PNEUMOSTAT 16100Recalls & reported problems

Atrium Medical Corporation · PNEUMOSTAT CHEST DRAIN VALVE · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
12
Deaths reported
3
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00650862161007), received between May 18, 2017 and Dec 10, 2020 — classified as 3 deaths, 2 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K984496); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Mar 1, 2023Open, Classified

    The Instructions for Use (IFU) for the Pneumostat Chest Drain Valve and the Express Mini 500 Dry Seal Chest Drains do not provide sufficient precaution against draining the device for continued use. Additionally, the IFUs do not provide sufficient precaution or warning against the use of the device outside the healthcare setting.

    Z-1302-2023 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Fluid/Blood Leak3
  • Leak/Splash3
  • Crack2
  • Break1
  • Detachment of Device or Device Component1
  • Gas/Air Leak1
  • Insufficient Device Problem Information1

Patient problems

  • No Known Impact Or Consequence To Patient4
  • Death3
  • Pneumothorax3
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 10, 2020Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT A PATIENT WENT HOME WITH A PNEUMOSTAT AND CAME BACK TO HOSPITAL WITH A PNEUMOTHORAX.

    Patient problems: Pneumothorax

    MAUDE report 10988973 ↗
  • DeathOct 3, 2020Adverse Event Without Identified Device or Use Problem

    REPORT RECEIVED FROM THERAPEUTIC GOODS ADMINISTRATION (TGA): PRETERM BABY DELIVERED AFTER HOURS AT 0432AM IN OPERATING THEATRES BY CAESAREAN SECTION AT A (B)(6) HOSPITAL. THE BABY REQUIRED PROLONGED RESUSCITATION AND WAS INTUBATED AND VENTILATED WITH DIFFICULTY. A LEFT TENSION PNEUMOTHORAX DEVELOPED WHILE THE BABY WAS STILL IN OPERATING THEATRES, AND AFTER INITIAL NEEDLE DRAINAGE, AN INTERCOSTAL CATHETER WAS INSERTED AND CONNECTED TO A PNEUMOSTAT CHEST DRAIN VALVE. THE BABY WAS MOVED TO THE NEONATAL INTENSIVE CARE UNIT (NICU) AT 90 MINUTES OF AGE. AT FIRST X-RAY IN THE NICU (0626AM), THERE WAS STILL LEFT PNEUMOTHORAX PRESENT ALTHOUGH MUCH SMALLER, AND A RIGHT TENSION PNEUMOTHORAX HAD DEVELOPED, AS WELL AS SURGICAL EMPHYSEMA. A RIGHT INTERCOSTAL CATHETER WAS INSERTED AND ALSO CONNECTED TO A PNEUMOSTAT CHEST DRAIN VALVE. ON THE NEXT X-RAY AT 0655AM, THE RIGHT PNEUMOTHORAX WAS RELIEVED WITH RIGHT LUNG RE EXPANSION, HOWEVER THE LEFT PNEUMOTHORAX WAS NOW INCREASED IN SIZE. ON CXR AT 0748, THERE WAS AGAIN A RIGHT TENSION PNEUMOTHORAX, AS WELL AS ONGOING LEFT PNEUMOTHORAX. ON X-RAY AT 0757, THERE WAS BILATERAL TENSION PNEUMOTHORAX. THE PATIENT PASSED AWAY THE SAME MORNING.

    Patient problems: Death

    MAUDE report 10624472 ↗
  • DeathOct 3, 2020Adverse Event Without Identified Device or Use Problem

    N/A.

    Patient problems: Death

    MAUDE report 10624471 ↗
  • DeathOct 3, 2020Labelling, Instructions for Use or Training Problem

    N/A.

    Patient problems: Death

    MAUDE report 10624469 ↗
  • MalfunctionAug 6, 2020Crack; Fluid/Blood Leak; Material Fragmentation

    IT WAS REPORTED THAT THE PLASTIC OF THE SECRETION CHAMBER ON THE LUER PORT WAS LEAKING, SHOWING CRACKS AND BREAKING OFF WITHOUT THE USE OF FORCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10377238 ↗
  • MalfunctionMar 2, 2020Leak/Splash

    REPORT RECEIVED STATED THAT THE NURSE FOUND THAT A PNEUMOSTAT WAS LEAKING FROM THE WHITE CHAMBER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9777727 ↗
  • InjurySep 26, 2019Gas/Air Leak

    N/A.

    Patient problems: Pneumothorax

    MAUDE report 9124680 ↗
  • MalfunctionJul 24, 2019Break; Detachment of Device or Device Component

    N/A.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8820466 ↗
  • MalfunctionJul 3, 2019Fluid/Blood Leak

    N/A.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8757535 ↗
  • MalfunctionJul 2, 2019Fluid/Blood Leak

    N/A.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8753900 ↗
  • MalfunctionSep 10, 2018Crack; Leak/Splash

    IT WAS REPORTED THE PATIENT HAD INTERCOSTAL TUBE DAMAGE FROM PNEUMOTHORAX. THE PNEUMOSTAT WAS CONNECTED TO THE EXISTING IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). POST ATTACHMENT AN EXTERNAL AIR LEAK WAS NOTICED AND THE PNEUMOSTAT WAS DISCONNECTED. ON EXAMINATION A SMALL CRACK ON THE INLET PART WAS NOTICED.

    Patient problems: Pneumothorax

    MAUDE report 7862467 ↗
  • MalfunctionMay 18, 2017Leak/Splash

    THE PNEUMOSTAT DRAIN LEAKED THROUGH THE AIR LEAK MONITOR WHEN IN A POSITION OTHER THAN UPRIGHT. PATIENT DRAINING MORE THAN 30 ML PER DAY. PATIENT WAS SWITCHED TO AN EXPRESS MINI WITHOUT ANY ISSUE AND PATIENT DISCHARGED THE NEXT DAY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6575471 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K984496. One submission often covers many models, so check the product and lot details in each record.

  • Z-2592-2021 · Aug 26, 2021 · Open, Classified
    Atrium Pneumostat Chest Drain Valve, Part Number 16100
  • Z-1208-2017 · Dec 23, 2016 · Terminated
    Maquet Express Dry Seal Suction Chest Drain, Single Collection w/AC Reference: 4000-100N Intended Use: To evacuate air and/or fluid from the chest cavity or mediastinum. " To help

Other BOTTLE, COLLECTION, VACUUM devices

All BOTTLE, COLLECTION, VACUUM devices →