Skip to content

FOGARTY 12TLW403FRecalls & reported problems

Edwards Lifesciences LLC · FOGARTY THRU-LUMEN EMBOLECTOMY CATHETER 3F 40CM · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00690103043457), received between Nov 27, 2024 and Oct 20, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (4 reports).

1 other recall cover devices cleared under the same FDA submission (K240078, K892410); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel4
  • Fluid/Blood Leak1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 20, 2025Burst Container or Vessel; Fluid/Blood Leak; Material Fragmentation; Burst Container or Vessel; Fluid/Blood Leak; Material Fragmentation

    AS REPORTED, DURING PREOPERATIVE PROCESS, WHEN FILLING THE BALLOON OF THIS EMBOLECTOMY FOGARTY CATHETER, SALINE LEAKAGE WAS OBSERVED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. AS PER EVALUATION RESULTS, THE BALLOON WAS FOUND TO BE RUPTURED AT CENTRAL AREA. THE RUPTURED EDGES DID NOT APPEAR TO MATCH AT THE RUPTURED LOCATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23334307 ↗
  • MalfunctionMar 13, 2025Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT DURING USE WITH A FOGARTY THRU LUMEN EMBOLECTOMY VASCULAR CATHETER, WHILE THE USER WAS PULLING THE THROMBUS WITH THE INFLATED BALLOON, THE BALLOON BURST. THE TIP OF THE CATHETER BECAME BROKEN. AN ADDITIONAL INCISION WAS REQUIRED TO REMOVE THE FRAGMENTS. ALL THE FRAGMENTS COULD NOT BE COMPLETELY REMOVED. THERE WAS NO PATIENT INJURY OR COMPLICATIONS OBSERVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21597725 ↗
  • MalfunctionDec 9, 2024Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT THE BALLOON OF THE 12TLW403F FOGARTY CATHETER RUPTURED BEFORE USE. PATIENT DEMOGRAPHICS WERE REQUESTED BUT UNAVAILABLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20879562 ↗
  • MalfunctionNov 27, 2024Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT THE BALLOON OF THIS FOGARTY CATHETER RUPTURED DURING USE. PATIENT DEMOGRAPHICS WERE REQUESTED BUT UNAVAILABLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20794215 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K240078, K892410. One submission often covers many models, so check the product and lot details in each record.

  • Z-0238-2024 · Sep 7, 2023 · Open, Classified
    Fogarty Thru-Lumen Embolectomy Catheter, REF: 12TLW805F35

Other CATHETER, EMBOLECTOMY devices

All CATHETER, EMBOLECTOMY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.