Skip to content

SWAN-GANZ 131F7 (REF 131F7P)Recalls & reported problems

Edwards Lifesciences LLC · SWAN-GANZ THERMODILUTION CATHETER · Class 2

All models of SWAN-GANZ (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00690103000412), received between Nov 29, 2024 and Jan 8, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Failure to Deflate” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K160084, K231248); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 8, 2026Open, Classified

    Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.

    Z-2202-2026 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Failure to Deflate1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 8, 2026Material Fragmentation

    AS REPORTED DURING USE THE SWAN GANZ CATHETER WAS TESTED BEFORE A CASE AND THE BALLOON INFLATED. IT INFLATED IN THE BODY AS WELL BUT ONCE IT WAS REMOVED THE BALLOON HAD ALREADY DEFLATED WITHOUT THE DOCTOR DOING SO., THEY TRIED TO INFLATE THE BALLOON AGAIN OUTSIDE THE BODY AND IT WOULD NOT INFLATE. THERE WAS NO PATIENT INJURY. THE DEVICE IS AVAILABLE FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24015691 ↗
  • MalfunctionNov 29, 2024Appropriate Device Problem Term/Code Not Available; Failure to Deflate; Appropriate Device Problem Term/Code Not Available; Failure to Deflate

    AS REPORTED, DURING DEVICE PREPARATION, THE BALLOON OF THIS SWAN-GANZ CATHETER DID NOT DEFLATE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20810716 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K160084, K231248. One submission often covers many models, so check the product and lot details in each record.

  • Z-2203-2026 · Apr 8, 2026 · Open, Classified
    Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P
  • Z-0765-2019 · Dec 21, 2018 · Terminated
    Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 151F7 Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determi
  • Z-0762-2019 · Dec 21, 2018 · Terminated
    Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF131F 7J Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determ
  • Z-0764-2019 · Dec 21, 2018 · Terminated
    Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131VF7P Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly deter
  • Z-0763-2019 · Dec 21, 2018 · Terminated
    Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians

Other Catheter, Intravascular, Diagnostic devices

All Catheter, Intravascular, Diagnostic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page