VersaCross Access Solution VXSK0021 (UDI 00685447017240)Recalls & reported problems
Baylis Médicale Cie Inc · 1 x VersaCross Transseptal Sheath (including 1 x J-tip Guidewire); 1 x VersaCross RF Wire (including 1 x Connector Cable (Single Use)) · Class 2
All models of VersaCross Access Solution (19)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00685447017240), received between Dec 4, 2024 and Apr 22, 2025 — classified as 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Failure to Advance” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Failure to Advance2
- Contamination1
- Difficult to Advance1
- Difficult to Remove1
- Display or Visual Feedback Problem1
- Leak/Splash1
- Material Frayed1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Low Blood Pressure/ Hypotension3
- Pericardial Effusion3
- Thrombosis/Thrombus2
- Cardiac Arrest1
- Cardiac Perforation1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 22, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING AN ATRIAL FIBRILLATION ABLATION PROCEDURE, A VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE. AFTER THE SHEATH WAS ACROSS THE SEPTUM, THE PHYSICIAN NOTICED THAT BLOOD WAS DRIPPING OUT OF THE BACK OF THE HEMOSTATIC VALVE. THE PHYSICIAN OPTED TO CONTINUE THE PROCEDURE WITHOUT REPLACING THE SHEATH WITH SOME FLUID LEAKING FOR THE REST OF THE CASE. ABOUT 25CC OF BLOOD LOSS OCCURRED IN TOTAL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS OCCURRED. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS (DISPOSED).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21882717 ↗InjuryFeb 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT EXPERIENCED HYPOTENSION AND A PERICARDIAL EFFUSION. DURING A PULSED FIELD ABLATION (PFA) PROCEDURE TO TREAT AN ATRIAL FIBRILLATION, A VERSACROSS CONNECT FOR FARADRIVE AND VERSACROSS ACCESS SOLUTION KITS WERE SELECTED FOR USE. THE PHYSICIAN HAD DIFFICULTIES GETTING TRANSEPTAL ACCESS USING VERSACROSS CONNECT WHERE TENTING WAS CONTINUOUSLY ANTERIOR. THE PHYSICIAN STRUGGLED TO CLOCK POSTERIOR TO GET RIGHT POSITIONING AND TRIED MULTIPLE TIMES TO ADVANCE RF WIRE OUT AS THE PHYSICIAN BELIEVED WIRE CROSSED WITHOUT APPLYING RADIO FREQUENCY (RF) BUT COULD NOT BE CONFIRMED ON INTRACARDIAC ECHOCARDIOGRAPHY (ICE) OR FLUOROSCOPY. THE PHYSICIAN REPLACED ALL DEVICES AND TRANSITIONED TO VERSACROSS ACCESS, SHAPED THE DILATOR AND WENT BACK INTO CROSS. ONCE CROSSED WITH THE DILATOR/SHEATH, IT SEEMED TO BE AT ANTERIOR POSITION AND THE WIRE ADVANCED INTO THE APPENDAGE. THE PHYSICIAN MANIPULATED THE SHEATH/WIRE TO ENSURE THE LEFT SUPERIOR VEIN WIRED BEFORE REMOVING THE DILATOR AND WIRE. THE CASE PROCEEDED, AND 50 APPLICATIONS WERE DELIVERED WITH THE FARAWAVE CATHETER ON ALL PULMONARY VEINS SUCCESSFULLY. THE PHYSICIAN THEN REMOVED THE CATHETER TO COMPLETE A POST MAP, AND TOW
Patient problems: Low Blood Pressure/ Hypotension; Pericardial Effusion; Thrombosis/Thrombus
MAUDE report 21454100 ↗InjuryFeb 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT EXPERIENCED HYPOTENSION AND A PERICARDIAL EFFUSION. DURING A PULSED FIELD ABLATION (PFA) PROCEDURE TO TREAT AN ATRIAL FIBRILLATION, A VERSACROSS CONNECT FOR FARADRIVE AND VERSACROSS ACCESS SOLUTION KITS WERE SELECTED FOR USE. THE PHYSICIAN HAD DIFFICULTIES GETTING TRANSEPTAL ACCESS USING VERSACROSS CONNECT WHERE TENTING WAS CONTINUOUSLY ANTERIOR. THE PHYSICIAN STRUGGLED TO CLOCK POSTERIOR TO GET RIGHT POSITIONING AND TRIED MULTIPLE TIMES TO ADVANCE RF WIRE OUT AS THE PHYSICIAN BELIEVED WIRE CROSSED WITHOUT APPLYING RADIO FREQUENCY (RF) BUT COULD NOT BE CONFIRMED ON INTRACARDIAC ECHOCARDIOGRAPHY (ICE) OR FLUOROSCOPY. THE PHYSICIAN REPLACED ALL DEVICES AND TRANSITIONED TO VERSACROSS ACCESS, SHAPED THE DILATOR AND WENT BACK INTO CROSS. ONCE CROSSED WITH THE DILATOR/SHEATH, IT SEEMED TO BE AT ANTERIOR POSITION AND THE WIRE ADVANCED INTO THE APPENDAGE. THE PHYSICIAN MANIPULATED THE SHEATH/WIRE TO ENSURE THE LEFT SUPERIOR VEIN WIRED BEFORE REMOVING THE DILATOR AND WIRE. THE CASE PROCEEDED, AND 50 APPLICATIONS WERE DELIVERED WITH THE FARAWAVE CATHETER ON ALL PULMONARY VEINS SUCCESSFULLY. THE PHYSICIAN THEN REMOVED THE CATHETER TO COMPLETE A POST MAP, AND TOW
