iCast Covered Stent System, 7mmX22mmX80cm 48722Recalls & reported problems
Atrium Medical Corporation · iCast Covered Stent System, 7mmX22mmX80cm · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00650862487220), received between Aug 21, 2025 and Sep 28, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Accessory Incompatible” (1 report).
1 other recall cover devices cleared under the same FDA submission (P120003); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Accessory Incompatible1
- Activation, Positioning or Separation Problem1
- Device Dislodged or Dislocated1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Device Dislodged or Dislocated
During placement of the iCast stent in the SMA, after successful cannulation and guidewire exchange to a Rosen wire, the physician advanced the iCast through a 7 Fr sheath. While attempting deployment, the stent and balloon detached from each other. The detached iCast was successfully retrieved using a Goose Neck snare. No patient injury was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50913723 ↗MalfunctionJun 8, 2026Accessory Incompatible
DURING A FEVAR PROCEDURE WITH A TERUMO TREO FENESTRATED DEVICE, FOUR VISCERAL VESSELS (BILATERAL RENAL ARTERIES, SMA, AND CELIAC) WERE STENTED USING ICAST STENTS. WHILE ATTEMPTING TO ADVANCE AN ICAST STENT INTO THE CELIAC ARTERY VIA A 6.5 FR BAYLISS STEERABLE SHEATH, THE STENT FAILED TO CROSS THE FENESTRATION. THE DEVICE WAS WITHDRAWN TO ALLOW PRE-DILATION; HOWEVER, DURING REMOVAL, THE STENT CAUGHT ON THE SHEATH¿S HEMOSTATIC VALVE AND DISLODGED. THE STENT WAS RETRIEVED INTACT. A SECOND ICAST STENT WAS SUCCESSFULLY DEPLOYED. NO HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25429594 ↗MalfunctionAug 21, 2025Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED THAT DURING AN UNSPECIFIED PROCEDURE AND ICAST STENT FELL OFF THE BALLOON AND WAS NOT DEPLOYED IN THE PATIENT. NO HARM, ADDITIONAL PROCEDURES OR EXTENDED SURGERY WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22863750 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120003. One submission often covers many models, so check the product and lot details in each record.
- Z-0291-2024 · Sep 21, 2023 · Open, ClassifiedAtrium Advanta V12 Covered Stent System, 10mm x 59mm x 80cm (OUS)-Indicated for restoring and improving the patency of the iliac andrenal arteries. Renal approval includes 5mm, 6m
Other Iliac Covered Stent, Arterial devices
- GORE VIABAHN VBX Balloon Expandable Endo BXB073902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths19 reports
- iCast Covered Stent System, 6mmX16mmX120cm 42616Atrium Medical Corporation · Iliac Covered Stent, Arterial1 recall15 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB057902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB067902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall14 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB075902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall11 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB095902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB077902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall8 reports
- GORE VIABAHN VBX Balloon Expandable Endo BXB085902AW. L. Gore & Associates, Inc. · Iliac covered stent, arterial1 recall2 deaths8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page