FOGARTY 12TLW805F35Recalls & reported problems
Edwards Lifesciences LLC · FOGARTY THRU-LUMEN EMBOLECTOMY CATHETER 5.5F 80CM · Class 2
All models of FOGARTY (32)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This is an Edwards Lifesciences Fogarty embolectomy catheter, a thin tube with a small balloon used to clear clots from blood vessels. The 44 reports mostly describe the balloon bursting, cracking, or tearing during use, sometimes breaking into pieces, and balloons that will not inflate or deflate properly; one report describes a foreign body embolism and one a laceration. There are no deaths and 1 injury. One open recall, started in September 2023, covers catheters whose balloon deflates more slowly than normal.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 43
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 00690103043532), received between Oct 5, 2021 and May 22, 2026 — classified as 1 injury, 42 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (11 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Sep 7, 2023Class IIOpen, Classified
Inadvertent release of embolectomy catheters that may experience slower than normal balloon deflation (greater-than 15 seconds).
Z-0238-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Burst Container or Vessel11
- Material Fragmentation9
- Detachment of Device or Device Component6
- Material Split, Cut or Torn6
- Failure to Deflate5
- Crack3
- Deflation Problem2
- Degraded1
Patient problems
- No Clinical Signs, Symptoms or Conditions41
- Foreign Body Embolism1
- Foreign Body In Patient1
- Laceration(s)1
Recent adverse event reports
The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 22, 2026Material Fragmentation
AS REPORTED, DURING USE IN PATIENT, THE BALLOON OF THIS FOGARTY EMBOLECTOMY CATHETER BURST. THERE IS NO ALLEGATION OF MISSING PIECES. NO ADDITIONAL INTERVENTION WAS REQUIRED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS RECEIVED FOR EVALUATION AND THE BALLOON EDGES DID NOT MATCH AT THE RUPTURED REGION. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25252216 ↗MalfunctionJan 30, 2026Material Fragmentation; Inflation Problem
AS REPORTED, DURING USE IN A PATIENT WITH THIS FOGARTY EMBOLECTOMY CATHETER, THE BALLOON INFLATION FAILED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS RECEIVED FOR EVALUATION. AS PER EVALUATION RESULTS, THE BALLOON WAS FOUND TO BE RUPTURED AT CENTRAL AREA. BALLOON EDGES DID NOT APPEAR TO MATCH AT THE RUPTURED REGION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24228901 ↗MalfunctionJan 29, 2026Burst Container or Vessel; Material Fragmentation
AS REPORTED, RIGHT AFTER INSERTION, THE BALLOON OF THIS EMBOLECTOMY FOGARTY CATHETER BURST. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. AS PER EVALUATION RESULTS, THE BALLOON WAS FOUND TO BE RUPTURED AT CENTRAL AREA. BALLOON EDGES DID NOT APPEAR TO MATCH AT THE RUPTURED REGION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24208394 ↗MalfunctionJan 9, 2026Material Fragmentation
AS REPORTED, DURING SURGERY, THE BALLOON OF THIS EMBOLECTOMY FOGARTY CATHETER LOST PRESSURE SUDDENLY. UPON REMOVAL, THE BALLOON WAS FOUND TO BE RUPTURED. THE ISSUE WAS SOLVED BY REPLACING THE DEVICE WITH A NEW ONE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. AS PER EVALUATION RESULTS, THE BALLOON WAS FOUND TO BE RUPTURED AT CENTRAL AREA. LATEX EDGES DID NOT APPEAR TO MATCH AT THE RUPTURED LOCATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24018788 ↗MalfunctionJan 2, 2026Material Fragmentation
AS REPORTED, DURING USE IN PATIENT, THE BALLOON OF THIS FOGARTY EMBOLECTOMY CATHETER BURST. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. AS PER PRODUCT EVALUATION, THE BALLOON WAS FOUND TO BE RUPTURED AT CENTRAL AREA AND THE RUPTURED EDGES DID NOT APPEARED TO MATCH. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23941627 ↗MalfunctionDec 11, 2025Failure to Deflate; Failure to Deflate
IT WAS REPORTED THAT BEFORE USE OF THE FOGARTY CATHETER, WHEN BENCH TESTING, IT WOULD EITHER NOT INFLATE OR WAS DIFFICULT TO INFLATE AND THEN NOT DEFLATE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23785827 ↗MalfunctionDec 11, 2025Failure to Deflate; Failure to Deflate
