Skip to content

VersaCross RF Wire VXW-35-230-PRecalls & reported problems

Baylis Médicale Cie Inc · 1 x VersaCross RF Wire; 1 x Connector Cable (Single Use) · Class 2

All models of VersaCross RF Wire (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00685447003311), received between Feb 4, 2021 and Jan 28, 2025 — classified as 10 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem10
  • Deformation Due to Compressive Stress1
  • Device-Device Incompatibility1
  • Difficult to Remove1
  • Failure to Advance1
  • Insufficient Device Problem Information1
  • Material Deformation1
  • Physical Resistance/Sticking1

Patient problems

  • Pericardial Effusion7
  • No Clinical Signs, Symptoms or Conditions5
  • Hemothorax1
  • Perforation1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 28, 2025Failure to Advance; Failure to Advance

    IT WAS REPORTED THAT BMC RF WIRE WAS SELECTED FOR PULSE FIELD ABLATION. DURING THE PROCEDURE, RADIO FREQUENCY WAS APPLIED ONCE, AND THEN, WHEN PERFORMING AGAIN THE TRANSEPTAL PUNCTURE, SINCE THE PHYSICIAN WAS NOT SURE ABOUT THE POSITION, THE RADIO FREQUENCY DID NOT WORK. THUS, THE DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS WERE REPORTED. THE DEVICE IS EXPECTED TO BE RETURN FOR ANALYSIS. NO OTHER ISSUES WERE REPORTED. THE DEVICE HAS RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21256482 ↗
  • MalfunctionMay 1, 2024Difficult to Remove; Material Deformation; Physical Resistance/Sticking; Difficult to Remove; Material Deformation; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING A TRANSSEPTAL PUNCTURE PROCEDURE, A VERSACROSS RF WIRE WAS SELECTED FOR USE. THE PHYSICIAN TRIED TO INSERT THE VERSACROSS RF WIRE DIRECTLY THROUGH THE NON-BOSTON SCIENTIFIC SYRINGE NEEDLE CAUSING BENDING OF THE WIRE TIP. THUS, THE DEVICE WAS REPLACED, AND THE ISSUE WAS RESOLVED. PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS WERE REPORTED. THE DEVICE IS STILL ON HOSPITAL FACILITIES AND THE PRODUCT WILL NOT RETURN FOR ANALYSIS. IT HAS BEEN FURTHER MENTIONED THAT WHEN THE VERSACROSS RF WIRE WAS INSERTED DIRECTLY INTO THE NON-BOSTON SCIENTIFIC NEEDLE UTILIZED FOR FEMORAL ACCESS, IT BECAME STUCK JUST AFTER THE TIP OF THE NON-BOSTON SCIENTIFIC NEEDLE. THE PHYSICIAN ATTEMPTED TO PULL IT OUT, BUT WAS UNABLE TO DO SO AND THEREFORE REMOVED THE VERSACROSS RF WIRE AND NEEDLE ALTOGETHER. OUTSIDE THE PATIENT'S BODY, THE PHYSICIAN WAS ABLE TO RETRACT THE VERSACROSS RF WIRE FROM THE NEEDLE, AND THAT'S WHEN THE KINK WAS NOTED. NO OTHER ISSUES WERE NOTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19226449 ↗
  • InjuryJan 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A PERICARDIAL EFFUSION WAS NOTED THAT LEAD TO A PROCEDURAL ABORTION WHERE VERSACROSS LARGE ACCESS TRANSSEPTAL DILATOR, VERSACROSS RF WIRE AND AGILIS SHEATH (NON-BAYLIS DEVICE) WERE USED. TRANSSEPTAL WAS UNSUCCESSFUL USING THE VERSCROSS LARGE ACCESS TRANSSEPTAL DILATOR AND VERSACROSS RF WIRE. BOTH TRANSESOPHAGEAL ECHOCARDIOGRAPHY (TEE) AND ICE IMAGING WERE USED BUT THE PHYSICIAN WAS UNABLE TO VIEW THE DILATOR TENT THE SEPTUM.. THE PHYSICIAN DECIDED TO USE THE AGILIS SHEATH WITH THE VERSACROSS RF WIRE. THE PHYSICIAN HAD DIFFICULTY VIEWING THE LOCATION OF THE CATHETER ON THE BICAVAL VIEW. TRANSSEPTAL WAS ATTEMPTED THREE TIMES HOWEVER, THE PHYSICIAN WAS UNABLE TO ADVANCE THE DEVICES INTO THE SEPTUM. AN EFFUSION WAS NOTED VIA TRANSESOPHAGEAL ECHOCARDIOGRAPHY (TEE) AND ON FLUOROSCOPY. THE PATIENT'S BLOOD PRESSURE DROPPED, AND A PERICARDIAL TAP WAS PERFORMED STABILIZING THE PATIENTS' BLOOD PRESSURE. BLOOD WAS RE-INFUSED INTO THE PATIENT. ONCE THE FLUID COLLECTION STOPPED, THE PHYSICIAN DECIDED TO REMOVE THE DRAIN. PATIENT WAS EXTUBATED, DRESSING WAS APPLIED TO SUBXYPHOID AREA AND THE PROCEDURE WAS ABORTED. THE PATIENT WAS SENT TO RECOVERY AREA. AS A BAYLIS MEDICAL DEVICE WAS REPORTED TO B

