Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00685447007449), received between Dec 19, 2024 and Jan 21, 2025 — classified as 1 death, 1 malfunction.

The most frequently reported specific device problem is “Thermal Decomposition of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Thermal Decomposition of Device1

Patient problems

  • Cardiac Arrest1
  • Cardiac Perforation1
  • Cardiac Tamponade1
  • No Clinical Signs, Symptoms or Conditions1
  • Pericardial Effusion1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • DeathJan 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED A PATIENT DEATH OCCURRED. THE PROCEDURE WAS CANCELLED. DURING A WATCHMAN FLX PRO PROCEDURE, A VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE. A TRACE EFFUSION WAS NOTED PRIOR TO PROCEDURE. THE PHYSICIAN OBTAINED RIGHT FEMORAL VEIN ACCESS AND AN 18GUAGE SHEATH WAS PLACED. THE VERSACROSS SHEATH WAS INSERTED. THE PHYSICIAN DROPPED BELOW THE SEPTUM ON FIRST ATTEMPT, SO IT WAS RE-ATTEMPTED TO RE-WIRE THE SUPERIOR VENA CAVA (SVC) TO COME BACK DOWN TO THE SEPTUM. THE TRANSSEPTAL PUNCTURE WAS PERFORMED AND THE VERSACROSS SHEATH WAS EXCHANGED FOR THE WATCHMAN TRUSTEER SHEATH OVER THE WIRE. A PIGTAIL CATHETER WENT IN OVER THE WIRE. AT THIS POINT, AN EFFUSION WAS NOTED. AUTO TRANSFUSION BEGAN WITH THAT PIGTAIL CATHETER WHILE A PERICARDIOCENTESIS WAS PERFORMED. AFTER THAT THERE WERE TWO PIGTAIL CATHETERS PLACED IN THE PERICARDIAL SPACE. 12 UNITS OF PACKED RED BLOOD CELLS (PRBCS), 8 UNITS OF FRESH FROZEN PLASMA (FFP) AND 1 UNIT OF PLATELETS WERE GIVEN AND THE PATIENT WAS TAKEN TO THE OPERATING ROOM (OR). THE PATIENT ARRESTED BEFORE GETTING TRANSFERRED THE OR TABLE, PATIENT CHEST WAS OPENED BUT WERE UNABLE TO RESUSCITATE. IT WAS DETERMINED THE CAUSE OF DEATH WAS DUE TO PERFO

    Patient problems: Cardiac Arrest; Cardiac Perforation; Cardiac Tamponade; Pericardial Effusion

    MAUDE report 21197451 ↗
  • MalfunctionDec 19, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    IT HAS BEEN REPORTED THAT A VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE FOR A PULSE FIELD ABLATION (PVI) PROCEDURE WITH FARAPULSE. AFTER TRANSSEPTAL CROSSING, WHEN THE VERSACROSS RF WIRE WAS REMOVED, THE PHYSICIAN NOTED THAT TIP OF THE WIRE WAS BLACK. THE CHARRING COULD NOT BE REMOVED. NO PATIENT COMPLICATIONS REPORTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. PRODUCT IS AVAILABLE FOR RETURN. NO OTHER ISSUES WERE REPORTED. IT WAS FURTHER REPORTED THAT THE RADIO FREQUENCY (RF) WAS APPLIED TWICE USING THE SAME RF WIRE ON THE 1 SEC CONSTANT SETTING. THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20977297 ↗

FDA clearance and classification

The FDA regulates product code DXF (catheter, septostomy) under 21 CFR 870.5175 as a Class II device. 21 CFR 870.5175 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VersaCross Access Solution VXSK0023

Has the VersaCross Access Solution VXSK0023 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00685447007449), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VersaCross Access Solution VXSK0023?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Dec 19, 2024 and Jan 21, 2025. In those reports, the most common specific device problem is “Thermal Decomposition of Device” (1); the most common patient problems are “Cardiac Arrest” (1), “Cardiac Perforation” (1) and “Cardiac Tamponade” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VersaCross Access Solution VXSK0023?

1 report is classified as a death, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the VersaCross Access Solution VXSK0023, and is it still sold?

The FDA's device database lists Baylis Médicale Cie Inc as the company responsible for it (brand VersaCross Access Solution, model VXSK0023). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, Septostomy devices

All Catheter, Septostomy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page