Fem-Flex II Femoral Arterial Cannula FEMII016ARecalls & reported problems
Edwards Lifesciences LLC · FEM-FLEX II Femoral Arterial Cannula · Class 2
All models of Fem-Flex II Femoral Arterial Cannula (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00690103031232), received between Dec 22, 2025 and Jan 20, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Mechanical Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
6 other recalls cover devices cleared under the same FDA submission (K140208, K200358); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 14, 2025Class IOpen, Classified
due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.
Z-2082-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Mechanical Problem1
Patient problems
- Cardiac Arrest2
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 20, 2026
PATIENT HAD BILATERAL LUNG TRANSPLANT AND REQUIRED VV ECMO [VENO-VENOUS EXTRACORPOREAL MEMBRANE OXYGENATION]. AROUND 12 HOURS LATER, THE CANULA WAS FOUND TO BE CRACKED AND THE PATIENT NEEDED TO GO BACK TO THE OR FOR EXCHANGE LEADING TO A CARDIAC ARREST AND CONVERSION TO CENTRAL ECMO.
Patient problems: Cardiac Arrest
MAUDE report 24114649 ↗InjuryDec 22, 2025Mechanical Problem; Mechanical Problem
EDWARDS RECEIVED A REPORT OF AN EDWARDS FEMII016A CANNULA THAT WAS FOUND TO BE LEAKING DURING USE. THE CANNULA WAS INSERTED AFTER A LUNG TRANSPLANT FOR V-V ECMO INITIATION. BLEEDING FROM THE CANNULA WAS NOTED THE NEXT DAY. THIS LED TO AN EXCHANGE OF CANNULA AND BLOOD LOSS FOR THE PATIENT. THE ESTIMATED TIME THE DEVICE WAS IN USE WAS 27 HOURS (21 HOURS PAST THE 6 HOURS IN THE IFU). THEY ALSO NOTED THEY HAD TO BRING THE PATIENT BACK TO THE OR THE NEXT DAY. IT APPEARS THERE WAS A CRACK/ FRACTURE IN THE LINE, AND BLOOD WAS COMING OUT OF THE LINE. THE CANNULA WAS PLACED, AND THE CRACK WAS NOT FOUND UNTIL A LATER TIME AFTER THE PACKAGING WAS DISPOSED OF. THEREFORE, THE LOT NUMBER IS UNKNOWN. THE CANNULA IS BEING HELD BY RISK MANAGEMENT AND THE INVESTIGATIVE LEAD HAS CONFIRMED NO FURTHER INFORMATION WILL BE SHARED OUTSIDE OF AN OFFICIAL LEGAL REQUEST. DEVICE NOT RETURNED. NO IMAGES PROVIDED. WHEN ASKED ABOUT THE TECHNIQUE USED FOR PLACEMENT/ SET UP FOR ECMO, THE PERFUSION DIRECTOR PROVIDED THE FOLLOWING: VV ECMO - DRAINAGE 23FR BIOMEDICUS FEM VEIN, REINFUSION - 16F FEM FLEX LEFT SUBCLAVIAN VEIN. THEY ALSO STATED THAT THE CANNULA WAS NOT AT AN ANGLE/ CURVED DURING USE. THE DEVICE WAS INSER
Patient problems: Cardiac Arrest; Hemorrhage/Blood Loss/Bleeding
MAUDE report 23871998 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K140208, K200358. One submission often covers many models, so check the product and lot details in each record.
- Z-2083-2025 · May 14, 2025 · Open, ClassifiedBrand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016AS Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion C
- Z-2084-2025 · May 14, 2025 · Open, ClassifiedBrand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018A Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion Ca
- Z-2085-2025 · May 14, 2025 · Open, ClassifiedBrand Name: FEM-FLEXII Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018AS Software Version: N/A Product Description: The Edwards FEMFLEXII Arterial Perfusion C
- Z-3033-2018 · Mar 7, 2018 · TerminatedEdwards Lifesciences Femoral Venous Cannula, 20F X 55CM, resterilized.
- Z-3032-2018 · Mar 7, 2018 · TerminatedEdwards Lifesciences Femoral Venous Cannula, 18FR, resterilized.
- Z-1371-2015 · Mar 23, 2015 · TerminatedEdwards Lifesciences Fem-Flex II Femoral Arterial Cannula 8, 10, 12 French,Sterile EO, Rx Only, Manufacturer Edwards Lifesciences LLC, Irvine, CA.-Model Numbers: FEMII008A, FEMII0
Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY2V10R33Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page