Skip to content

DeRoyal PHESL-1000BRecalls & reported problems

Deroyal Industries, Inc. · SafeLiner Suction Canister 1000cc · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00749756377041), received between Aug 7, 2024 and Dec 6, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Suction Problem” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 9, 2023Open, Classified

    The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.

    Z-1283-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Suction Problem3
  • Collapse1
  • Decrease in Suction1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 6, 2024Suction Problem; Suction Problem

    A USER FACILITY REPORTED, "THE DEROYAL SUCTION LINERS WERE INITIALLY UNABLE TO SUPPORT INTERMITTENT SUCTION DUE TO AN IMPROPER SEAL WITH THE SUCTION REGULATORS IN THE DEROYAL CANISTERS, THOUGH CONTINUOUS SUCTION WAS FUNCTIONAL. BIOMED FIRST ASSESSED, SUSPECTING A WALL LEAK, BUT PLANT OPERATIONS CONFIRMED NO WALL ISSUE. THEY FOUND THAT PROPERLY SEALING THE LINER TOP TO THE CANISTER ALLOWED INTERMITTENT SUCTION TO WORK. MULTIPLE TROUBLESHOOTING ATTEMPTS INCLUDED CHANGING OUT SUCTION MACHINES, PLASTIC CANISTERS, LINERS, TOPS, AND TUBING."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20866926 ↗
  • MalfunctionOct 29, 2024Collapse; Collapse

    A USER FACILITY REPORTED, "CANISTER ARE USED BY ANESTHESIA FOR SUCTION FOR [PATIOENT CARE. CANISTERS ARE COLLAPSING IN THE PATIENT ROOM. DELAYS CAUSED BY NEED TO CHANGE THE CANISTER DURING PROCEDURES."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20559207 ↗
  • MalfunctionAug 7, 2024Suction Problem; Suction Problem

    THEY ARE HAVING ISSUES WITH GETTING ADEQUATE SUCTION WITH SOME OF THEIR LINERS/CANISTERS. THEY HAD SOME OLD REGULATORS AND IDENTIFIED THAT AS PART OF THE PROBLEM. BUT STILL HAVING ISSUES W/ SOME OF THE NEW REGULATORS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19928891 ↗
  • MalfunctionAug 7, 2024Decrease in Suction; Suction Problem; Decrease in Suction; Suction Problem

    NURSE COULDN'T GET GOOD STRONG SUCTION WITH LINER ONLY VERY LOW PRESSURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19928709 ↗