    Patient problems: Pericardial Effusion

    MAUDE report 16154491 ↗
  • InjuryJul 8, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A CASE OF AN AORTIC PERFORATION LEADING TO A PROCEDURAL ABORTION DURING A LEFT ATRIAL APPENDAGE CLOSURE (LAAC) PROCEDURE WAS REPORTED WHERE THE BAYLIS MEDICAL VERSACROSS RF WIRE (RF WIRE) WAS USED FOR THE INITIAL TRANSSEPTAL PUNCTURE OF THE PROCEDURE. THE PATIENT'S ATRIAL SEPTUM WAS NOTED TO BE HYPER ELASTIC ON TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE). THE RF WIRE WAS POSITIONED ON THE SEPTUM. THE RF WIRE CROSSED THE SEPTUM AND WAS NOTED TO BE ADVANCED IN THE AORTA ON TEE. THE RF WIRE WAS RETRACTED INTO THE RIGHT ATRIUM. A PIGTAIL CATHETER WAS INTRODUCED INTO THE AORTA. CONTRAST WAS ADMINISTERED, AND A PUNCTURE TO THE AORTA WAS CONFIRMED. THE LAAC PROCEDURE WAS ABORTED. THE PATIENT WAS STABLE AND UNDERWENT COMPUTED TOMOGRAPHY IMAGING. MINIMAL PERICARDIAL EFFUSION WAS NOTED. THE AORTIC PERFORATION WAS REPORTED TO HEAL AS THE PATIENT CONTINUED TO REMAIN IN STABLE CONDITION 12 HOURS LATER. AS A BAYLIS MEDICAL DEVICE WAS REPORTED TO BE AMONG THE DEVICES USED IN THE PROCEDURE, BAYLIS MEDICAL HAS DECIDED TO SUBMIT THIS REPORT.

