLER THAT THIS COULD BE INDICATIVE OF A RISE IN THE PATIENT¿S INTRATHORACIC PRESSURE, AND POSSIBLE CLINICAL REASONS. NO DRAINAGE WAS OBSERVED IN COLLECTION CHAMBER D. AFTER REPEAT CHEST XRAY IT WAS DETERMINED THAT THE PATIENT¿S PNEUMOTHORAX HAD INCREASED IN SIZE AND CARE TEAM ATTEMPTED ASPIRATION FROM THE CHEST TUBE. REPORTEDLY RESISTANCE WAS ENCOUNTERED DURING ASPIRATION, BUT EVERYTHING RETURNED TO NORMAL ONCE THE DRAIN WAS RECONNECTED.

    Patient problems: Pneumothorax

    MAUDE report 21146421 ↗
  • InjuryJan 9, 2025Overfill; Overfill

    PICU RN CALLED FOR ASSISTANCE WITH ATRIUM OASIS PEDIATRIC DRAIN. INFANT HAD BEEN ADMITTED FOR PNEUMOTHORAX AND HAD A CHEST TUBE PLACED. PATIENT THEN DEVELOPED A LARGER PNEUMOTHORAX BELOW THE FIRST, SO A SECOND CHEST TUBE WAS PLACED AFTER NEEDLE ASPIRATION. THE CALLER DESCRIBED THAT THE BLUE WATER IN THE AIR LEAK MONITOR CHAMBER C OF THE SECOND CHEST DRAIN HAD OVERFILLED TO WATER SEAL CHAMBER B, MEANING THAT THERE WAS AN ACUTE RISE IN THE WATER SEAL CHAMBER LEVEL. THE DRAIN WAS CHANGED OUT AND THE SAME THING WAS OBSERVED ON THE REPLACEMENT DRAIN. SUPPORT TEAM DISCUSSED WITH THE CALLER THAT THIS COULD BE INDICATIVE OF A RISE IN THE PATIENT¿S INTRATHORACIC PRESSURE, AND POSSIBLE CLINICAL REASONS. NO DRAINAGE WAS OBSERVED IN COLLECTION CHAMBER D. AFTER REPEAT CHEST XRAY IT WAS DETERMINED THAT THE PATIENT¿S PNEUMOTHORAX HAD INCREASED IN SIZE AND CARE TEAM ATTEMPTED ASPIRATION FROM THE CHEST TUBE. REPORTEDLY RESISTANCE WAS ENCOUNTERED DURING ASPIRATION, BUT EVERYTHING RETURNED TO NORMAL ONCE THE DRAIN WAS RECONNECTED.

    Patient problems: Pneumothorax

    MAUDE report 21110803 ↗
  • MalfunctionAug 25, 2023Product Quality Problem; Product Quality Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17623272 ↗
  • MalfunctionAug 25, 2023Product Quality Problem; Product Quality Problem

    REPORT RECEIVED STATED THAT THE CUSTOMER HAS A PEDI OASIS DRAIN THAT HAS A WET-LOOKING SPOT WHERE THE STICKY DOT (THAT HOLDS THE BLUE CAP IN PLACE) HAS ¿LIQUEFIED¿ SOME, AND BLED THROUGH ONTO THE OUTER SIDE OF THE SLEEVE. IT IS SHOWING THROUGH ONTO THE OUTER PORTION OF THE PAPER, APPEARING WET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17623264 ↗
  • MalfunctionAug 25, 2023Product Quality Problem; Product Quality Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17623058 ↗
  • MalfunctionAug 22, 2023Product Quality Problem; Product Quality Problem

    REPORT RECEIVED STATED THAT THE CUSTOMER HAS A PEDI OASIS DRAIN THAT HAS A WET-LOOKING SPOT WHERE THE STICKY DOT (THAT HOLDS THE BLUE CAP IN PLACE) HAS ¿LIQUEFIED¿ SOME, AND BLED THROUGH ONTO THE OUTER SIDE OF THE SLEEVE. IT IS SHOWING THROUGH ONTO THE OUTER PORTION OF THE PAPER, APPEARING WET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17600122 ↗
  • InjuryJun 28, 2023Device Handling Problem; Device Handling Problem

    N/A.

    Patient problems: Aspiration Pneumonitis

    MAUDE report 17214123 ↗
  • MalfunctionFeb 4, 2022Appropriate Device Problem Term/Code Not Available; Component Missing; Appropriate Device Problem Term/Code Not Available

    RECEIVED A USER FACILITY MDR REPORT (3633050000-2021-8033) ON (B)(6) 2022. THE REPORT RECEIVED STATED THAT THE BLUE CAPS WERE MISSING FROM CT DRAINAGE DEVICE. ITEM DISCARDED, ONE FROM SAME LOT NUMBER WAS AVAILABLE. NOTE: THE RESULTS FROM THE DECISION TREE STATE THAT THE EVENT IS NOT REPORTABLE. SINCE THE COMPLAINT WAS OPENED DUE TO A FDA MEDWATCH FORM RECEIVED, THE COMPLAINT IS BEING REPORTED TO THE FDA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13455983 ↗
  • InjuryOct 5, 2021Insufficient Device Problem Information; Therapeutic or Diagnostic Output Failure; Insufficient Device Problem Information

    N/A.

    Patient problems: Low Oxygen Saturation

    MAUDE report 12575711 ↗
  • MalfunctionMar 27, 2020Suction Problem

    NURSE REPORTED THAT OVER THE WEEKEND SHE COULD NOT TURN UP THE SUCTION DIAL FROM -20CMH2O UP TO -30CMH2O. THERE WAS NO IMPACT TO THE PATIENT AND ANOTHER DRAIN WAS ATTACHED WHICH WORKED CORRECTLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9889475 ↗
  • MalfunctionMar 27, 2020Break; Physical Resistance/Sticking

    NURSE EDUCATOR REPORTED THEY CHECKED THE SAMPLE USED FOR TRAINING AND THE REGULATOR WAS STUCK. THEY FORCIBLY TURNED THE DIAL AND IT BROKE. DEVICE WAS NOT USED ON A PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9889470 ↗
  • MalfunctionMar 27, 2020Calibration Problem

    NURSE REPORTED THAT OVER THE WEEKEND SHE COULD NOT TURN UP THE SUCTION DIAL FROM -20CMH2O UP TO -30CMH2O. THERE WAS NO IMPACT TO THE PATIENT AND ANOTHER DRAIN WAS ATTACHED WHICH WORKED CORRECTLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9889395 ↗
  • MalfunctionOct 22, 2019Fluid/Blood Leak

    N/A.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9217804 ↗
  • MalfunctionJun 13, 2019Leak/Splash

    N/A.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8696897 ↗

Other Bottle, Collection, Vacuum devices

All Bottle, Collection, Vacuum devices →