VersaCross Access Solution VXSK0033 (UDI 00685447017363)Recalls & reported problems
Baylis Médicale Cie Inc · 1 x VersaCross Transseptal Sheath (including 1 x J-tip Guidewire); 1 x VersaCross RF Wire (including 1 x Connector Cable (Single Use)) · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00685447017363), received between Jun 9, 2025 and Jun 9, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Positioning Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Positioning Problem1
Patient problems
- Pericardial Effusion2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJun 9, 2025Positioning Problem; Positioning Problem
IT WAS REPORTED A PERICARDIAL EFFUSION (PE) WAS NOTED. DURING A LEFT ATRIAL APPENDAGE CLOSURE (LAAC) PROCEDURE, A VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. THERE WAS A TRACE EFFUSION NOTED BEFORE THE PROCEDURE STARTED AND IT WAS LOCATED AT THE APEX OF THE HEART. THE VERSACROSS WIRE WAS INSERTED IN THE SUPERIOR VENA CAVA (SVC), WITH THE SHEATH TO FOLLOW. THE PHYSICIAN BROUGHT THE WHOLE SYSTEM DOWN ONTO THE SEPTUM IN ORDER TO GET IN THE RIGHT POSITION TO CROSS AND GO INTO THE LEFT ATRIUM (LA). IN THE PROCESS OF TRYING TO GET THE CORRECT PLACEMENT ON THE SEPTUM THE ACCESS SYSTEM WAS PLACED IN THE RIGHT ATRIAL APPENDAGE AND THEN DROPPED ONTO THE SEPTUM. THE PHYSICIAN WENT TO CROSS THE SEPTUM AND HEAD INTO THE LEFT ATRIUM. ONCE ACROSS THE LEFT ATRIUM, THE RF WIRE WAS INSERTED INTO THE LEFT ATRIAL APPENDAGE AND THE VERSACROSS SYSTEM WAS TAKEN OUT, NEXT THE WATCHMAN DOUBLE CURVE SHEATH WAS INSERTED. ONCE THE DOUBLE CURVE WAS INSERTED THE NON-BOSTON SCIENTIFIC PIGTAIL CATHETER OVER THE RF WIRE AND THE WIRE WAS TAKEN OUT. LA PRESSURE WAS OBTAINED AT THIS POINT, AND A CONTRAST SHOT WAS TAKEN OF THE LEFT ATRIAL APPENDAGE TO DETERMINE WHAT SIZE WATCHMAN DEVICE. THE DEVICE WAS PREPPED, T
Patient problems: Pericardial Effusion
MAUDE report 22170061 ↗InjuryJun 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED A PERICARDIAL EFFUSION (PE) WAS NOTED. DURING A LEFT ATRIAL APPENDAGE CLOSURE (LAAC) PROCEDURE, A VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. THERE WAS A TRACE EFFUSION NOTED BEFORE THE PROCEDURE STARTED AND IT WAS LOCATED AT THE APEX OF THE HEART. THE VERSACROSS WIRE WAS INSERTED IN THE SUPERIOR VENA CAVA (SVC), WITH THE SHEATH TO FOLLOW. THE PHYSICIAN BROUGHT THE WHOLE SYSTEM DOWN ONTO THE SEPTUM IN ORDER TO GET IN THE RIGHT POSITION TO CROSS AND GO INTO THE LEFT ATRIUM (LA). IN THE PROCESS OF TRYING TO GET THE CORRECT PLACEMENT ON THE SEPTUM THE ACCESS SYSTEM WAS PLACED IN THE RIGHT ATRIAL APPENDAGE AND THEN DROPPED ONTO THE SEPTUM. THE PHYSICIAN WENT TO CROSS THE SEPTUM AND HEAD INTO THE LEFT ATRIUM. ONCE ACROSS THE LEFT ATRIUM, THE RF WIRE WAS INSERTED INTO THE LEFT ATRIAL APPENDAGE AND THE VERSACROSS SYSTEM WAS TAKEN OUT, NEXT THE WATCHMAN DOUBLE CURVE SHEATH WAS INSERTED. ONCE THE DOUBLE CURVE WAS INSERTED THE NON-BOSTON SCIENTIFIC PIGTAIL CATHETER OVER THE RF WIRE AND THE WIRE WAS TAKEN OUT. LA PRESSURE WAS OBTAINED AT THIS POINT, AND A CONTRAST SHOT WAS TAKEN OF THE LEFT ATRIAL APPENDAGE TO DETERMINE WHAT SIZE WATCHMAN DEVICE. THE DEVICE WAS PREPPED, T
Patient problems: Pericardial Effusion
MAUDE report 22170008 ↗
Other Catheter, Septostomy devices
- VersaCross Access Solution VXSK0031Baylis Médicale Cie Inc · Catheter, Septostomy1 recall16 reports
- VersaCross Access Solution VXSK0021Baylis Médicale Cie Inc · Catheter, Septostomy1 recall1 death15 reports
- VersaCross Access Solution VXSK0032Baylis Médicale Cie Inc · Catheter, Septostomy1 recall14 reports
- VersaCross Access Solution VXSK0022Baylis Médicale Cie Inc · Catheter, Septostomy1 recall13 reports
- VersaCross Access Solution VXSK0024Baylis Médicale Cie Inc · Catheter, Septostomy1 recall3 reports
- VersaCross Access Solution VXSK0034Baylis Médicale Cie Inc · Catheter, Septostomy1 recall1 report
- VersaCross Access Solution VXSK0025Baylis Médicale Cie Inc · Catheter, Septostomy1 recall1 report
- NRG Transseptal Needle NRG-E-HF-71-C0Boston Scientific Corporation · Catheter, Septostomy22 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.