    Patient problems: Perforation

    MAUDE report 14957870 ↗
  • InjuryJun 3, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A CASE OF PERICARDIAL EFFUSION REQUIRING INTERVENTION WAS REPORTED DURING A LEFT ATRIAL APPENDAGE CLOSURE (LAAC) PROCEDURE, WHERE THE BAYLIS MEDICAL, VERSACROSS STEERABLE SHEATH (VST), VERSACROSS TRANSSEPTAL SHEATH (VXS) AND VERSACROSS RF WIRE (VXW) WERE USED FOR THE INITIAL TRANSSEPTAL PUNCTURE. THE VST WAS WIRED WITH THE VXW AND INSERTED IN THE PATIENT. DUE TO COMPLEX PATIENT ANATOMY, THE PHYSICIAN EXPERIENCED DIFFICULTY MANEUVERING THE TRANSESOPHAGEAL ECHOCARDIOGRAPHY (TEE) PROBE TO VISUALIZE TENTING OF THE SEPTUM. THE PHYSICIAN ATTEMPTED TO REWIRE AND RE-POSITION THE VST ON THE SEPTUM SIX MORE TIMES. TENTING OF THE SEPTUM COULD NOT BE CONFIRMED DUE TO POOR IMAGING AND THE VST WAS EXCHANGED WITH THE VXS. THE VXS WAS SHAPED, WIRED AND INSERTED FOR POSITIONING. WITH NO CLEAR IMAGING OF THE SEPTUM OR THE VXS, THE PHYSICIAN REQUESTED ANOTHER SIZED SHEATH. BEFORE THE SHEATH COULD BE EXCHANGED, THE PATIENT'S BLOOD PRESSURE BEGAN TO DROP. A PERFORATION LEADING TO PERICARDIAL EFFUSION WAS NOTED IN THE RIGHT ATRIAL APPENDAGE. THE LAAC PROCEDURE WAS ABORTED. PERICARDIOCENTESIS WAS PERFORMED. CARDIOTHORACIC SURGERY WAS CONSULTED. MORE FLUID WAS REMOVED AND THE PATIENT WAS STABILIZED. THERE

    Patient problems: Pericardial Effusion

    MAUDE report 14597532 ↗
  • InjuryMay 4, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A CASE OF PERICARDIAL EFFUSION REQUIRING INTERVENTION WAS REPORTED DURING A LEFT ATRIAL APPENDAGE CLOSURE (LAAC) PROCEDURE WHERE THE VERSACROSS LARGE ACCESS TRANSSEPTAL DILATOR AND VERSACROSS RF WIRE WERE USED AMONG OTHER DEVICES INCLUDING THE WATCHMAN DEVICES (BOSTON SCIENTIFIC CORPORATION) THE PHYSICIAN EXPERIENCED DIFFICULTY VISUALIZING THE POSITIONING OF THE DEVICES DUE TO POOR TRANSESOPHAGEAL ECHOCARDIOGRAPHY (TEE) IMAGING. THE VERSACROSS DEVICES WERE POSITIONED ON THE SEPTUM AND TRANSSEPTAL PUNCTURE WAS SUCCESSFULLY COMPLETED. THE VERSACROSS RF WIRE, VERSACROSS LARGE ACCESS TRANSSEPTAL DILATOR AND WATCHMAN SHEATH WERE ADVANCED INTO THE LEFT ATRIUM. UNABLE TO VISUALIZE THE DEVICES USING TEE, THE PHYSICIAN USED CONTRAST FOR FLUOROSCOPIC VISUALIZATION. DARKENING IN THE PERICARDIAL SPACE WAS NOTED AND A PERICARDIAL EFFUSION WAS NOTED ON ECHO. THE LAAC PROCEDURE WAS ABORTED, ALL DEVICES WERE REMOVED AND PERICARDIOCENTESIS WAS PERFORMED. THE PATIENT REMAINED STABLE THROUGHOUT PROCEDURE AND WAS KEPT FOR OVERNIGHT OBSERVATION. THERE IS NO REASONABLE EVIDENCE TO SUGGEST THAT THE BAYLIS MEDICAL DEVICES CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. HOWEVER, AS BAYLIS MEDICAL DEVICES W