Patient problems: Low Blood Pressure/ Hypotension; Pericardial Effusion; Thrombosis/Thrombus
MAUDE report 21454057 ↗MalfunctionFeb 6, 2025Material Frayed; Difficult to Advance; Material Integrity Problem; Physical Resistance/Sticking; Material Frayed; Difficult to Advance; Material Integrity Problem; Physical Resistance/Sticking
IT HAS BEEN REPORTED THAT A VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE FOR AN ATRIAL FIBRILLATION (A FIB) PROCEDURE. DURING INSERTION OF THE VERSACROSS RF WIRE THROUGH THE DILATOR, THE USER MENTIONED THAT THEY HAD DIFFICULTY ADVANCING THE RF WIRE THROUGH DILATOR THAT'S WHEN THEY PULLED THE WIRES ALL THE WAY OUT OF ITS ORIGINAL PACKAGING. UPON FURTHER INSPECTION, THE TIP OF THE RF WIRE HAD FRAYING ELECTRICAL COATING THAT WAS FLAKING OFF OF THE WIRE. A NEW KIT WAS THEN OPENED TO REPLACE THE DEVICE, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS REPORTED. PRODUCT IS NOT EXPECTED TO RETURN FOR ANALYSIS (DISPOSED). THE PHYSICIAN HAD A HARD TIME ADVANCING THE RF WIRE, AND WHEN LOOKED AT THE PROXIMAL END OF THE WIRE, IT WAS NOTICED THE INSULATION WAS FLAKING OFF AND STICKING OUT FROM THE CATHETER, GETTING STUCK ON THE DILATOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21323398 ↗MalfunctionJan 9, 2025Contamination; Unsealed Device Packaging; Contamination; Unsealed Device Packaging
IT WAS REPORTED THAT A VERSACROSS ACCESS SOLUTION WAS SENT OUT TO THE CUSTOMER AND THE STERILE PACKAGE WAS COMPROMISED. THE POUCH WAS MISSING. THE DISCOVERY WAS MADE BEFORE THE PROCEDURE AND WAS A DEVICE PULLED FROM THE CURRENT INVENTORY. NO PATIENT COMPLICATIONS WERE REPORTED. NO DAMAGE WAS NOTED ON THE DEVICES. THE DEVICE IS NOT RETURNING FOR ANALYSIS (DISPOSED).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21119492 ↗MalfunctionJan 8, 2025Failure to Advance; Failure to Advance
IT WAS REPORTED THAT VERSACROSS ACCESS SOLUTION WAS SELECTED FOR PULMONARY VEIN ISOLATION. DURING THE PROCEDURE, IT WAS MENTIONED THAT THE PHYSICIAN HAD DIFFICULT DURING ATTEMPT TO GO TRANSEPTAL USING VERSACROSS WIRE THROUGH NON-BOSTON SCIENTIFIC VIZIGO SHEATH AND DILATOR). THE BMC RF GENERATOR HAD DELIVERED THE THERAPY, BUT NO BUBBLES WERE NOTED IN LEFT ATRIUM (LA) AND WIRE WAS NOT ABLE TO ADVANCE. IT WAS ATTEMPTED TO REPOSITIONING A FEW TIMES AND ENSURED UNDER FLUORO THAT TIP OF THE VERSACROSS RF WIRE WAS EXPOSED AND IN CONTACT WITH SEPTUM. THUS, THE NEW VERSACROSS KIT WAS USED, AND STILL THE ISSUE PERSISTED. LATER, THE PROCEDURE WAS COMPLETED USING A NRG RF NEEDLE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS WERE REPORTED. THE DEVICE IS NOT EXPECTED TO BE RETURN FOR ANALYSIS (DISPOSED).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21106959 ↗MalfunctionJan 8, 2025Failure to Advance; Failure to Advance
IT WAS REPORTED THAT VERSACROSS ACCESS SOLUTION WAS SELECTED FOR PULMONARY VEIN ISOLATION. DURING THE PROCEDURE, IT WAS MENTIONED THAT THE PHYSICIAN HAD DIFFICULT DURING ATTEMPT TO GO TRANSEPTAL USING VERSACROSS WIRE THROUGH NON-BOSTON SCIENTIFIC VIZIGO SHEATH AND DILATOR). THE BMC RF GENERATOR HAD DELIVERED THE THERAPY, BUT NO BUBBLES WERE NOTED IN LEFT ATRIUM (LA) AND WIRE WAS NOT ABLE TO ADVANCE. IT WAS ATTEMPTED TO REPOSITIONING A FEW TIMES AND ENSURED UNDER FLUORO THAT TIP OF THE VERSACROSS RF WIRE WAS EXPOSED AND IN CONTACT WITH SEPTUM. THUS, THE NEW VERSACROSS KIT WAS USED, AND STILL THE ISSUE PERSISTED. LATER, THE PROCEDURE WAS COMPLETED USING A NRG RF NEEDLE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS WERE REPORTED. THE DEVICE IS NOT EXPECTED TO BE RETURN FOR ANALYSIS (DISPOSED).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21106941 ↗InjuryDec 4, 2024Display or Visual Feedback Problem; Difficult to Remove; Display or Visual Feedback Problem; Difficult to Remove
