VersaCross Access Solution VXSK0003Recalls & reported problems

Baylis Médicale Cie Inc · 1 x VersaCross Transseptal Sheath (including 1 x J-tip Guidewire); 1 x VersaCross RF Wire (including 1 x Connector Cable (Single Use)) · Class 2

All models of VersaCross Access Solution (19)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00685447017066), received between Jun 5, 2025 and Jun 5, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Display or Visual Feedback Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Display or Visual Feedback Problem2
  • Positioning Problem1

Patient problems

  • Cardiac Tamponade2
  • Low Blood Pressure/ Hypotension2
  • Pericardial Effusion2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 5, 2025Display or Visual Feedback Problem; Positioning Problem; Display or Visual Feedback Problem; Positioning Problem

    IT WAS REPORTED A PERICARDIAL EFFUSION (PE) AND CARDIAC TAMPONADE OCCURRED. THE PROCEDURE WAS CANCELLED. DURING A PULMONARY VEIN ISOLATION PULSE FIELD ABLATION PROCEDURE, A VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE. THE PATIENT WAS PLACED UNDER GENERAL ANESTHESIA. A PRE-PROCEDURE TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) CONFIRMED THERE WAS NO CLOT IN THE APPENDAGE AND THERE WAS NO MENTION OF A BASELINE EFFUSION. THE PHYSICIAN OBTAINED GROIN ACCESS AND INTRACARDIAC ECHOCARDIOGRAPHY (ICE) WAS PLACED. THE ATRIUM WAS EXTREMELY DILATED AND THE PHYSICIAN STRUGGLED TO GET THE TIP OF THE SHEATH ON FOSSA. THE DILATOR WAS WITHDRAWN OUT OF THE BODY, MORE CURVE WAS ADDED AND PLACED IT BACK INTO THE BODY THROUGH THE SHEATH THAT WAS IN THE BODY. WITH THE EXTRA CURVE PLACED ON THE DILATOR, THE PHYSICIAN WAS ABLE TO GET THE SHEATH TO REACH THE FOSSA. ONCE DROPPED ONTO THE FOSSA, THE RADIO FREQUENCY (RF) WAS APPLIED FOR TRANSSEPTAL PUNCTURE (TSP). THE TSP APPEARED TO VERY SUPERIOR AND THERE WAS NO CLEAR TENTING ON ICE. A BUBBLE STUDY WAS PERFORMED AND THERE WERE BUBBLES IN THE LEFT ATRIUM (LA) AND IT WAS CONFIRMED THE VERSACROSS SHEATH IN THE LA. AN ARTERIOGRAM CONTRAST SHOT WAS PERFORMED. AT

    Patient problems: Cardiac Tamponade; Low Blood Pressure/ Hypotension; Pericardial Effusion

    MAUDE report 22149117 ↗
  • InjuryJun 5, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED A PERICARDIAL EFFUSION (PE) AND CARDIAC TAMPONADE OCCURRED. THE PROCEDURE WAS CANCELLED. DURING A PULMONARY VEIN ISOLATION PULSE FIELD ABLATION PROCEDURE, A VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE. THE PATIENT WAS PLACED UNDER GENERAL ANESTHESIA. A PRE-PROCEDURE TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) CONFIRMED THERE WAS NO CLOT IN THE APPENDAGE AND THERE WAS NO MENTION OF A BASELINE EFFUSION. THE PHYSICIAN OBTAINED GROIN ACCESS AND INTRACARDIAC ECHOCARDIOGRAPHY (ICE) WAS PLACED. THE ATRIUM WAS EXTREMELY DILATED AND THE PHYSICIAN STRUGGLED TO GET THE TIP OF THE SHEATH ON FOSSA. THE DILATOR WAS WITHDRAWN OUT OF THE BODY, MORE CURVE WAS ADDED AND PLACED IT BACK INTO THE BODY THROUGH THE SHEATH THAT WAS IN THE BODY. WITH THE EXTRA CURVE PLACED ON THE DILATOR, THE PHYSICIAN WAS ABLE TO GET THE SHEATH TO REACH THE FOSSA. ONCE DROPPED ONTO THE FOSSA, THE RADIO FREQUENCY (RF) WAS APPLIED FOR TRANSSEPTAL PUNCTURE (TSP). THE TSP APPEARED TO VERY SUPERIOR AND THERE WAS NO CLEAR TENTING ON ICE. A BUBBLE STUDY WAS PERFORMED AND THERE WERE BUBBLES IN THE LEFT ATRIUM (LA) AND IT WAS CONFIRMED THE VERSACROSS SHEATH IN THE LA. AN ARTERIOGRAM CONTRAST SHOT WAS PERFORMED. AT

    Patient problems: Cardiac Tamponade; Low Blood Pressure/ Hypotension; Pericardial Effusion

    MAUDE report 22149041 ↗

FDA clearance and classification

The FDA regulates product code DXF (catheter, septostomy) under 21 CFR 870.5175 as a Class II device. 21 CFR 870.5175 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VersaCross Access Solution VXSK0003

Has the VersaCross Access Solution VXSK0003 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00685447017066), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VersaCross Access Solution VXSK0003?

The FDA's MAUDE database lists 2 adverse event reports for this device, received on Jun 5, 2025. In those reports, the most common specific device problems are “Display or Visual Feedback Problem” (2) and “Positioning Problem” (1); the most common patient problems are “Cardiac Tamponade” (2), “Low Blood Pressure/ Hypotension” (2) and “Pericardial Effusion” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VersaCross Access Solution VXSK0003?

None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the VersaCross Access Solution VXSK0003, and is it still sold?

The FDA's device database lists Baylis Médicale Cie Inc as the company responsible for it (brand VersaCross Access Solution, model VXSK0003). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.