VersaCross Steerable Access Solution VSTK0021Recalls & reported problems
Baylis Médicale Cie Inc · 1 x VersaCross Steerable Sheath (including 1 x Dilator and 1 x J-tip Guidewire); 1 x VersaCross RF Wire (including 1 x Connector Cable (Single Use)) · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00685447012122), received between Jul 12, 2023 and Jun 25, 2024 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Difficult to Advance” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Advance2
- Failure to Advance2
- Leak/Splash1
Patient problems
- Pericardial Effusion3
- Cardiac Tamponade2
- Low Blood Pressure/ Hypotension2
- Perforation2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJun 25, 2024Difficult to Advance; Difficult to Advance
IT WAS REPORTED A PERICARDIAL EFFUSION (PE) OCCURRED. DURING AN ATRIAL FIBRILLATION (A FIB) ABLATION PROCEDURE, A VERSACROSS STEERABLE KIT WAS SELECTED FOR USE. THE PATIENT HAD A VERY THICK SEPTUM CAUSING DIFFICULTY TO CROSS FOR TRANSSEPTAL ACCESS. INITIALLY, THERE WERE THREE ATTEMPTS TO CROSS THE SEPTUM MADE BY THE PHYSICIAN AND THEN ANOTHER PHYSICIAN AS CONSULTED. AT THIS TIME, THEY CHOSE TO INCREASE THE RADIO FREQUENCY (RF) DELIVERY OF THE VERSACROSS RF WIRE FROM 1 TO 2 SECONDS/CONSTANT TO ACCOMMODATE FOR THE THICK SEPTUM. THE FOURTH ATTEMPT WAS THEN SUCCESSFUL. THE PATIENT REMAINED STABLE THROUGHOUT THE ENTIRE PROCEDURE HOWEVER LATER IT WAS FOUND TO HAVE A PERICARDIAL EFFUSION UPON THE ROUTINE POST PROCEDURAL IMAGING. HENCE, A PERICARDIOCENTESIS WAS PERFORMED TO ADDRESS IT. THE PATIENT LEFT THE LAB IN A STABLE STATE, WAS ADMITTED FOR MONITORING OF THE EFFUSION, AND IT WAS LATER DISCHARGED IN STABLE CONDITION. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS (DISPOSED). THE PROCEDURE TOOK ABOUT 3 HOURS IN TOTAL WHERE THE TRANSSEPTAL ACCESS WAS OBTAINED IN AROUND TWO HOURS. NO ISSUES WITH THE VERSACROSS RF WIRE WERE NOTED DURING THE FIRST THREE ATTEMPTS. THE FIRST THREE ATT
Patient problems: Pericardial Effusion
MAUDE report 19609449 ↗MalfunctionJun 10, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THAT A VERSACROSS STEERABLE ACCESS SOLUTION KIT WAS SELECTED FOR USE. UPON OPENING THE PACKAGE, A LEAK WAS NOTED IN ONE OF THE DEVICES. HENCE, THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. NO PATIENT COMPLICATIONS OCCURRED. THE DEVICE IS EXPECTED TO BE RETURNED FOR ANALYSIS. NO OTHER ISSUES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19505237 ↗InjuryJul 12, 2023Failure to Advance; Difficult to Advance; Failure to Advance; Difficult to Advance
IT WAS REPORTED THAT DURING A WATCHMAN PROCEDURE TO TREAT ATRIAL FIBRILLATION (A FIB) IN LEFT ATRIAL APPENDAGE (LAA), A VERSACROSS CONNECT SOLUTION WAS SELECTED FOR USE AND PERFORATION HAPPENED ON THE RIGHT SIDE OF THE HEART WHILE ATTEMPTING TO GO TRANSSEPTAL. THERE WAS A BASELINE TRACE EFFUSION WHEN WE STARTED. JUST A NORMAL AMOUNT OF FLUID IN THE PERICARDIAL SPACE. THE PATIENT HAD A LIPOMATOUS SEPTUM (SUPERIORLY) AND A LARGE EUSTACHIAN VALVE THAT KEPT PUSHING THE TRANSSEPTAL SYSTEM OFF (WATCHMAN FXD DOUBLE CURVE AND VERSACROSS CONNECT). HENCE IN ORDER TO ATTEMPT TO GET AND MAINTAIN CONTACT WITH THE FOSSA OVALIS, A VERSACROSS STEERABLE WAS OPENED TO TRY TO GET TRANSSEPTAL. IT MADE CONTACT WITH THE SEPTUM INFERIORLY AND POSTERIORLY ON TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) AND RADIO FREQUENCY WAS NOT APPLIED AND HENCE NEVER CROSSED THE SEPTUM. APPROXIMATELY 10 MINUTES AFTER INSERTING THE STEERABLE VERSACROSS, THE PATIENT'S PRESSURE DROPPED AND A PERICARDIAL EFFUSION WAS NOTED ON THE ECHO. A DRAIN WAS PLACED AND PATIENT WAS TAKEN TO THE OPERATION ROOM (OR) WHERE A STERNOTOMY WITH A REPAIR AND A LEFT ATRIAL APPENDAGE LIGATION WAS PERFORMED. THE SURGEON NOTED A 4MM HOLE IN A THE TRANSVE
