FOGARTY 12TLW806FRecalls & reported problems

Edwards Lifesciences LLC · FOGARTY THRU-LUMEN EMBOLECTOMY CATHETER 6F 80CM · Class 2

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00690103043549), received between Dec 21, 2023 and Sep 8, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K240078, K892410); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel3
  • Break1
  • Failure to Deflate1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 8, 2025Burst Container or Vessel; Material Fragmentation; Burst Container or Vessel; Material Fragmentation

    AS REPORTED, DURING USE IN PATIENT, THE BALLOON OF THIS EMBOLECTOMY FOGARTY CATHETER BURST. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. AS PER EVALUATION RESULTS, THE BALLOON WAS FOUND TO BE RUPTURED IN THE MIDDLE AREA. THE EDGES OF LATEX DID NOT APPEARED TO MATCH AT THE RUPTURED REGION. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22992994 ↗
  • MalfunctionJul 17, 2025Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, DURING TEST PRIOR TO USE IN PATIENT, THE BALLOON OF THIS EMBOLECTOMY FOGARTY CATHETER BURST. THE DEVICE WAS RECEIVED FOR EVALUATION, THE BALLOON WAS RUPTURED NEAR THE PROXIMAL WINDING AND THE EDGES OF LATEX DID NOT MATCH AT THE RUPTURED REGION. THERE WAS NO ALLEGATION OF PATIENT INJURY. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22528713 ↗
  • MalfunctionApr 25, 2025Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, DURING TESTING OF THIS FOGARTY CATHETER, THE BALLOON BURST. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS RECEIVED FOR EVALUATION AND THE EDGES OF LATEX DID NOT MATCH AT THE RUPTURED LOCATION. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21907150 ↗
  • MalfunctionJan 24, 2025Failure to Deflate; Failure to Deflate

    AS REPORTED, DURING USE IN PATIENT OF THIS FOGARTY EMBOLECTOMY CATHETER, THE BALLOON DID NOT DEFLATE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21227870 ↗
  • MalfunctionDec 21, 2023Break; Break

    IT WAS REPORTED THAT A MODEL 12TLW806F FOGARTY CATHETER BROKE IN HALF DURING A PROCEDURE. CUSTOMER WAS ABLE TO SUCCESSFULLY REMOVE CATHETER WITHOUT AN ADDITIONAL PROCEDURE. NO FURTHER TROUBLESHOOTING WAS ATTEMPTED, A NEW FOGARTY WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18378874 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K240078, K892410. One submission often covers many models, so check the product and lot details in each record.

  • Z-0238-2024 · Sep 7, 2023 · Open, Classified
    Fogarty Thru-Lumen Embolectomy Catheter, REF: 12TLW805F35

Other CATHETER, EMBOLECTOMY devices

All CATHETER, EMBOLECTOMY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.