    Patient problems: Pericardial Effusion

    MAUDE report 14278755 ↗
  • InjuryMay 4, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A CASE OF PERICARDIAL EFFUSION REQUIRING INTERVENTION WAS REPORTED DURING A LEFT ATRIAL APPENDAGE CLOSURE (LAAC) PROCEDURE WHERE THE VERSACROSS LARGE ACCESS TRANSSEPTAL DILATOR AND VERSACROSS RF WIRE WERE USED AMONG OTHER DEVICES INCLUDING THE WATCHMAN DEVICES (BOSTON SCIENTIFIC CORPORATION) THE PHYSICIAN EXPERIENCED DIFFICULTY VISUALIZING THE POSITIONING OF THE DEVICES DUE TO POOR TRANSESOPHAGEAL ECHOCARDIOGRAPHY (TEE) IMAGING. THE VERSACROSS DEVICES WERE POSITIONED ON THE SEPTUM AND TRANSSEPTAL PUNCTURE WAS SUCCESSFULLY COMPLETED. THE VERSACROSS RF WIRE, VERSACROSS LARGE ACCESS TRANSSEPTAL DILATOR AND WATCHMAN SHEATH WERE ADVANCED INTO THE LEFT ATRIUM. UNABLE TO VISUALIZE THE DEVICES USING TEE, THE PHYSICIAN USED CONTRAST FOR FLUOROSCOPIC VISUALIZATION. DARKENING IN THE PERICARDIAL SPACE WAS NOTED AND A PERICARDIAL EFFUSION WAS NOTED ON ECHO. THE LAAC PROCEDURE WAS ABORTED, ALL DEVICES WERE REMOVED AND PERICARDIOCENTESIS WAS PERFORMED. THE PATIENT REMAINED STABLE THROUGHOUT PROCEDURE AND WAS KEPT FOR OVERNIGHT OBSERVATION. THERE IS NO REASONABLE EVIDENCE TO SUGGEST THAT THE BAYLIS MEDICAL DEVICES CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. HOWEVER, AS BAYLIS MEDICAL DEVICES W

    Patient problems: Pericardial Effusion

    MAUDE report 14278754 ↗
  • InjuryApr 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING A LEFT ATRIAL APPENDAGE CLOUSE (LAAC) PROCEDURE, THE VERSACROSS TRANSSEPTAL SHEATH AND VERSACROSS RF WIRE (BAYLIS MEDICAL COMPANY INC.) WERE USED FOR THE INITIAL TRANSSEPTAL PUNCTURE OF THE PROCEDURE WITH NO OBSERVED INITIAL COMPLICATIONS. AFTER TRANSSEPTAL PUNCTURE, THE WATCHMAN FLX (BOSTON SCIENTIFIC CORP.) DEVICES WERE USED TO COMPLETE THE LAAC PROCEDURE. UPON POST-PROCEDURE VERIFICATION, THE PATIENT WAS NOTED TO HAVE A HEMOTHORAX. THE HEMOTHORAX WAS DRAINED AND THE PATIENT WAS KEPT FOR OBSERVATION. FOLLOW-UP REQUESTS REVELED THAT THE PATIENT IS DOING WELL AND IS EXPECTED TO RECOVER. AS THE HEMOTHROAX WAS IDENTIFIED AFTER COMPLETION OF THE CLOSURE PROCEDURE, THERE IS NO OBJECTIVE EVIDENCE TO SUGGEST THAT THE BAYLIS MEDICAL DEVICES CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. HOWEVER, AS THE BAYLIS DEVICES WERE AMONG THE DEVICES USED IN THE PROCEDURE, BAYLIS MEDICAL HAS DECIDED TO SUBMIT THIS REPORT. A SEPARATE PROBLEM REPORT HAS BEEN SUBMITTED FOR THE VERSACROSS TRANSSEPTAL SHEATH WITH REPORT NUMBER 9710452-2022-00024.

