Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00690103043280), received between Mar 3, 2020 and Sep 19, 2025 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel3
  • Break1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body Embolism1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 19, 2025Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, DURING TESTING OF THIS FOGARTY EMBOLECTOMY CATHETER, AN AIR LEAKAGE FROM THE BALLOON WAS NOTICED. WHEN TESTING AGAIN, THE BALLOON BURST. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS RECEIVED FOR EVALUATION AND THE SPRING TIP WIRE WAS VISIBLE. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23093707 ↗
  • MalfunctionMay 28, 2025Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT, DURING USE, WHEN WITHDRAWING THIS FOGARTY CATHETER FOR THE SECOND TIME, THE BALLOON BURST. THERE WAS NO PATIENT INJURY. THE DEVICE WILL BE AVAILABLE FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22102451 ↗
  • InjuryApr 30, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, WHEN TRAWLING DURING A POPLITEAL ARTERY EMBOLECTOMY OF THIS FOGARTY EMBOLECTOMY CATHETER, THE BALLOON DISLODGED IN THE VESSEL. THE BALLOON COULD NOT BE RETRIEVED USING A NEW FOGARTY EMBOLECTOMY CATHETER.

    Patient problems: Foreign Body Embolism

    MAUDE report 21933800 ↗
  • MalfunctionSep 24, 2024Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT THE FOGARTY ARTERIAL EMBOLECTOMY 120803F BALLOON TIP BURST IN A PATIENT DURING A PROCEDURE. NO PATIENT INJURY WAS REPORTED AND NO DEMOGRAPHICS WERE AVAILABLE. THERE WILL BE NO PRODUCT RETURN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20297195 ↗
  • MalfunctionMar 3, 2020Break

    WE ARE PERFORMING A FEMORAL POPLITEAL THROMBECTOMY AND USING #3 FOGARTY CATHETER, THE TIP OF THE CATHETER BROKE OFF IN THE PATIENT'S ARTERY.