VersaCross Steerable Access Solution VSTK0016 (REF NA)Recalls & reported problems

Baylis Médicale Cie Inc · 1 x VersaCross Steerable Sheath (including 1 x Dilator and 1 x J-tip Guidewire); 1 x VersaCross RF Wire (including 1 x Connector Cable (Single Use)); 1 x Dispersive Electrode · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00685447007135), received between Jan 28, 2025 and Apr 14, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Integrity Problem1
  • Material Perforation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 14, 2025Material Perforation; Material Perforation

    IT WAS REPORTED THAT DURING A PULSE FIELD ABLATION PROCEDURE, A VERSACROSS STEERABLE ACCESS SOLUTION WAS SELECTED FOR USE. THE PHYSICIAN PERFORMED TRANSSEPTAL PUNCTURE SUCCESSFULLY AND UPON THE DEVICE REMOVAL, IT WAS NOTED THAT THE SHEATH TIP WAS DAMAGED. THERE WAS A TEAR NOTED IN THE PRODUCT. THE PHYSICIAN COMPLETED THE PROCEDURE, AND NO PATIENT COMPLICATIONS OCCURRED. THE DEVICE IS EXPECTED TO BE RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21836720 ↗
  • MalfunctionJan 28, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT VERSACROSS STEERABLE ACCESS SOLUTION WAS SELECTED FOR PULMONARY VEIN ISOLATION PROCEDURE. DURING THE PROCEDURE, WHEN VERSACROSS RF WIRE WAS INTRODUCED INTO THE PATIENT BODY FROM THE TRANSSEPTAL SHEATH, AND THE OUTER COATING OF THE TRANSSEPTAL WIRE CAME OUT FROM RIGHT SIDE OF THE MARKER BAND. THUS, THE VERSACROSS RF WIRE WAS REPLACED AND NEW TRANSSEPTAL WIRE WAS USED FORM THE DIFFERENT VERSACROSS KIT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS WERE REPORTED. THE DEVICE IS EXPECTED TO BE RETURN FOR ANALYSIS. THE DEVICE WAS INTACT AND WAS REMOVED WITHOUT ANY FURTHER COMPROMISE. ITS INSULATION STRIPPED OFF AT ITS PROXIMAL END. THE ISSUE WAS NOTED WHEN THE PHYSICIAN WAS INITIALLY FINDING THE MARKER BAND FOR REFERENCE TO MOVE TO THE GROIN ACCESS SITE, A SMALL TAB ATTACHED FROM THE PROXIMAL END WAS NOTED. THE DISTAL TIP WAS NOT TO THE HEART WHEN THIS WAS DISCOVERED AND THE DE-INSULATED PART NEVER WENT INTO ANY OTHER DEVICE, ONLY CAME OUT OF THE HOOP. THE RF WIRE WAS INSERTED INTO A HEMOSTATIC VALVE ON A SHORT SHEATH. THE PRODUCT WAS RECEIVED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21256488 ↗

Other Catheter, Septostomy devices