Skip to content

ESHEATH OPTIMA INTRODUCER SET (14F) 14000ES14Recalls & reported problems

Edwards Lifesciences LLC · eSheath Optima Introducer Set (14F) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00690103220629), received between Jun 12, 2025 and Jul 14, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert1
  • Material Integrity Problem1
  • Material Puncture/Hole1

Patient problems

  • Insufficient Information1
  • Perforation of Vessels1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 14, 2025Difficult to Insert; Difficult to Insert

    AS REPORTED BY THE FIELD CLINICAL SPECIALIST (FCS), DURING IMPLANT OF A 23 MM SAPIEN 3 ULTRA RESILIA VALVE IN THE AORTIC POSITION, THE SURGEON USED THE 21F IN-SHEATH DILATOR (ISD) TO PREDILATE THE SHEATH BEFORE PUTTING IN THE COMMANDER DELIVERY SYSTEM. THERE WAS QUITE A BIT OF FORCE WHILE PUTTING IN THE 21F ISD. DURING THE PRE-DILATION USING THE 21F THE TIP OF THE ISD WAS ABLE TO PROGRESS PAST THE ILIAC BIFURCATION IN THE ABDOMINAL AORTA TO THE END OF THE SHEATH. WHEN THE PHYSICIAN REMOVED THE SHEATH AT THE END OF THE CASE, THERE WAS AN ILIAC PERFORATION. THE SURGEON THOUGHT PERHAPS PREDILATING THE SHEATH WITH THE OPTIMA ISD CAUSED THE PERFORATION. THE SHEATH DID NOT HAVE ANY VISIBLE DAMAGE. THE SURGEON PUT IN A COVERED STENT AND CONTROLLED THE PERFORATION. THE PATIENT IS STABLE AND DOING OK.

    Patient problems: Perforation of Vessels

    MAUDE report 22501107 ↗
  • MalfunctionJun 12, 2025Material Puncture/Hole; Material Integrity Problem; Material Puncture/Hole; Material Integrity Problem

    AS REPORTED BY THE FIELD CLINICAL SPECIALIST (FCS), DURING THE TRANSFEMORAL TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVR) PROCEDURE WITH A 26 MM SAPIEN 3 ULTRA VALVE, POST SUCCESSFUL VALVE DEPLOYMENT, A PINHOLE WAS NOTED IN THE 14FR ESHEATH OPTIMA WHEN IT WAS WITHDRAWN FROM THE PATIENT. IT WAS BELIEVED TO HAVE BEEN CAUSED BY THE CALCIUM NODULE AFTER THE IN-SHEATH DILATOR (ISD) WAS USED TO HELP WITH DIFFUSING THE DISEASE IN THE ANATOMY. THERE WAS NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 22195868 ↗

Other Introducer, Catheter devices

All Introducer, Catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page