HANCOCK® HC15020Recalls & reported problems
Medtronic, Inc. · CONDUIT HC15020 HKI MO 10L · Class 3 · Implant
All models of HANCOCK (22)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00681490908795), received between Nov 30, 2018 and Feb 19, 2021 — classified as 9 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information8
- Calcified1
- Degraded1
- Device Appears to Trigger Rejection1
- Patient Device Interaction Problem1
Patient problems
- Pulmonary Valve Stenosis5
- No Information4
- Calcium Deposits/Calcification2
- Pulmonary Insufficiency2
- Endocarditis1
- High Blood Pressure/ Hypertension1
- Insufficient Information1
- Pulmonary Regurgitation1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryFeb 19, 2021Degraded; Degraded
MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 6 MONTHS POST IMPLANT OF THIS 20MM PULMONARY VALVED CONDUIT IN A (B)(6)-YEAR-OLD PEDIATRIC PATIENT, IT WAS EXPLANTED AND REPLACED WITH A 25MM CONDUIT OF THE SAME MODEL. THE CONDUIT WAS REPLACED DUE TO LEAFLET DEGENERATION RESULTING IN WIDE-OPEN PULMONARY INSUFFICIENCY. THERE WAS ALSO "SOME STENOSIS" REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS RELATED TO THIS PULMONARY VALVED CONDUIT WERE REPORTED.
Patient problems: Pulmonary Valve Insufficiency/ Regurgitation; Pulmonary Valve Stenosis
MAUDE report 11355428 ↗InjuryFeb 2, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 6 YEARS AND 6 MONTHS POST IMPLANT OF THIS 20MM PULMONARY VALVED CONDUIT IN AN (B)(6)-YEAR-OLD PEDIATRIC PATIENT, IT WAS EXPLANTED AND REPLACED WITH A 25MM CONDUIT OF THE SAME MODEL. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 11266502 ↗InjuryOct 12, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS POST IMPLANT OF THIS 20 MM PULMONARY VALVED CONDUIT IN A (B)(6)-MONTH-OLD PEDIATRIC PATIENT, A TRANSCATHETER PULMONARY BIOPROSTHETIC VALVE (TPBV) WAS IMPLANTED VALVE-IN-VALVE. THE VALVE WAS REPLACED DUE TO SEVERE REGURGITATION. OF NOTE, IT WAS REPORTED THAT WHEN THE CONDUIT WAS ORIGINALLY IMPLANTED, THE SURGEON REMOVED THE STEEL SUPPORTING RING FROM THE CONDUIT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Pulmonary Regurgitation
MAUDE report 10662477 ↗InjuryAug 30, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS POST IMPLANT OF THIS 20MM PULMONARY VALVED CONDUIT IN A (B)(6) PEDIATRIC PATIENT, IT WAS EXPLANTED AND REPLACED WITH A 25MM CONDUIT OF THE SAME MODEL. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Information
MAUDE report 10469906 ↗InjuryApr 15, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 2 YEARS AND 6 MONTHS POST IMPLANT OF THIS PULMONARY VALVED CONDUIT IMPLANTED IN AN APPROXIMATELY (B)(6) PATIENT, A TRANSCATHETER PULMONARY VALVE (TPV) WAS IMPLANTED VALVE-IN-VALVE DUE TO STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Pulmonary Valve Stenosis
MAUDE report 9961073 ↗InjuryOct 18, 2019Insufficient Device Problem Information; Patient Device Interaction Problem; Insufficient Device Problem Information; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 14 YEARS AND 9 MONTHS POST IMPLANT OF THIS 20MM PULMONARY VALVED CONDUIT IN A (B)(6) PEDIATRIC PATIENT, IT WAS REPLACED VALVE-IN-VALVE WITH A TRANSCATHETER BIOPROSTHETIC PULMONARY VALVE. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Calcium Deposits/Calcification; Endocarditis; No Information; Pulmonary Insufficiency; Pulmonary Valve Stenosis
MAUDE report 9207685 ↗InjuryFeb 26, 2019Calcified; Device Appears to Trigger Rejection; Insufficient Device Problem Information; Calcified; Device Appears to Trigger Rejection; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS POST IMPLANT OF THIS 20MM PULMONARY VALVED CONDUIT IMPLANTED IN A SIX YEAR PATIENT, IT WAS EXPLANTED AND REPLACED WITH A 25MM CONDUIT OF THE SAME MODEL. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Calcium Deposits/Calcification; No Information; Pulmonary Insufficiency; Pulmonary Valve Stenosis
MAUDE report 8370323 ↗InjuryFeb 12, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 5 MONTHS POST IMPLANT OF THIS PULMONARY VALVED CONDUIT IN A (B)(6) YEAR OLD PEDIATRIC PATIENT, A TRANSCATHETER PULMONARY BIOPROSTHETIC VALVE (TPBV) WAS IMPLANTED VALVE-IN-VALVE DUE TO SEVERE CONDUIT STENOSIS AND SUPRA-SYSTEMIC RIGHT VENTRICLE (RV) PRESSURE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: High Blood Pressure/ Hypertension; Pulmonary Valve Stenosis
MAUDE report 8332704 ↗InjuryNov 30, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS 4 MONTHS POST IMPLANT OF THIS 20MM BIOPROSTHETIC VALVED CONDUIT IN THIS PEDIATRIC PATIENT, IT WAS REPLACED VALVE-IN-VALVE WITH A TRANSCATHETER PULMONARY VALVE (TPV) FOR UNKNOWN REASONS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Information
MAUDE report 8121340 ↗
Other heart-valve, non-allograft tissue devices
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue2 deaths404 reports
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue385 reports
- PERCEVAL PVS23Corcym Canada Corp · heart-valve, non-allograft tissue26 deaths337 reports
- MOSAIC® 305C23Medtronic, Inc. · heart-valve, non-allograft tissue327 reports
- MOSAIC® 305C25Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths324 reports
- FREESTYLE® FR995-27Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths318 reports
- FREESTYLE® FR995-29Medtronic, Inc. · heart-valve, non-allograft tissue288 reports
- FREESTYLE® FR99525Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths280 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page