Skip to content

SureFlex Steerable Guiding Sheath TSK85-32-BD-71M (REF NA)Recalls & reported problems

Boston Scientific Corporation · SureFlex Steerable Sheath; Transseptal Dilator; J-Tip Guidewire · Class 2

All models of SureFlex Steerable Guiding Sheath (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00685447006312), received between Mar 29, 2023 and Dec 14, 2023 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Flushing Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Flushing Problem1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Transient Ischemic Attack1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 14, 2023Flushing Problem; Flushing Problem

    IT WAS REPORTED THE PATIENT STROKES OCCURRED. DURING AN ABLATION PROCEDURE, A SUREFLEX STEERABLE GUIDING SHEATH WAS SELECTED FOR USE. THE SHEATH WAS PREPARED AND THEN INTRODUCED INTO THE LEFT ATRIUM. THE SHEATH WAS ASPIRATED TO PREVENT BUBBLES. ONCE THE DEVICE WERE CONNECTED WITH A NON-BOSTON SCIENTIFIC MAPPING CATHETER AND ABLATING DEVICE, ABLATION WAS PERFORMED AND THE CATHETER WAS PULLED OUT FROM THE SHEATH. AT THIS POINT WHEN THE PHYSICIAN TRIED TO ASPIRATE, A RESISTANCE WAS FELT. THE PHYSICIAN IMMEDIATELY STOPPED THE PROCEDURE AND PULLED THE SHEATH OUT. OUTSIDE THE PATIENT BODY, THE SHEATH WAS FLUSHED BUT THERE WAS NO BLOOD COAGULUM. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS. THE SHEATH WAS FLUSHED/ASPIRATED WITHOUT ANYTHING IN IT, IT WAS AFTER GETTING THE ABLATION CATHETER OUT OF THE SHEATH IT COULDN'T BE ASPIRATED. NO DAMAGES OR ISSUES NOTED WITH THE STOPCOCK. AFTER GETTING THE SHEATH OUT THERE WERE NOTHING ON THE TIP. THE "LITTLE STROKES" WERE SLOWLY TRANSIENT AND THE PATIENT SEEM TO BE RECOVER.

    Patient problems: Transient Ischemic Attack

    MAUDE report 18325795 ↗
  • MalfunctionMar 29, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE TO TREAT ATRIAL FIBRILLATION (AFIB), A SUREFLEX STEERABLE SHEATH WAS SELECTED FOR USE. DURING THE PROCEDURE A LEAK WAS NOTED IN THE SHEATH. THE SHEATH WAS REPLACED, AND THE PROCEDURE WAS COMPLETED WITHOUT ANY PATIENT COMPLICATIONS. THE PRODUCT WILL BE RETURNED FOR LABORATORY ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16636685 ↗

Other Introducer, Catheter devices

All Introducer, Catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page