Skip to content

Fogarty 120803FPRecalls & reported problems

Edwards Lifesciences LLC · FOGARTY ARTERIAL EMBOLECTOMY CATHETER 3F 80CM · Class 2

All models of FOGARTY (32)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00690103205091), received between Aug 30, 2022 and Jan 31, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 6, 2022Class IIOpen, Classified

    Due to an increase of confirmed latex deterioration complaints for arterial embolectomy catheters packaged in pouches.

    Z-0146-2023 · Root cause: Packaging · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Fragmentation2
  • Burst Container or Vessel1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 31, 2025Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, DURING PROCEDURE WITH THIS FOGARTY EMBOLECTOMY CATHETER, RIGHT AFTER INFLATING THE BALLOON, IT BURST. AS PER PRODUCT EVALUATION FINDINGS, BALLOON WAS FOUND TO BE RUPTURED AT CENTRAL AREA AND THE SPRING TIP WIRE WAS EXPOSED. THERE IS NO ALLEGATION OF PATIENT INJURY. PATIENT DEMOGRAPHICS WERE NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21283017 ↗
  • MalfunctionOct 27, 2022Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT A TEAM OF EW REPRESENTATIVES PERFORMED AN ON-SITE VISIT TO INVESTIGATE COMPLAINTS RELATED TO FOGARTY CATHETERS IN THE VASCUPOUCH. THE INITIAL INTERVIEW WITH HOSPITAL STAFF STATED THAT CATHETERS WERE STORED EITHER IN CENTRAL SUPPLY OR THE OR CORE. THE EDWARDS TEAM VISITED THE LOCATIONS WITHIN THE HOSPITAL WHERE CATHETERS WERE RECEIVED, STORED, AND ULTIMATELY USED. DURING THE FACILITY WALK THROUGH, IT WAS FOUND THAT THE CATHETERS WERE ALSO STORED IN TRAUMA OR 1. CATHETERS WERE TESTED IN LOCATIONS WHERE BOTH THE OUTER CORRUGATED PACKAGING HAD BEEN REMOVED AND PRIOR TO THE LOCATION OF THE REPORTED FAILURE, SPECIFICALLY, OR CORE AND TRAUMA OR 1. THE TEAM FOUND EVIDENCE OF XRAY IMAGING AND UV SANITIZATION BEING PRESENT IN TRAUMA OR 1, WHICH CORRESPONDS WITH OTHER FACILITY VISIT FINDINGS OF HIGH ENERGY SOURCES WHERE THE CATHETERS ARE STORED. THIS REPORT IS FOR ONE OF THE MODEL 120803FP, LOT 63883786 CATHETERS FROM TRAUMA OR 1 WHICH WAS NOTED TO HAVE A BALLOON THAT WOULD NOT INFLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15680157 ↗
  • MalfunctionAug 30, 2022Material Fragmentation; Material Rupture; Material Fragmentation; Material Rupture

    IT WAS REPORTED THAT PRIOR TO USE, THE BALLOONS ON SEVEN ARTERIAL EMBOLECTOMY CATHETERS WERE NOTED TO HAVE HOLES IN THEM WHEN INFLATED. IT WAS STATED THAT THE SURGEON WAS CONCERNED AND HAS A BUNCH OF EXPERIENCE WITH THESE CATHETERS, SO THE CUSTOMER FELT THAT IT WAS NOT USER ERROR. THE ONLY TROUBLESHOOTING WAS TO GET ANOTHER CATHETER HOWEVER, THE TECH STATED THAT AS SHE INFLATED THE BALLOON WITH FLUID THERE WERE MULTIPLE PIN HOLES IN THE BALLOON. THERE WAS NO PATIENT INJURY. THE EXACT DATE OF THE EVENT IS UNKNOWN. THIS REPORT IS FOR ONE MODEL 120803FP CATHETER FROM LOT 63954397.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15321025 ↗

FDA clearance and classification

The FDA regulates product code DXE (catheter, embolectomy) under 21 CFR 870.5150 as a Class II device. 21 CFR 870.5150 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Fogarty 120803FP

Has the Fogarty 120803FP been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0146-2023 (Class II), began on Oct 6, 2022: Due to an increase of confirmed latex deterioration complaints for arterial embolectomy catheters packaged in pouches. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Fogarty 120803FP?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Aug 30, 2022 and Jan 31, 2025. In those reports, the most common specific device problems are “Material Fragmentation” (2), “Burst Container or Vessel” (1) and “Material Rupture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Fogarty 120803FP?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Fogarty 120803FP, and is it still sold?

The FDA's device database lists Edwards Lifesciences LLC as the company responsible for it (brand Fogarty, model 120803FP). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, Embolectomy devices

All Catheter, Embolectomy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page