IT WAS REPORTED THAT DURING USE OF THE FOGARTY CATHETER IT WOULD EITHER NOT INFLATE OR WAS DIFFICULT TO INFLATE AND THEN NOT DEFLATE. THE CATHETER WAS REMOVED FROM THE PATIENT ONCE THE ISSUE WAS NOTICED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23785791 ↗MalfunctionDec 10, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING SET-UP THIS FOGARTY EMBOLECTOMY CATHETER BODY WAS BROKEN IN HALF. NO FURTHER INFORMATION IS AVAILABLE. THERE WAS NO ALLEGATION OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23766725 ↗MalfunctionDec 8, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, BEFORE USING THIS FOGARTY EMBOLECTOMY CATHETER, THE CATHETER WAS BROKEN INTO TWO PIECES. THERE IS NO ALLEGATION OF PATIENT INJURY. PATIENT DEMOGRAPHICS WERE UNABLE TO BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23750655 ↗MalfunctionNov 7, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING USE IN PATIENT, THE BALLOON OF THIS FOGARTY EMBOLECTOMY CATHETER FAILED THE INFLATION. NO FURTHER INFORMATION WAS AVAILABLE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS RECEIVED FOR EVALUATION AND THE CATHETER BODY WAS BROKEN INTO TWO PIECES. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23494701 ↗MalfunctionOct 20, 2025Degraded; Material Fragmentation; Degraded; Material Fragmentation
AS REPORTED, DURING USE IN PATIENT OF THIS FOGARTY CATHETER, THE BALLOON DID NOT INFLATE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS RECEIVED FOR EVALUATION. AS PER PRODUCT EVALUATION RESULTS, MULTIPLE CRACKS WERE EVIDENT ON THE BALLOON LATEX AND CRACKS DID NOT APPEAR TO MATCH AT THE RUPTURED LOCATION. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23344241 ↗MalfunctionOct 17, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, BEFORE USE IN PATIENT, THIS FOGARTY EMBOLECTOMY CATHETER BROKE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. AS PER EVALUATION FINDINGS, THE CATHETER BODY WAS FOUND SEPARATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23323977 ↗MalfunctionSep 12, 2025Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, DURING USE AFTER SUCCESSFUL TESTING OF THIS FOGARTY EMBOLECTOMY CATHETER, BALLOON BURST. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS RECEIVED FOR EVALUATION. AS PER EVALUATION RESULTS, BALLOON WAS FOUND TO BE RUPTURED AT CENTRAL AREA. BALLOON EDGES DID NOT APPEAR TO MATCH AT THE RUPTURED REGION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23035189 ↗InjurySep 9, 2025Material Fragmentation; Material Fragmentation
AS REPORTED, DURING THE TEST OF THIS FOGARTY EMBOLECTOMY CATHETER, THE BALLOON BURST. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED. AS PER EVALUATION FINDINGS, THE BALLOON WAS FOUND TO BE RUPTURED AT THE CENTRAL AREA. BALLOON EDGES DID NOT APPEAR TO MATCH AT THE RUPTURED REGION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23002022 ↗MalfunctionSep 9, 2025Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, DURING USE OF THIS FOGARTY THRU-LUMEN EMBOLECTOMY CATHETER, BALLOON BURST. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS RECEIVED FOR EVALUATION. AS PER EVALUATION RESULTS, BALLOON WAS FOUND TO BE RUPTURED IN THE CENTRAL AREA. THE BALLOON EDGES DID NOT APPEAR TO MATCH AT THE RUPTURED REGION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23001850 ↗MalfunctionAug 5, 2025Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, DURING TESTING OF THIS FOGARTY EMBOLECTOMY CATHETER A SALINE LEAKAGE WAS NOTICED. NO FURTHER INFORMATION AVAILABLE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. PATIENT DEMOGRAPHICS NOT AVAILABLE. AS PER PRODUCT EVALUATION, THE BALLOON WAS FOUND TO BE RUPTURED AT CENTRAL AREA AND THE RUPTURED EDGES DID NOT APPEARED TO MATCH AT THE RUPTURED LOCATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22715130 ↗MalfunctionJul 31, 2025Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, DURING TESTING, THIS FOGARTY EMBOLECTOMY CATHETER LEAKED SALINE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. A PER PRODUCT EVALUATION FINDINGS, THE BALLOON WAS FOUND TO BE RUPTURED IN THE CENTRAL AREA. THE RUPTURED EDGES DID NOT APPEAR TO MATCH UP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22673724 ↗MalfunctionJul 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT BEFORE USE OF THIS FOGARTY CATHETER, THE BALLOON WAS BROKEN. THE BALLOON DID NOT INFLATE WHEN TESTED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22569912 ↗MalfunctionJul 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT BEFORE USE OF THIS FOGARTY CATHETER, THE BALLOON WAS BROKEN. THE BALLOON DID NOT INFLATE WHEN TESTED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22569891 ↗MalfunctionJul 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT BEFORE USE OF THIS FOGARTY CATHETER, THE BALLOON WAS BROKEN. THE BALLOON DID NOT INFLATE WHEN TESTED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22569521 ↗MalfunctionJul 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT BEFORE USE OF THIS FOGARTY CATHETER, THE BALLOON WAS BROKEN. THE BALLOON DID NOT INFLATE WHEN TESTED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22569402 ↗MalfunctionJul 22, 2025Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, WHILE TESTING THIS FOGARTY EMBOLECTOMY CATHETER, THE BALLOON BURST. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS RECEIVED FOR EVALUATION. AS PER EVALUATION RESULTS, THE BALLOON WAS FOUND TO BE RUPTURED AT CENTRAL AREA. THE RUPTURED EDGES DID NOT APPEARED TO MATCH AT THE RUPTURED LOCATION. PATIENT DEMOGRAPHIC DATA UNABLE TO BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22563494 ↗MalfunctionJul 17, 2025Failure to Deflate; Failure to Deflate
