Skip to content

Fogarty 120806FPRecalls & reported problems

Edwards Lifesciences LLC · FOGARTY ARTERIAL EMBOLECTOMY CATHETER 6F 80CM · Class 2

All models of FOGARTY (32)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00690103205237), received between Oct 27, 2022 and Feb 7, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K193379); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Oct 6, 2022Class IIOpen, Classified

    Due to an increase of confirmed latex deterioration complaints for arterial embolectomy catheters packaged in pouches.

    Z-0146-2023 · Root cause: Packaging · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Fragmentation1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 7, 2023Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED WHILE PREPPING A FOGARTY CATHETER, MODEL 120806FP, LOT 63612683, THE BALLOON POPPED, OR HAD HOLES IN THE BALLOON. THERE WAS NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16315118 ↗
  • MalfunctionOct 27, 2022Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT A TEAM OF EW REPRESENTATIVES PERFORMED AN ON-SITE VISIT TO INVESTIGATE COMPLAINTS RELATED TO FOGARTY CATHETERS IN THE VASCUPOUCH. THE INITIAL INTERVIEW WITH HOSPITAL STAFF STATED THAT CATHETERS WERE STORED EITHER IN CENTRAL SUPPLY OR THE OR CORE. THE EDWARDS TEAM VISITED THE LOCATIONS WITHIN THE HOSPITAL WHERE CATHETERS WERE RECEIVED, STORED, AND ULTIMATELY USED. DURING THE FACILITY WALK THROUGH, IT WAS FOUND THAT THE CATHETERS WERE ALSO STORED IN TRAUMA OR 1. CATHETERS WERE TESTED IN LOCATIONS WHERE BOTH THE OUTER CORRUGATED PACKAGING HAD BEEN REMOVED AND PRIOR TO THE LOCATION OF THE REPORTED FAILURE, SPECIFICALLY, OR CORE AND TRAUMA OR 1. THE TEAM FOUND EVIDENCE OF XRAY IMAGING AND UV SANITIZATION BEING PRESENT IN TRAUMA OR 1, WHICH CORRESPONDS WITH OTHER FACILITY VISIT FINDINGS OF HIGH ENERGY SOURCES WHERE THE CATHETERS ARE STORED. THIS REPORT IS FOR ONE OF THE MODEL 120806FP, LOT 63852678 CATHETERS FROM TRAUMA OR 1 WHICH WAS NOTED TO HAVE A BALLOON THAT WOULD NOT INFLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15680194 ↗

FDA clearance and classification

Cleared through 510(k) K193379 on Jun 12, 2020 (decision: Substantially Equivalent). Submitted by Edwards Lifesciences, LLC as “Fogarty Arterial Embolectomy Catheter, Fogarty Biliary Balloon Probe”. The FDA review took 190 days. FDA record ↗

The FDA regulates product code DXE (catheter, embolectomy) under 21 CFR 870.5150 as a Class II device. 21 CFR 870.5150 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K193379. One submission often covers many models, so check the product and lot details in each record.

  • Z-0147-2023 · Oct 6, 2022 · Open, Classified
    Fogarty Biliary Balloon Probes Model Number: 410235FP 410405FP 410236FP

Common questions about the Fogarty 120806FP

Has the Fogarty 120806FP been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0146-2023 (Class II), began on Oct 6, 2022: Due to an increase of confirmed latex deterioration complaints for arterial embolectomy catheters packaged in pouches. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Fogarty 120806FP?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Oct 27, 2022 and Feb 7, 2023. In those reports, the most common specific device problems are “Material Fragmentation” (1) and “Material Puncture/Hole” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Fogarty 120806FP?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Fogarty 120806FP FDA approved?

It was cleared through 510(k) K193379 on Jun 12, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Fogarty 120806FP, and is it still sold?

The FDA's device database lists Edwards Lifesciences LLC as the company responsible for it (brand Fogarty, model 120806FP). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, Embolectomy devices

All Catheter, Embolectomy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page