Patient problems: Cardiac Tamponade; Low Blood Pressure/ Hypotension; Perforation; Pericardial Effusion
MAUDE report 17305765 ↗InjuryJul 12, 2023Failure to Advance; Failure to Advance
IT WAS REPORTED THAT DURING A WATCHMAN PROCEDURE TO TREAT ATRIAL FIBRILLATION (A FIB) IN LEFT ATRIAL APPENDAGE (LAA), A VERSACROSS CONNECT SOLUTION WAS SELECTED FOR USE AND PERFORATION HAPPENED ON THE RIGHT SIDE OF THE HEART WHILE ATTEMPTING TO GO TRANSSEPTAL. THERE WAS A BASELINE TRACE EFFUSION WHEN WE STARTED. JUST A NORMAL AMOUNT OF FLUID IN THE PERICARDIAL SPACE. THE PATIENT HAD A LIPOMATOUS SEPTUM (SUPERIORLY) AND A LARGE EUSTACHIAN VALVE THAT KEPT PUSHING THE TRANSSEPTAL SYSTEM OFF (WATCHMAN FXD DOUBLE CURVE AND VERSACROSS CONNECT). HENCE IN ORDER TO ATTEMPT TO GET AND MAINTAIN CONTACT WITH THE FOSSA OVALIS, A VERSACROSS STEERABLE WAS OPENED TO TRY TO GET TRANSSEPTAL. IT MADE CONTACT WITH THE SEPTUM INFERIORLY AND POSTERIORLY ON TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) AND RADIO FREQUENCY WAS NOT APPLIED AND HENCE NEVER CROSSED THE SEPTUM. APPROXIMATELY 10 MINUTES AFTER INSERTING THE STEERABLE VERSACROSS, THE PATIENT'S PRESSURE DROPPED AND A PERICARDIAL EFFUSION WAS NOTED ON THE ECHO. A DRAIN WAS PLACED AND PATIENT WAS TAKEN TO THE OPERATION ROOM (OR) WHERE A STERNOTOMY WITH A REPAIR AND A LEFT ATRIAL APPENDAGE LIGATION WAS PERFORMED. THE SURGEON NOTED A 4MM HOLE IN A THE TRANSVE
Patient problems: Cardiac Tamponade; Low Blood Pressure/ Hypotension; Perforation; Pericardial Effusion
MAUDE report 17305749 ↗
FDA clearance and classification
The FDA regulates product code DXF (catheter, septostomy) under 21 CFR 870.5175 as a Class II device. 21 CFR 870.5175 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the VersaCross Steerable Access Solution VSTK0021
Has the VersaCross Steerable Access Solution VSTK0021 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00685447012122), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the VersaCross Steerable Access Solution VSTK0021?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jul 12, 2023 and Jun 25, 2024. In those reports, the most common specific device problems are “Difficult to Advance” (2), “Failure to Advance” (2) and “Leak/Splash” (1); the most common patient problems are “Pericardial Effusion” (3), “Cardiac Tamponade” (2) and “Low Blood Pressure/ Hypotension” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VersaCross Steerable Access Solution VSTK0021?
None of the reports are classified as deaths, 3 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the VersaCross Steerable Access Solution VSTK0021, and is it still sold?
The FDA's device database lists Baylis Médicale Cie Inc as the company responsible for it (brand VersaCross Steerable Access Solution, model VSTK0021). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, Septostomy devices
- VersaCross Access Solution VXSK0031Baylis Médicale Cie Inc · Catheter, Septostomy1 recall16 reports
- VersaCross Access Solution VXSK0021Baylis Médicale Cie Inc · Catheter, Septostomy1 recall1 death15 reports
- VersaCross Access Solution VXSK0032Baylis Médicale Cie Inc · Catheter, Septostomy1 recall14 reports
- VersaCross Access Solution VXSK0022Baylis Médicale Cie Inc · Catheter, Septostomy1 recall13 reports
- VersaCross Access Solution VXSK0024Baylis Médicale Cie Inc · Catheter, Septostomy1 recall3 reports
- VersaCross Access Solution VXSK0034Baylis Médicale Cie Inc · Catheter, Septostomy1 recall1 report
- VersaCross Access Solution VXSK0025Baylis Médicale Cie Inc · Catheter, Septostomy1 recall1 report
- NRG Transseptal Needle NRG-E-HF-71-C0Boston Scientific Corporation · Catheter, Septostomy22 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page