The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00690103043303), received between Feb 14, 2019 and Aug 20, 2024 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component5
- Break2
- Failure to Deflate1
Patient problems
- Foreign Body In Patient2
- No Clinical Signs, Symptoms or Conditions2
- Device Embedded In Tissue or Plaque1
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
- Vascular Dissection1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryAug 20, 2024Failure to Deflate; Failure to Deflate
AS REPORTED, WHEN INSERTING THIS FOGARTY EMBOLECTOMY CATHETER DURING AN ENDOVASCULAR SURGERY, AFTER INFLATING THE BALLOON, THE BALLOON SLIPPED TO THE TIP OF THE CATHETER AND COULD NOT BE DEFLATED. THERE WAS A DISSECTION OF AN ARTERY AND THE INTERVENTION WAS PROLONGED. NO FURTHER INFORMATION PROVIDED. PATIENT DEMOGRAPHICS WERE UNABLE TO BE OBTAINED.
Patient problems: Insufficient Information; Vascular Dissection
MAUDE report 20007062 ↗MalfunctionAug 14, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE PATIENT WAS IN THE OR FOR A RIGHT AXILLARY BILATERAL FEMORAL THROMBECTOMY WITH ANGIOGRAM. A 4FR FOGARTY CATHETER WAS USED DURING ANGIOGRAM. THE PROVIDER PLACED THE FOGARTY CATHETER IN PATIENTS RIGHT LEG AND INFLATED THE BALLOON AS INDICATED ON THE PACKAGING AND CONFIRMED WITH X-RAY. WHEN THE CATHETER WAS REMOVED IT WAS DISCOVERED THAT THE BALLOON WAS NOT INTACT ON THE CATHETER AND HAD BEEN DISPLACED INSIDE OF THE PATIENT. PROVIDER STATED THAT THE FIRST BALLOON GOT HUNG UP ON WHAT SHE BELIEVES TO BE THE CHRONIC OCCLUDED PLAQUE. SHE HAD TO PULL HARD AND THE BALLOON BROKE AWAY FROM THE CATHETER AND WAS LEFT BEHIND. SHE THEN USED A SECOND BALLOON TO RETRIEVE THE FIRST AND THE EXACT SAME THING HAPPENED SO SHE DID NOT TRY AGAIN. TWO BALLOONS WERE LEFT BEHIND IN THE PATIENT. PROVIDER ALSO STATED THAT IN HER OPINION THE BALLOONS ARE NOT CAUSING ANY OCCLUSION BECAUSE THE VESSEL WAS ALREADY OCCLUDED. THERE WAS NO ALLEGATION OF PATIENT INJURY. DURING SET UP, THE INFLATION TEST WAS COMPLETED WITHOUT ISSUE. THE TWO BALLOONS ARE STILL INSIDE THE PATIENT. THIS COMPLAINT CAPTURES THE SECOND FOGARTY CATHETER USED IN THIS CASE. DEVICE WAS DISCARDED AFTER USE.
Patient problems: Foreign Body In Patient
MAUDE report 19986719 ↗MalfunctionAug 14, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE PATIENT WAS IN THE OR FOR A RIGHT AXILLARY BILATERAL FEMORAL THROMBECTOMY WITH ANGIOGRAM. A 4FR FOGARTY CATHETER WAS USED DURING ANGIOGRAM. THE PROVIDER PLACED THE FOGARTY CATHETER IN PATIENTS RIGHT LEG AND INFLATED THE BALLOON AS INDICATED ON THE PACKAGING AND CONFIRMED WITH X-RAY. WHEN THE CATHETER WAS REMOVED IT WAS DISCOVERED THAT THE BALLOON WAS NOT INTACT ON THE CATHETER AND HAD BEEN DISPLACED INSIDE OF THE PATIENT. PROVIDER STATED THAT THE FIRST BALLOON GOT HUNG UP ON WHAT SHE BELIEVES TO BE THE CHRONIC OCCLUDED PLAQUE. SHE HAD TO PULL HARD AND THE BALLOON BROKE AWAY FROM THE CATHETER AND WAS LEFT BEHIND. SHE THEN USED A SECOND BALLOON TO RETRIEVE THE FIRST AND THE EXACT SAME THING HAPPENED SO SHE DID NOT TRY AGAIN. TWO BALLOONS WERE LEFT BEHIND IN THE PATIENT. PROVIDER ALSO STATED THAT IN HER OPINION THE BALLOONS ARE NOT CAUSING ANY OCCLUSION BECAUSE THE VESSEL WAS ALREADY OCCLUDED. THERE WAS NO ALLEGATION OF PATIENT INJURY. DURING SET UP, THE INFLATION TEST WAS COMPLETED WITHOUT ISSUE. THE TWO BALLOONS ARE STILL INSIDE THE PATIENT. THIS COMPLAINT CAPTURES THE FIRST FOGARTY CATHETER USED IN THIS CASE. DEVICE WAS DISCARDED AFTER USE.
Patient problems: Foreign Body In Patient
MAUDE report 19986064 ↗MalfunctionJul 23, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, WHEN BRINGING THIS FOGARTY CATHETER OUT FROM PATIENT, WHICH WAS USED FOR A BLOCKED FEMORAL ARTERY, THE BALLOON DETACHED FROM THE CATHETER, LEAVING IT INSIDE THE PATIENT. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19803813 ↗MalfunctionJun 28, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN AXILLARY BIFEMORAL BYPASS PROCEDURE THE BALLOON BECAME DETACHED FROM A FOGARTY 4FR MODEL 120804F CATHETER AND WAS RETAINED IN THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A NEW FOGARTY CATHETER. PER THE CUSTOMER, IT WAS DETERMINED THAT NO ADDITIONAL PROCEDURE WAS PERFORMED BASED ON RISK TO THE PATIENT. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE IS NOT AVAILABLE FOR RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19635784 ↗MalfunctionJun 12, 2020Break
FROM STAFF: WHILE USING A FOGARTY EMBOLECTOMY CATHETER THE TIP BROKE OFF IN THE PATIENT. THE TIP WAS RETRIEVED BY THE SURGEON.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10147895 ↗InjuryFeb 14, 2019Break; Detachment of Device or Device Component
THE BALLOON ON THE FOGARTY CATHETER BROKE AND REMAINED INSIDE THE PATIENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8343631 ↗
Other CATHETER, EMBOLECTOMY devices
- Indigo System CAT6Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths408 reports
- Indigo System CATRXKITPenumbra, Inc. · Catheter, Embolectomy1 death339 reports
- Fogarty 120804FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall1 death200 reports
- LeMaitre Embolectomy Catheter 1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy1 recall161 reports
- FOGARTY 120404FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall61 reports
- Indigo System CAT6KIT-APenumbra, Inc. · Catheter, Embolectomy109 reports
- LeMaitre Embolectomy Catheter e1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy2 recalls4 reports
- Indigo System CAT8XTORQ115Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths98 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.