    Patient problems: Hemothorax

    MAUDE report 14249848 ↗
  • MalfunctionFeb 28, 2022Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    A 30-MINUTE PROCEDURAL DELAY WAS REPORTED DUE TO KINKING OF THE VERSACROSS RF WIRE (RF WIRE) AFTER MULTIPLE ATTEMPTS WERE MADE TO CROSS A THICK CALCIFIED SEPTUM. FOLLOWING THIS ISSUE, THE RF WIRE WAS REMOVED FROM THE PATIENT TOGETHER WITH THE VERSACROSS SHEATH/DILATOR ASSEMBLY. UPON INSPECTION OF THE DEVICES, DAMAGE TO THE RF WIRE WAS NOTED. THE PROCEDURE WAS ABLE TO CONTINUE AFTER A NEW VERSACROSS RF WIRE WAS USED TO PERFORM A SUCCESSFUL TRANSSEPTAL PUNCTURE. WHILE THERE WAS NO PATIENT INJURY REPORTED, BAYLIS MEDICAL COMPANY INC. HAS DECIDED TO REPORT THIS EVENT DUE TO THE PROCEDURAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13624931 ↗
  • InjuryNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE VERSACROSS RF WIRE AND VERSACROSS TRANSSEPTAL SHEATH WERE USED DURING A MITRACLIP PROCEDURE. UPON TRANSSEPTAL ACCESS, A BLOOD CLOT WAS OBSERVED ON THE DEVICES ON THE LEFT SIDE OF THE HEART. HEPARIN WAS ADMINISTERED AND THE VERSACROSS RF WIRE WAS RETRACTED WHILE ASPIRATING. THE CLOT WAS CAPTURED AND NO ADVERSE REACTIONS OCCURRED. THERE IS NO EVIDENCE TO SUGGEST THAT THE BAYLIS MEDICAL DEVICES CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. HOWEVER, AS A BAYLIS MEDICAL DEVICES WERE REPORTED TO BE AMONG THE DEVICES USED IN THE PROCEDURE, BAYLIS MEDICAL HAS DECIDED TO SUBMIT THIS REPORT. A SEPARATE PROBLEM REPORT HAS BEEN SUBMITTED FOR THE VERSACROSS TRANSSEPTAL SHEATH WITH THE REPORT NUMBER 9710452-2021-00051.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 12804159 ↗
  • InjuryOct 21, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A CASE OF PERICARDIAL EFFUSION WAS REPORTED IN A PROCEDURE WHERE THE VERSACROSS RF WIRE AND VERSACROSS TRANSSEPTAL SHEATH WERE USED AMONG OTHER DEVICES, INCLUDING THE WATCHMAN SHEATH, A PIGTAIL CATHETER AND A 31 MM FLEX DEVICE. A TRANSSEPTAL PUNCTURE USING BAYLIS MEDICAL DEVICES WAS COMPLETED FOLLOWED BY THE DEPLOYMENT OF THE FLEX DEVICE INTO THE LEFT ATRIAL APPENDAGE (LAA) USING THE WATCHMAN DEPLOYMENT CATHETERS. IMMEDIATELY UPON RETRACTION OF THE DEPLOYMENT CATHETER SHEATH FROM THE LAA INTO THE RIGHT ATRIUM AND INFERIOR VENA CAVA, A PERICARDIAL EFFUSION WAS OBSERVED. PERICARDIOCENTESIS WAS PERFORMED AFTER PATIENT OBSERVATION FOR 25-30MIN, FOLLOWED BY OPEN HEART SURGERY. THE SURGERY WAS SUCCESSFUL AND THE PATIENT WAS DEEMED TO BE STABLE. THERE IS NO EVIDENCE TO SUGGEST THAT THE BAYLIS MEDICAL DEVICE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. HOWEVER, AS A BAYLIS MEDICAL DEVICE WAS REPORTED TO BE AMONG THE DEVICES USED IN THE PROCEDURE, BAYLIS MEDICAL HAS DECIDED TO SUBMIT THIS REPORT. THERE IS NO SUSPECTED DEVICE FAILURE. A SEPARATE MDR HAS BEEN SUBMITTED FOR THE VERSACROSS TRANSSEPTAL SHEATH UNDER MFR REPORT NUMBER: 9710452-2021-00048.

