Skip to content

HemoSphere Tissue Oximetry Module HEMTOM10Recalls & reported problems

Edwards Lifesciences LLC · HemoSphere Tissue Oximetry Module · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00690103200355), received between Jun 30, 2022 and Feb 12, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 12, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT A HEMTOM10 HAD INTERMITTENT CONNECTION WHICH CAUSED INCONSISTENT VALUES. DURING A CASE, STAFF NOTICED THE VALUE DECREASE. IT IS UNKNOWN WHICH HEMODYNAMIC VALUE DECREASED. THE ISSUE WAS RESOLVED BY EXCHANGING THE MODULE. THERE WERE NO PATIENT INJURIES. THE DEVICE IS AVAILABLE FOR RETURN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21370434 ↗
  • MalfunctionAug 3, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT THE HEMTOM10 HAS BEEN GENERATING INTERMITTENT AND FLUCTUATING STO2 VALUES. THE ISSUE HAS BEEN WITH THE A1 L VALUES. THE A2 R VALUE STAYED STEADY FROM 88 89 DURING A CASE. WHEN THERE IS A VALUE THE A1 L VALUE FLUCTUATED BETWEEN 65 TO 88 CHANGING EVERY FEW SECONDS. AFTER A FEW MINUTES OF DISPLAYED VALUES THE A1 L SUDDENLY STOPS GENERATING VALUES. AFTER SEVERAL MINUTES THE CYCLE STARTS AGAIN. THIS WAS DURING USE. THERE IS NO PATIENT INJURY. PATIENT DEMOGRAPHICS WERE REQUESTED BUT UNAVAILABLE. A HEM1 HEMFSM10 AND FORESIGHT SENSOR WILL ALSO BE SENT IN SEPARATE MDRS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15161315 ↗
  • MalfunctionJun 30, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT THE HEMFSM10 CHANNEL 2 STO2 VALUE DROPPED 15 TO 20 POINTS AFTER STARTING A NEW PATIENT IN HEMOSPHERE. THEY STARTED THE CASE WITH FORESIGHT ONLY AND THE INITIAL STO2 VALUES FOR BOTH LEFT AND RIGHT BRAIN WERE IN THE MID 70S. THE ATTENDING ANESTHESIOLOGIST DECIDED TO CONNECT THE SWAN GANZ. THEY TURNED THE HEMOSPHERE OFF AND STARTED A NEW PATIENT. ONCE THE HEMOSPHERE CAME UP AGAIN THE LEFT STO2 CHANNEL 1 VALUES WERE THE SAME 72 TO 74 BUT THE RIGHT STO2 CHANNEL 2 DROPPED 15 TO 20 POINTS. THE DIFFERENCE REMAINED THE ENTIRE CASE. THIS WAS DURING USE. THERE IS NO PATIENT INJURY. THERE WILL BE NO PRODUCT RETURN PER THE REP. PATIENT DEMOGRAPHICS WERE REQUESTED BUT UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14879625 ↗

Other Oximeter, Tissue Saturation devices

All Oximeter, Tissue Saturation devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.