Skip to content

TiMesh® 015-042Recalls & reported problems

Medtronic Ps Medical, Inc. · PLATE 015-042 STRAIGHT 1H X 1H 8MM · Class 2 · Implant

All models of TiMesh (15)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00673978002668), received between Jul 28, 2023 and Oct 9, 2024 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 5 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Fracture7

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Unspecified Tissue Injury1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 9, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING A PLATE FOR CRANIAL REPAIR. IT WAS REPORTED THAT THE SURGEON USED THE CRANIOMAXILLOFACIAL FIXATION SYSTEM TO INSTALL A 2-HOLE 8MM STANDARD STRAIGHT PLATE. THE FIRST SCREW WAS SCREWED IN NORMALLY. AFTER THE SECOND SCREW WAS SCREWED IN, IT WAS FOUND THAT THE CONNECTION PLATE WAS BROKEN. THE BROKEN PART WAS IMMEDIATELY REMOVED, AND THE CIRCULATING NURSE REPLACED THE CRANIOMAXILLOFACIAL FIXATION SYSTEM, WHICH THE SURGEON USED NORMALLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20407483 ↗
  • MalfunctionJan 4, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING CRANIAL REPAIR PRODUCTS. IT WAS REPORTED THAT WHEN USING THE HAND TOOLS TO TIGHTEN THE SCREW DURING SURGERY, DUE TO THE LACK OF SELF-TAPPING PROPERTIES OF THE SCREW, THE DOCTOR NEEDED TO USE MORE FORCE TO TIGHTEN THE SCREW CAUSING THE SNOWFLAKE PIECE, CONNECTING PIECE, AND SCREW TO DEFORM AND MAKING IT IMPOSSIBLE TO CONTINUE USING IT. IT WAS NOTED THAT THE HOSPITAL SUPPLY ROOM COLLECTED DAMAGED PRODUCTS FROM VARIOUS MANUFACTURERS ON A DAILY BASIS AND PROVIDED THEM TO THE DISTRIBUTOR AT ONE TIME DURING THE INVENTORY PERIOD AT THE END OF THE YEAR. IT WAS ASKED FOR THE DOCTORS TO USE MEDTRONIC PRODUCTS OF THE SAME MODEL INSTEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18446863 ↗
  • MalfunctionJan 4, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING CRANIAL REPAIR PRODUCTS. IT WAS REPORTED THAT WHEN USING THE HAND TOOLS TO TIGHTEN THE SCREW DURING SURGERY, DUE TO THE LACK OF SELF-TAPPING PROPERTIES OF THE SCREW, THE DOCTOR NEEDED TO USE MORE FORCE TO TIGHTEN THE SCREW CAUSING THE SNOWFLAKE PIECE, CONNECTING PIECE, AND SCREW TO DEFORM AND MAKING IT IMPOSSIBLE TO CONTINUE USING IT. IT WAS NOTED THAT THE HOSPITAL SUPPLY ROOM COLLECTED DAMAGED PRODUCTS FROM VARIOUS MANUFACTURERS ON A DAILY BASIS AND PROVIDED THEM TO THE DISTRIBUTOR AT ONE TIME DURING THE INVENTORY PERIOD AT THE END OF THE YEAR. IT WAS ASKED FOR THE DOCTORS TO USE MEDTRONIC PRODUCTS OF THE SAME MODEL INSTEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18446804 ↗
  • MalfunctionJan 4, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING CRANIAL REPAIR PRODUCTS. IT WAS REPORTED THAT WHEN USING THE HAND TOOLS TO TIGHTEN THE SCREW DURING SURGERY, DUE TO THE LACK OF SELF-TAPPING PROPERTIES OF THE SCREW, THE DOCTOR NEEDED TO USE MORE FORCE TO TIGHTEN THE SCREW CAUSING THE SNOWFLAKE PIECE, CONNECTING PIECE, AND SCREW TO DEFORM AND MAKING IT IMPOSSIBLE TO CONTINUE USING IT. IT WAS NOTED THAT THE HOSPITAL SUPPLY ROOM COLLECTED DAMAGED PRODUCTS FROM VARIOUS MANUFACTURERS ON A DAILY BASIS AND PROVIDED THEM TO THE DISTRIBUTOR AT ONE TIME DURING THE INVENTORY PERIOD AT THE END OF THE YEAR. IT WAS ASKED FOR THE DOCTORS TO USE MEDTRONIC PRODUCTS OF THE SAME MODEL INSTEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18446763 ↗
  • MalfunctionJan 3, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING CRANIAL REPAIR PRODUCTS. IT WAS REPORTED THAT WHEN USING THE HAND TOOLS TO TIGHTEN THE SCREW DURING SURGERY, DUE TO THE LACK OF SELF-TAPPING PROPERTIES OF THE SCREW, THE DOCTOR NEEDED TO USE MORE FORCE TO TIGHTEN THE SCREW CAUSING THE SNOWFLAKE PIECE, CONNECTING PIECE, AND SCREW TO DEFORM AND MAKING IT IMPOSSIBLE TO CONTINUE USING IT. IT WAS NOTED THAT THE HOSPITAL SUPPLY ROOM COLLECTED DAMAGED PRODUCTS FROM VARIOUS MANUFACTURERS ON A DAILY BASIS AND PROVIDED THEM TO THE DISTRIBUTOR AT ONE TIME DURING THE INVENTORY PERIOD AT THE END OF THE YEAR. IT WAS ASKED FOR THE DOCTORS TO USE MEDTRONIC PRODUCTS OF THE SAME MODEL INSTEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18439020 ↗
  • MalfunctionJan 3, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING CRANIAL REPAIR PRODUCTS. IT WAS REPORTED THAT WHEN USING THE HAND TOOLS TO TIGHTEN THE SCREW DURING SURGERY, DUE TO THE LACK OF SELF-TAPPING PROPERTIES OF THE SCREW, THE DOCTOR NEEDED TO USE MORE FORCE TO TIGHTEN THE SCREW CAUSING THE SNOWFLAKE PIECE, CONNECTING PIECE, AND SCREW TO DEFORM AND MAKING IT IMPOSSIBLE TO CONTINUE USING IT. IT WAS NOTED THAT THE HOSPITAL SUPPLY ROOM COLLECTED DAMAGED PRODUCTS FROM VARIOUS MANUFACTURERS ON A DAILY BASIS AND PROVIDED THEM TO THE DISTRIBUTOR AT ONE TIME DURING THE INVENTORY PERIOD AT THE END OF THE YEAR. IT WAS ASKED FOR THE DOCTORS TO USE MEDTRONIC PRODUCTS OF THE SAME MODEL INSTEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18438921 ↗
  • InjuryJul 28, 2023Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING A DURAREPAIR. IT WAS REPORTED ON (B)(6) 2023, 17:45 ACCORDING TO THE NEEDS OF THE SURGERY, THE SURGEON USED MEDTRONIC CONNECTING PLATES AND NAILS. AT 17:50, WHEN THE SURGEON INSTALLED THE NAIL, THE CONNECTING PLATE BROKE. AT 18:00, A NEW MEDTRONIC CONNECTING PLATE WAS REPLACED, AND THE SURGERY PROCEEDED NORMALLY.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17413273 ↗

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →