FOGARTY 120404FRecalls & reported problems

Edwards Lifesciences LLC · FOGARTY ARTERIAL EMBOLECTOMY CATHETER 4F 40CM · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00690103043266), received between Jun 19, 2024 and Jan 10, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Break1
  • Deflation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 10, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT BEFORE USE DURING THE INFLATION TEST, THE TIP OF THIS FOGARTY CATHETER WAS DETACHED WHEN THE BALLOON WAS INFLATED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21126728 ↗
  • MalfunctionOct 11, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE FOGARTY CATHETER WAS BROKEN DURING USE. THE DETAILED INFORMATION ON THE LOCATION AND CONDITION OF THE DAMAGE WERE NOT AVAILABLE. NO ADDITIONAL INTERVENTION SUCH AS ADDITIONAL INCISION WAS PERFORMED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20433849 ↗
  • MalfunctionJun 19, 2024Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT THE BALLOON OF A VASCULAR CATHETER DID NOT DEFLATE. STAFF HAD TO PUNCTURE THE BALLOON THROUGH THE SKIN TO DEFLATE BALLOON. THERE WERE NO PATIENT INJURIES. A PRODUCT EVALUATION WAS COMPLETED ON THE RETURNED 120404F CATHETER. THE CUSTOMER REPORTED OF "DID NOT DEFLATE" WAS UNABLE TO BE CONFIRMED. AS RECEIVED, BALLOON WAS RECEIVED RUPTURED. BALLOON LATEX WAS RELEASED FROM PROXIMAL WINDING TO CHECK BALLOON EDGES. THE EDGES DID NOT APPEAR TO MATCH AT THE RUPTURE. THE PROXIMAL WINDING HAD PARTIALLY UNRAVELED AND SLIPPED DISTAL ON THE CATHETER TIP. BOTH BALLOON AND PART OF THE PROXIMAL WINDING HAD SLIPPED DISTAL OF THE INFLATION PORTS. BALLOON MAY BE UNABLE TO DEFLATE WHEN SLIPPED DISTAL OF THE INFLATION PORTS. DISTAL WINDING WAS INTACT. NO VISIBLE DAMAGE WAS OBSERVED FROM CATHETER BODY. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED AND DOCUMENTED THAT DEVICE MET ALL SPECIFICATIONS UPON DISTRIBUTION. AN ENGINEERING EVALUATION WAS INITIATED TO ASSESS FOR ANY MANUFACTURING RELATED PROCESSES WHICH COULD BE CORRELATED TO THE COMPLAINT. BASED ON THE AVAILABLE INFORMATION, THERE IS NO EVIDENCE THAT SUPPORTS OR CONFIRMS THE FAILURE MODE IS ASSOCIATED TO A MANUFACTURING OR DESI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19567063 ↗