Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00690103000566), received between Oct 27, 2020 and Jun 29, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Deflation Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
4 other recalls cover devices cleared under the same FDA submission (K160084); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Apr 8, 2026Class IIOpen, Classified
Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
Z-2202-2026 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Deflation Problem1
- Difficult to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 29, 2023Deflation Problem; Deflation Problem
IT WAS REPORTED THAT A 151F7 SWAN-GANZ CATHETER FROM LOT 64830010 HAD A BALLOON FAILURE. THE PATIENT CAME TO CCU FROM THE CATH LAB, WHERE THE CATHETER WAS INSERTED. UPON ARRIVAL TO CCU, THE RN ATTEMPTED TO WEDGE THE BALLOON WITH 1.5CC OF AIR TO RECORD A WEDGE PRESSURE. ACCORDING TO THE RN, THE DEVICE WOULD NOT WEDGE AS INDICATED BY NO CHANGE IN WAVEFORM WHILE ATTEMPTING THE WEDGE. THE CATHETER BALLOON WOULD ALSO NOT PASSIVELY DEFLATE WHEN THE SYRINGE PLUNGER WAS RELEASED. THERE WAS NO RESISTANCE ON THE SYRINGE PLUNGER WHEN THE PLUNGER WAS RELEASED. THERE WAS NO PATIENT INJURY. THE DEVICE IS NOT AVAILABLE FOR RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17232003 ↗MalfunctionOct 27, 2020Difficult to Advance
ELDERLY MALE WITH HISTORY OF CORONARY ARTERY DISEASE AND HYPERTENSION, RECENT VENTRICULAR TACHYCARDIA AND VENTRICULAR FIBRILLATION. WHILE HAVING A HEART CATH, THE PERSON INSERTING THE SWAN GANZ HAD PROBLEMS THREADING THE 0.25 WIRE. IT WAS REMOVED FROM THE PATIENT WITHOUT KNOWN HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10739847 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K160084. One submission often covers many models, so check the product and lot details in each record.
- Z-0765-2019 · Dec 21, 2018 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF 151F7 Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly dete
- Z-0764-2019 · Dec 21, 2018 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131VF7P Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly de
- Z-0763-2019 · Dec 21, 2018 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians
- Z-0762-2019 · Dec 21, 2018 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF131F 7J Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly det
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H749085263001Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.