Skip to content

FOGARTY 12TLW404FRecalls & reported problems

Edwards Lifesciences LLC · FOGARTY THRU-LUMEN EMBOLECTOMY CATHETER, 4F 40CM · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00690103043464), received between Feb 11, 2025 and Aug 26, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K240078, K892410); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel2
  • Detachment of Device or Device Component1
  • Failure to Deflate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2026Failure to Deflate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50454702 ↗
  • MalfunctionSep 11, 2025Burst Container or Vessel; Detachment of Device or Device Component; Burst Container or Vessel; Detachment of Device or Device Component

    AS REPORTED, DURING SURGERY, THE BALLOON OF THIS FOGARTY EMBOLECTOMY CATHETER BURST. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. AS PER PRODUCT EVALUATION FINDINGS, PART OF THE CATHETER TIP, BALLOON LATEX, DISTAL BUSHING, AND DISTAL WINDING WERE COMPLETELY DETACHED FROM THE CATHETER BODY. AS CONFIRMED, NO ADDITIONAL INTERVENTION WAS NEEDED IN ORDER TO RETRIEVE THE PIECES DETACHED AND THE PATIENT DID NOT SUFFER ANY HURT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23026557 ↗
  • MalfunctionFeb 11, 2025Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT THE BALLOON OF THIS FOGARTY CATHETER RUPTURED WHEN IT WAS INFLATED AND PULLED TO REMOVE THE BLOOD CLOT AT THE ANASTOMOTIC SITE DURING USE. PATIENT DEMOGRAPHICS WERE REQUESTED BUT UNAVAILABLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21356571 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K240078, K892410. One submission often covers many models, so check the product and lot details in each record.

  • Z-0238-2024 · Sep 7, 2023 · Open, Classified
    Fogarty Thru-Lumen Embolectomy Catheter, REF: 12TLW805F35

Other CATHETER, EMBOLECTOMY devices

All CATHETER, EMBOLECTOMY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.