IT WAS REPORTED A CARDIAC PERFORATION LEADING TO PERICARDIAL EFFUSION (PE) OCCURRED. THE PROCEDURE WAS CANCELLED. DURING AN ATRIAL FIBRILLATION ABLATION PROCEDURE, A VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE. THE PHYSICIAN WENT UP WITH THE VERSACROSS RF WIRE WITH INTRACARDIAC ECHOCARDIOGRAPHY (ICE) IMAGING ONLY. THE RF WIRE WENT UP WAY HIGH AND COULD NOT BE VISUALIZED ON ICE. THE WIRE WAS ATTEMPTED TO BRING DOWN TO SUPERIOR VENA CAVA (SVC) HOWEVER DIFFICULTY WAS EXPERIENCED. THE PATIENT BLOOD PRESSURE TANKED TO 36/21 AND A LARGE EFFUSION WAS SEEN ON ICE. THREE PERICARDIOCENTESIS KITS WERE USED WITH LITTLE CHANGE TO EFFUSION. THE PATIENT DID GO INTO CARDIAC ARREST AND SUBSEQUENT ACTIONS WERE TAKEN TO STABILIZE THE PATIENT. OPEN STERNOTOMY WAS PERFORMED TO SUTURE THE PERFORATION. A PERFORATION WAS FOUND IN THE RIGHT VENTRICLE AND IT WAS SUTURED AND A PATCH PLACED OVER TO SECURE THE INJURY. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS (DISPOSED). THE PATIENT STABILIZED AFTER CT SURGERY. THE PERFORATION WAS A RESULT OF THE ICE CATHETER GOING THROUGH THE RIGHT VENTRICLE. THE VERSACROSS WIRE DID NOT RELATE TO THE CAUSE OF INCIDENT. THE PHYSICIAN BELIEVES THAT WHILE MA
Patient problems: Cardiac Arrest; Cardiac Perforation; Low Blood Pressure/ Hypotension; Pericardial Effusion
MAUDE report 20846169 ↗
FDA clearance and classification
The FDA regulates product code DXF (catheter, septostomy) under 21 CFR 870.5175 as a Class II device. 21 CFR 870.5175 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the VersaCross Access Solution VXSK0021
Has the VersaCross Access Solution VXSK0021 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00685447017240), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the VersaCross Access Solution VXSK0021?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Dec 4, 2024 and Apr 22, 2025. In those reports, the most common specific device problems are “Failure to Advance” (2), “Contamination” (1) and “Difficult to Advance” (1); the most common patient problems are “Low Blood Pressure/ Hypotension” (3), “Pericardial Effusion” (3) and “Thrombosis/Thrombus” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VersaCross Access Solution VXSK0021?
None of the reports are classified as deaths, 3 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the VersaCross Access Solution VXSK0021, and is it still sold?
The FDA's device database lists Baylis Médicale Cie Inc as the company responsible for it (brand VersaCross Access Solution, model VXSK0021). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, Septostomy devices
- VersaCross Access Solution VXSK0031Baylis Médicale Cie Inc · Catheter, Septostomy1 recall16 reports
- VersaCross Access Solution VXSK0021Baylis Médicale Cie Inc · Catheter, Septostomy1 recall1 death15 reports
- VersaCross Access Solution VXSK0032Baylis Médicale Cie Inc · Catheter, Septostomy1 recall14 reports
- VersaCross Access Solution VXSK0022Baylis Médicale Cie Inc · Catheter, Septostomy1 recall13 reports
- VersaCross Access Solution VXSK0024Baylis Médicale Cie Inc · Catheter, Septostomy1 recall3 reports
- VersaCross Access Solution VXSK0034Baylis Médicale Cie Inc · Catheter, Septostomy1 recall1 report
- VersaCross Access Solution VXSK0025Baylis Médicale Cie Inc · Catheter, Septostomy1 recall1 report
- NRG Transseptal Needle NRG-E-HF-71-C0Boston Scientific Corporation · Catheter, Septostomy22 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page