Fem-Flex II Femoral Arterial Cannula recallZ-2082-2025 · Edwards Lifesciences, LLC · initiated May 14, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Edwards Lifesciences, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.
FDA root cause category: Under Investigation by firm
What the company is doing
On 05/16/2025, and 05/19/2025, the firm sent via FedEx overnight an "URGENT: PRODUCT RECALL" Letter to customers informing them that Edwards has confirmed a small number of occurrences impacting the OptiSite Arterial Perfusion cannula in which a 3mm to 4mm section of wire, from the wire-reinforcement coil at the cannula tip, was found to be released from the cannula body. The risk for the wire-reinforcement coil protruding out of the cannula body is major tissue damage and hemolysis. Customers are instructed to: " Verify their inventory on the attached customer acknowledgement form. " Share this notice with appropriate clinical staff at their site. " No patient follow-up or notification is necessary. " Return a completed Customer Acknowledgment Form to Edwards eCV Representative or via email to Edwards Customer Service at [email protected] within 15 days of receipt of this notification. " After receiving RGA number, return any impacted product to Edwards at the address below. A credit will be issued upon receipt of returned product. enableCV Distribution Center 6644 W. 2100 S. Suite D West Valley, UT 84128 For questions/assistance contact - Customer Service at 1-888-943-2783, or email: [email protected]
Recall details
- Recall number
- Z-2082-2025
- Status
- Open, Classified
- Initiated by the company
- May 14, 2025
- Posted by the FDA
- Jul 10, 2025
- Recalling firm
- Edwards Lifesciences, LLC
- Quantity
- 6801
- Product code
- DQR
- FDA submissions
- K140208
- Distribution
- Worldwide - US Nationwide distribution including in the states of ALABAMA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, HAWAII, IDAHO, ILLINOIS, INDIANA, IOWA, KANSAS, KENTUCKY, LOUISIANA, MAINE, MARYLAND, MASSACHUSETTS, MICHIGAN, MINNESOTA, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEVADA, NEW HAMPSHIRE, NEW JERSEY, NEW MEXICO, NEW YORK, NORTH CAROLINA, NORTH DAKOTA, OHIO, OKLAHOMA, OREGON, PENNSYLVANIA, Puerto Rico, RHODE ISLAND, SOUTH CAROLINA, SOUTH DAKOTA, TENNESSEE, TEXAS, UTAH, VIRGINIA, WASHINGTON, WEST VIRGINIA, WISCONSIN and the countries of Albania, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, CANARY ISLANDS, Chile, Colombia, Cyprus, Czech Republic, Denmark, France, Germany, Greece, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Malta, Netherlands, Northern Ireland, Norway, Pakistan, Panama, Poland, Portugal, Romania, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates (UAE), United Kingdom.
Product description
Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016A Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No
Code information (models, lots, serial numbers)
Lot Code: Model No: FEMII016A; UDI: 00690103031232; All lots up to: BSLC8179 Serial Numbers: 130544 130757 131105 131313 131511 131535 131888 131925 132504 BSLC0184 BSLC0386 BSLC0596 BSLC1311 BSLC3741 BSLC4162 BSLC5040 BSLC6487 BSLC8179
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Fem-Flex II Femoral Arterial Cannula FEMII016AEdwards Lifesciences LLC · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS1 recall2 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K140208 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Fem-Flex II Femoral Arterial Cannula FEMII018ASEdwards Lifesciences LLC · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS1 recall0 reports
- Fem-Flex II Femoral Arterial Cannula FEMII020AEdwards Lifesciences LLC · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS1 report
- Edwards Lifesciences Femoral Venous Cannula VFEM018Edwards Lifesciences LLC · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS1 report
- Edwards Lifesciences Femoral Venous Cannula VFEM022Edwards Lifesciences LLC · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 96943).
- Z-2080-2025 · Open, ClassifiedBrand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI16 Software Version: N/A Product Description: Cardiac cannula. The Edwards O…
- Z-2081-2025 · Open, ClassifiedBrand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI18 Software Version: N/A Product Description: Cardiac cannula. The Edwards O…
- Z-2083-2025 · Open, ClassifiedBrand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016AS Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion C…
- Z-2084-2025 · Open, ClassifiedBrand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018A Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion Ca…
- Z-2085-2025 · Open, ClassifiedBrand Name: FEM-FLEXII Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018AS Software Version: N/A Product Description: The Edwards FEMFLEXII Arterial Perfusion C…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page