    Patient problems: Pericardial Effusion

    MAUDE report 12672244 ↗
  • MalfunctionAug 17, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    A 20-MINUTE PROCEDURAL DELAY WAS REPORTED IN A CASE WHERE THE VERSACROSS RF WIRE CAUSED AN ERROR CODE IN THE BAYLIS RADIOFREQUENCY PERFORATION GENERATOR. THE ERROR CODE PERSISTED AFTER USING ANOTHER GENERATOR. THE ISSUE WAS RESOLVED WHEN ANOTHER VERSACROSS RF WIRE WAS USED FOR THE PROCEDURE. WHILE THERE WAS NO PATIENT INJURY REPORTED, BAYLIS MEDICAL COMPANY INC. HAS DECIDED TO REPORT THIS EVENT DUE TO THE PROCEDURAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12334121 ↗
  • MalfunctionMay 12, 2021Device-Device Incompatibility; Device-Device Incompatibility

    A 30-MINUTE PROCEDURAL DELAY WAS REPORTED DUE TO THE BUCKLING OF THE VERSACROSS RF WIRE COATING. THE PROCEDURE WAS CONTINUED AFTER USER MANIPULATION TO ALLOW FOR THE VERSACROSS TRANSSEPTAL SHEATH TO BE EXCHANGED OVER THE WIRE. AFTER THE TRANSSEPTAL ACCESS, THE VERSACROSS RF WIRE WAS EXCHANGED FOR A DIFFERENT WIRE FOR THE REMAINING MITRACLIP PROCEDURE. WHILE THERE WAS NO PATIENT INJURY REPORTED, BAYLIS MEDICAL COMPANY INC. HAS DECIDED TO REPORT THIS EVENT DUE TO THE PROCEDURAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11814363 ↗
  • InjuryApr 16, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A CASE OF PERICARDIAL EFFUSION WAS REPORTED IN A MITRACLIP PROCEDURE WHERE THE VERSACROSS TRANSSEPTAL SHEATH AND THE VERSACROSS RF WIRE WERE USED AMONG OTHER DEVICES INCLUDING THE SUPRACORE WIRE (ABBOTT). A 3MM CUT WAS LOCATED BETWEEN THE RIGHT ATRIUM AND THE SUPERIOR VENA CAVA. PERICARDIOCENTESIS WAS PERFORMED TO DRAIN THE EFFUSION AND THE PROCEDURE WAS ABORTED. THERE IS NO EVIDENCE TO SUGGEST THAT THE BAYLIS MEDICAL DEVICES CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. HOWEVER, AS A BAYLIS MEDICAL DEVICES WERE REPORTED TO BE AMONG THE DEVICES USED IN THE PROCEDURE, BAYLIS MEDICAL HAS DECIDED TO SUBMIT THIS REPORT. A SEPARATE PROBLEM REPORT HAS BEEN SUBMITTED FOR THE VERSACROSS TRANSSEPTAL SHEATH WITH THE REPORT NUMBER 9710452-2021-0023.

    Patient problems: Pericardial Effusion

    MAUDE report 11679605 ↗
  • InjuryFeb 4, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A CASE OF PERICARDIAL EFFUSION WAS REPORTED DURING A PROCEDURE WHERE THE VERSACROSS RF WIRE, VERSACROSS TRANSSEPTAL SHEATH, NRG TRANSSEPTAL NEEDLE AND TORFLEX TRANSSEPTAL GUIDING SHEATH WERE USED. THE PATIENT HAD AN EXISTING PERICARDIAL EFFUSION THAT GREW DURING THE PROCEDURE DUE TO A SUSPECTED RIGHT ATRIUM PUNCTURE. THE PROCEDURE WAS ABORTED AND A PERICARDIOCENTESIS WAS PERFORMED. THE PATIENT STABILIZED AND WAS KEPT OVERNIGHT. THERE IS NO EVIDENCE TO SUGGEST THAT THE BAYLIS MEDICAL DEVICE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. HOWEVER, AS A BAYLIS MEDICAL DEVICE WAS REPORTED TO BE AMONG THE DEVICES USED IN THE PROCEDURE, BAYLIS MEDICAL HAS DECIDED TO SUBMIT THIS REPORT. SEPARATE PROBLEM REPORTS HAVE BEEN SUBMITTED FOR THE VERSACROSS TRANSSEPTAL SHEATH, NRG TRANSSEPTAL NEEDLE AND TORFLEX TRANSSEPTAL GUIDING SHEATH.

    Patient problems: Pericardial Effusion

    MAUDE report 11278193 ↗

Other Catheter, Septostomy devices

All Catheter, Septostomy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page