iCast Covered Stent System, 6mmX22mmX120cm 42622Recalls & reported problems
Atrium Medical Corporation · iCast Covered Stent System, 6mmX22mmX120cm · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00650862426229), received between Mar 21, 2025 and May 4, 2026 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (P120003); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation, Positioning or Separation Problem1
- Device Fell1
- Failure to Advance1
- Off-Label Use1
- Patient-Device Incompatibility1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Calcium Deposits/Calcification1
- Foreign Body In Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryMay 4, 2026Patient-Device Incompatibility
DURING THE PROCEDURE, THE PHYSICIAN ATTEMPTED TO CROSS A STENOTIC AREA USING A 6-45 ANSEL SHEATH; HOWEVER, THIS WAS UNSUCCESSFUL VIA THE INITIAL APPROACH. ALTERNATE ACCESS WAS OBTAINED VIA THE LEFT EXTERNAL APPROACH, AND THE STENOSIS IN THE PROXIMAL LEFT COMMON ILIAC ARTERY WAS SUCCESSFULLY CROSSED USING AN AMPLATZ WIRE. AN ICAST STENT WAS ADVANCED BUT BECAME DISLODGED FROM THE BALLOON PRIOR TO DEPLOYMENT. THE DISLODGED STENT WAS SUCCESSFULLY SNARED AND RETRIEVED. SHOCKWAVE THERAPY WAS THEN PERFORMED TO TREAT THE STENOSIS, AND NO STENT WAS IMPLANTED. THE SHEATH AND DISLODGED STENT WERE REMOVED FROM THE PATIENT. NO HARM WAS REPORTED.
Patient problems: Calcium Deposits/Calcification
MAUDE report 25067582 ↗MalfunctionNov 7, 2025Off-Label Use; Device Fell; Off-Label Use; Device Fell
A PATIENT UNDERWENT AN UNSPECIFIED PROCEDURE USING THE ICAST COVERED STENT SYSTEM, DURING WHICH THREE ICAST STENTS FELL OFF THE BALLOON CATHETER. THE PHYSICIAN PLACED AN AORTIC GRAFT AND LASER FENESTRATION TO INSERT THE STENTS INTO THE RENAL ARTERIES, BUT COULDN'T GET THE SHEATH IN THE VESSELS SO THE PHYSICIAN PLACED THE ICAST STENTS VIA BARE BACK. ONCE THE ICAST STENTS WERE OUT OF THE SHEATH THE STENTS BECAME CAUGHT ON EITHER A STENT IN THE AORTIC GRAFT OR THE BURNT HOLES. PRIOR TO THE ICAST STENTS, THE PHYSICIAN TRIED TO USE A COMPETITOR'S STENT WITH SIMILAR ISSUES. A 6MM X 22MM ICAST STENT WAS SUCCESSFULLY PLACED IN THE RIGHT RENAL ARTERY, WHILE A VBX STENT WAS USED IN THE LEFT RENAL ARTERY AFTER ATTEMPTS TO PLACE TWO 7MM X 22MM ICAST STENTS WERE UNSUCCESSFUL. THERE WAS NO ADVERSE EVENT OR HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23497525 ↗InjuryApr 28, 2025Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
THE ICAST STENT FELL OFF THE BALLOON WHILE TRYING TO POSITION IT IN THE RIGHT RENAL AND SMA. THE STENTS WERE RECOVERED AND TAKEN OUT OF THE BODY BY SNARING THEM. NO ADVERSE OUTCOMES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21923458 ↗MalfunctionMar 21, 2025Failure to Advance; Failure to Advance
N/A.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 21660202 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120003. One submission often covers many models, so check the product and lot details in each record.
- Z-0291-2024 · Sep 21, 2023 · Open, ClassifiedAtrium Advanta V12 Covered Stent System, 10mm x 59mm x 80cm (OUS)-Indicated for restoring and improving the patency of the iliac andrenal arteries. Renal approval includes 5mm, 6m
Other Iliac Covered Stent, Arterial devices
- GORE VIABAHN VBX Balloon Expandable Endo BXB073902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths19 reports
- iCast Covered Stent System, 6mmX16mmX120cm 42616Atrium Medical Corporation · Iliac Covered Stent, Arterial1 recall15 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB057902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB067902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB075902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall11 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB095902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB077902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB085902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page