VersaCross Steerable Access Solution VSTK0021Recalls & reported problems

Baylis Médicale Cie Inc · 1 x VersaCross Steerable Sheath (including 1 x Dilator and 1 x J-tip Guidewire); 1 x VersaCross RF Wire (including 1 x Connector Cable (Single Use)) · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00685447012122), received between Jul 12, 2023 and Jun 25, 2024 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Advance” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance2
  • Failure to Advance2
  • Leak/Splash1

Patient problems

  • Pericardial Effusion3
  • Cardiac Tamponade2
  • Low Blood Pressure/ Hypotension2
  • Perforation2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 25, 2024Difficult to Advance; Difficult to Advance

    IT WAS REPORTED A PERICARDIAL EFFUSION (PE) OCCURRED. DURING AN ATRIAL FIBRILLATION (A FIB) ABLATION PROCEDURE, A VERSACROSS STEERABLE KIT WAS SELECTED FOR USE. THE PATIENT HAD A VERY THICK SEPTUM CAUSING DIFFICULTY TO CROSS FOR TRANSSEPTAL ACCESS. INITIALLY, THERE WERE THREE ATTEMPTS TO CROSS THE SEPTUM MADE BY THE PHYSICIAN AND THEN ANOTHER PHYSICIAN AS CONSULTED. AT THIS TIME, THEY CHOSE TO INCREASE THE RADIO FREQUENCY (RF) DELIVERY OF THE VERSACROSS RF WIRE FROM 1 TO 2 SECONDS/CONSTANT TO ACCOMMODATE FOR THE THICK SEPTUM. THE FOURTH ATTEMPT WAS THEN SUCCESSFUL. THE PATIENT REMAINED STABLE THROUGHOUT THE ENTIRE PROCEDURE HOWEVER LATER IT WAS FOUND TO HAVE A PERICARDIAL EFFUSION UPON THE ROUTINE POST PROCEDURAL IMAGING. HENCE, A PERICARDIOCENTESIS WAS PERFORMED TO ADDRESS IT. THE PATIENT LEFT THE LAB IN A STABLE STATE, WAS ADMITTED FOR MONITORING OF THE EFFUSION, AND IT WAS LATER DISCHARGED IN STABLE CONDITION. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS (DISPOSED). THE PROCEDURE TOOK ABOUT 3 HOURS IN TOTAL WHERE THE TRANSSEPTAL ACCESS WAS OBTAINED IN AROUND TWO HOURS. NO ISSUES WITH THE VERSACROSS RF WIRE WERE NOTED DURING THE FIRST THREE ATTEMPTS. THE FIRST THREE ATT

    Patient problems: Pericardial Effusion

    MAUDE report 19609449 ↗
  • MalfunctionJun 10, 2024Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT A VERSACROSS STEERABLE ACCESS SOLUTION KIT WAS SELECTED FOR USE. UPON OPENING THE PACKAGE, A LEAK WAS NOTED IN ONE OF THE DEVICES. HENCE, THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. NO PATIENT COMPLICATIONS OCCURRED. THE DEVICE IS EXPECTED TO BE RETURNED FOR ANALYSIS. NO OTHER ISSUES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19505237 ↗
  • InjuryJul 12, 2023Failure to Advance; Difficult to Advance; Failure to Advance; Difficult to Advance

    IT WAS REPORTED THAT DURING A WATCHMAN PROCEDURE TO TREAT ATRIAL FIBRILLATION (A FIB) IN LEFT ATRIAL APPENDAGE (LAA), A VERSACROSS CONNECT SOLUTION WAS SELECTED FOR USE AND PERFORATION HAPPENED ON THE RIGHT SIDE OF THE HEART WHILE ATTEMPTING TO GO TRANSSEPTAL. THERE WAS A BASELINE TRACE EFFUSION WHEN WE STARTED. JUST A NORMAL AMOUNT OF FLUID IN THE PERICARDIAL SPACE. THE PATIENT HAD A LIPOMATOUS SEPTUM (SUPERIORLY) AND A LARGE EUSTACHIAN VALVE THAT KEPT PUSHING THE TRANSSEPTAL SYSTEM OFF (WATCHMAN FXD DOUBLE CURVE AND VERSACROSS CONNECT). HENCE IN ORDER TO ATTEMPT TO GET AND MAINTAIN CONTACT WITH THE FOSSA OVALIS, A VERSACROSS STEERABLE WAS OPENED TO TRY TO GET TRANSSEPTAL. IT MADE CONTACT WITH THE SEPTUM INFERIORLY AND POSTERIORLY ON TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) AND RADIO FREQUENCY WAS NOT APPLIED AND HENCE NEVER CROSSED THE SEPTUM. APPROXIMATELY 10 MINUTES AFTER INSERTING THE STEERABLE VERSACROSS, THE PATIENT'S PRESSURE DROPPED AND A PERICARDIAL EFFUSION WAS NOTED ON THE ECHO. A DRAIN WAS PLACED AND PATIENT WAS TAKEN TO THE OPERATION ROOM (OR) WHERE A STERNOTOMY WITH A REPAIR AND A LEFT ATRIAL APPENDAGE LIGATION WAS PERFORMED. THE SURGEON NOTED A 4MM HOLE IN A THE TRANSVE

    Patient problems: Cardiac Tamponade; Low Blood Pressure/ Hypotension; Perforation; Pericardial Effusion

    MAUDE report 17305765 ↗
  • InjuryJul 12, 2023Failure to Advance; Failure to Advance

    IT WAS REPORTED THAT DURING A WATCHMAN PROCEDURE TO TREAT ATRIAL FIBRILLATION (A FIB) IN LEFT ATRIAL APPENDAGE (LAA), A VERSACROSS CONNECT SOLUTION WAS SELECTED FOR USE AND PERFORATION HAPPENED ON THE RIGHT SIDE OF THE HEART WHILE ATTEMPTING TO GO TRANSSEPTAL. THERE WAS A BASELINE TRACE EFFUSION WHEN WE STARTED. JUST A NORMAL AMOUNT OF FLUID IN THE PERICARDIAL SPACE. THE PATIENT HAD A LIPOMATOUS SEPTUM (SUPERIORLY) AND A LARGE EUSTACHIAN VALVE THAT KEPT PUSHING THE TRANSSEPTAL SYSTEM OFF (WATCHMAN FXD DOUBLE CURVE AND VERSACROSS CONNECT). HENCE IN ORDER TO ATTEMPT TO GET AND MAINTAIN CONTACT WITH THE FOSSA OVALIS, A VERSACROSS STEERABLE WAS OPENED TO TRY TO GET TRANSSEPTAL. IT MADE CONTACT WITH THE SEPTUM INFERIORLY AND POSTERIORLY ON TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) AND RADIO FREQUENCY WAS NOT APPLIED AND HENCE NEVER CROSSED THE SEPTUM. APPROXIMATELY 10 MINUTES AFTER INSERTING THE STEERABLE VERSACROSS, THE PATIENT'S PRESSURE DROPPED AND A PERICARDIAL EFFUSION WAS NOTED ON THE ECHO. A DRAIN WAS PLACED AND PATIENT WAS TAKEN TO THE OPERATION ROOM (OR) WHERE A STERNOTOMY WITH A REPAIR AND A LEFT ATRIAL APPENDAGE LIGATION WAS PERFORMED. THE SURGEON NOTED A 4MM HOLE IN A THE TRANSVE

    Patient problems: Cardiac Tamponade; Low Blood Pressure/ Hypotension; Perforation; Pericardial Effusion

    MAUDE report 17305749 ↗

Other Catheter, Septostomy devices