Skip to content

iCast Covered Stent System, 6mmX22mmX120cm 42622Recalls & reported problems

Atrium Medical Corporation · iCast Covered Stent System, 6mmX22mmX120cm · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00650862426229), received between Mar 21, 2025 and May 4, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (P120003); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation, Positioning or Separation Problem1
  • Device Fell1
  • Failure to Advance1
  • Off-Label Use1
  • Patient-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Calcium Deposits/Calcification1
  • Foreign Body In Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 4, 2026Patient-Device Incompatibility

    DURING THE PROCEDURE, THE PHYSICIAN ATTEMPTED TO CROSS A STENOTIC AREA USING A 6-45 ANSEL SHEATH; HOWEVER, THIS WAS UNSUCCESSFUL VIA THE INITIAL APPROACH. ALTERNATE ACCESS WAS OBTAINED VIA THE LEFT EXTERNAL APPROACH, AND THE STENOSIS IN THE PROXIMAL LEFT COMMON ILIAC ARTERY WAS SUCCESSFULLY CROSSED USING AN AMPLATZ WIRE. AN ICAST STENT WAS ADVANCED BUT BECAME DISLODGED FROM THE BALLOON PRIOR TO DEPLOYMENT. THE DISLODGED STENT WAS SUCCESSFULLY SNARED AND RETRIEVED. SHOCKWAVE THERAPY WAS THEN PERFORMED TO TREAT THE STENOSIS, AND NO STENT WAS IMPLANTED. THE SHEATH AND DISLODGED STENT WERE REMOVED FROM THE PATIENT. NO HARM WAS REPORTED.

    Patient problems: Calcium Deposits/Calcification

    MAUDE report 25067582 ↗
  • MalfunctionNov 7, 2025Off-Label Use; Device Fell; Off-Label Use; Device Fell

    A PATIENT UNDERWENT AN UNSPECIFIED PROCEDURE USING THE ICAST COVERED STENT SYSTEM, DURING WHICH THREE ICAST STENTS FELL OFF THE BALLOON CATHETER. THE PHYSICIAN PLACED AN AORTIC GRAFT AND LASER FENESTRATION TO INSERT THE STENTS INTO THE RENAL ARTERIES, BUT COULDN'T GET THE SHEATH IN THE VESSELS SO THE PHYSICIAN PLACED THE ICAST STENTS VIA BARE BACK. ONCE THE ICAST STENTS WERE OUT OF THE SHEATH THE STENTS BECAME CAUGHT ON EITHER A STENT IN THE AORTIC GRAFT OR THE BURNT HOLES. PRIOR TO THE ICAST STENTS, THE PHYSICIAN TRIED TO USE A COMPETITOR'S STENT WITH SIMILAR ISSUES. A 6MM X 22MM ICAST STENT WAS SUCCESSFULLY PLACED IN THE RIGHT RENAL ARTERY, WHILE A VBX STENT WAS USED IN THE LEFT RENAL ARTERY AFTER ATTEMPTS TO PLACE TWO 7MM X 22MM ICAST STENTS WERE UNSUCCESSFUL. THERE WAS NO ADVERSE EVENT OR HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23497525 ↗
  • InjuryApr 28, 2025Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem

    THE ICAST STENT FELL OFF THE BALLOON WHILE TRYING TO POSITION IT IN THE RIGHT RENAL AND SMA. THE STENTS WERE RECOVERED AND TAKEN OUT OF THE BODY BY SNARING THEM. NO ADVERSE OUTCOMES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21923458 ↗
  • MalfunctionMar 21, 2025Failure to Advance; Failure to Advance

    N/A.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21660202 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P120003. One submission often covers many models, so check the product and lot details in each record.

  • Z-0291-2024 · Sep 21, 2023 · Open, Classified
    Atrium Advanta V12 Covered Stent System, 10mm x 59mm x 80cm (OUS)-Indicated for restoring and improving the patency of the iliac andrenal arteries. Renal approval includes 5mm, 6m

Other Iliac Covered Stent, Arterial devices

All Iliac Covered Stent, Arterial devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page