Fine Osteotomy system - Distal femur 015-330-007Recalls & reported problems
Laboratoires Bodycad Inc · Femoral plate , Cortical Locking Screws, Cortical Compression Screws and disposable instrument · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00688346002140), received between Nov 14, 2022 and Oct 17, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Migration” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Migration2
- Material Fragmentation1
- No Apparent Adverse Event1
Patient problems
- Bone Fracture(s)1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 17, 2023Material Fragmentation; No Apparent Adverse Event; Migration; Material Fragmentation; No Apparent Adverse Event; Migration
THE PATIENT UNDERWENT A LATERAL OPENING WEDGE DISTAL FEMUR OSTEOTOMY (DFO) TO ADDRESS A VALGUS DEFORMITY AROUND THE KNEE. ADDITIONALLY, THE PATIENT RECEIVED A LATERAL FEMORAL CONDYLE OSTEOCHONDRAL ALLOGRAFT. THE DFO PROCEDURE WAS PERFORMED IN THE STANDARD MANNER, AND ALTHOUGH THE MEDIAL HINGE CRACKED, IT WAS NOT STABILIZED. THE DFO OSTEOTOMY WAS SECURED WITH A BODYCAD PERSONALIZED PLATE, AND THE GAP WAS FILLED WITH A BONE GRAFT. THE INITIAL POST-OPERATIVE X-RAY SHOWED THAT THE PLATE WAS WELL-POSITIONED, AND THE SCREWS WERE OF THE APPROPRIATE LENGTH AS PLANNED.THE PATIENT RETURNED FOR A ROUTINE CHECKUP SIX WEEKS POST-OP, WHERE IT WAS NOTED THAT ONE OF THE DISTAL SCREWS (D) HAD SLIGHTLY BACKED OUT, AND ANOTHER SCREW (B) WAS SLIGHTLY BENT. HOWEVER, THE PLATE REMAINED STABLE, AND THE MICRO-FRACTURE AT THE HINGE SITE WAS SHOWING GOOD CONSOLIDATION. THERE WAS EXCELLENT BONE HEALING AND PROPER CONSOLIDATION OF THE OSTEOTOMY, WITH NO LOSS OF CORRECTION.AT THE SECOND ROUTINE CHECKUP, 12 WEEKS POST-OP, NO SIGNS OF INFECTION WERE OBSERVED. TWO SCREWS (C AND D) HAD BACKED OUT SLIGHTLY BUT REMAINED STABLE AND HAD NOT MOVED BEYOND THE LOCKING MECHANISM (3 MM). ALL OTHER SCREWS' THREADS WERE WITH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17950052 ↗MalfunctionNov 14, 2022Migration; Migration
SCREWS BACKING UP. DURING A ROUTINE APPOINTMENT, THE DOCTOR NOTICED ON THE X-RAY THAT 3 SCREWS WERE MIGRATING. NO INFECTION IS PRESENT. NOTE ALSO THAT WE SEE A FRACTURE ON THE OPPOSITE SIDE OF THE CORTEX. WE SEE A SMALL FRACTURE INITIATION ON THE POST-OP PHOTO AND IT SEEMS TO HAVE WORSENED ON THE 6 WEEK POST-OP PHOTO.
Patient problems: Bone Fracture(s)
MAUDE report 15792648 ↗
Other Plate, Fixation, Bone devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-MLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-MRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page