AS REPORTED, WHEN TESTING THE BALLOON OF THIS FOGARTY EMBOLECTOMY CATHETER, IT WORKED CORRECTLY, HOWEVER, DURING USE, THE BALLOON WAS DIFFICULT TO DEFLATE. AS PER PRODUCT EVALUATION FINDINGS, THE PROXIMAL BUSHING AND WINDINGS HAD SLIPPED DISTAL ON THE CATHETER TIP. PROXIMAL WINDINGS WERE LOCATED ON THE BALLOON INFLATION PORT. ADHESIVE WAS VISIBLE AT THE BOND LOCATION ON THE CATHETER TIP. BALLOON WAS UNABLE TO BE DEFLATED AS IT WAS SLIPPED DISTAL OF THE INFLATION PORTS. THERE IS NO ALLEGATION OF PATIENT INJURY. THERE IS NO MORE AVAILABLE INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22530347 ↗MalfunctionJun 30, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT BEFORE USE OF THIS FOGARTY CATHETER, THE BALLOON WAS BROKEN. THE BALLOON DID NOT INFLATE WHEN TESTED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22367124 ↗MalfunctionJun 12, 2025Burst Container or Vessel; Device Damaged Prior to Use; Burst Container or Vessel; Device Damaged Prior to Use
AS REPORTED, WHEN OPENING THE PACKAGE OF THIS FOGARTY CATHETER, THE CATHETER BODY WAS FOUND KINKED. THERE WAS NO ALLEGATION OF PATIENT INJURY. AS PER PRODUCT EVALUATION FINDINGS BALLOON EDGES DID NOT MATCH AT THE RUPTURED REGION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22191969 ↗
FDA clearance and classification
Cleared through 510(k) K240078 on Aug 28, 2024 (decision: Substantially Equivalent). Submitted by Edwards Lifesciences, LLC as “Fogarty Thru-Lumen Embolectomy Catheter”. The FDA review took 230 days. FDA record ↗
Cleared through 510(k) K892410 on Jul 20, 1989 (decision: Substantially Equivalent). Submitted by Baxter Healthcare Corp as “Fogarty(R) Thru-Lumen Embolectomy Catheter”. The FDA review took 104 days. FDA record ↗
The FDA regulates product code DXE (catheter, embolectomy) under 21 CFR 870.5150 as a Class II device. 21 CFR 870.5150 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the FOGARTY 12TLW805F35
Has the FOGARTY 12TLW805F35 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0238-2024 (Class II), began on Sep 7, 2023: Inadvertent release of embolectomy catheters that may experience slower than normal balloon deflation (greater-than 15 seconds). If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the FOGARTY 12TLW805F35?
The FDA's MAUDE database lists 43 adverse event reports for this device, received between Oct 5, 2021 and May 22, 2026. In those reports, the most common specific device problems are “Burst Container or Vessel” (11), “Material Fragmentation” (9) and “Detachment of Device or Device Component” (6); the most common patient problems are “Foreign Body Embolism” (1), “Foreign Body In Patient” (1) and “Laceration(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the FOGARTY 12TLW805F35?
None of the reports are classified as deaths, 1 as an injury and 42 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the FOGARTY 12TLW805F35 FDA approved?
It was cleared through 510(k) K892410 on Jul 20, 1989; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the FOGARTY 12TLW805F35, and is it still sold?
The FDA's device database lists Edwards Lifesciences LLC as the company responsible for it (brand FOGARTY, model 12TLW805F35). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, EMBOLECTOMY devices
- Indigo System CAT6Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths408 reports
- Indigo System CATRXKITPenumbra, Inc. · Catheter, Embolectomy1 death339 reports
- Fogarty 120804FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall1 death200 reports
- LeMaitre Embolectomy Catheter 1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy1 recall153 reports
- FOGARTY 120404FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall61 reports
- Indigo System CAT6KIT-APenumbra, Inc. · Catheter, Embolectomy109 reports
- LeMaitre Embolectomy Catheter e1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy2 recalls4 reports
- Indigo System CAT8XTORQ115Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths98 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page