Skip to content

ESHEATH OPTIMA 14000ES16ARecalls & reported problems

Edwards Lifesciences LLC · eSheath Optima Introducer Set (16F) · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00690103220636), received between Dec 8, 2025 and Dec 8, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Puncture/Hole” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 8, 2025Material Puncture/Hole; Material Puncture/Hole

    AS REPORTED BY THE FIELD CLINICAL SPECIALIST (FCS), DURING IMPLANT OF A 29MM SAPIEN 3 ULTRA RESILIA VALVE IN THE AORTIC POSITION VIA RIGHT TRANSFEMORAL APPROACH, IT'S NOTED THERE WAS SOME RESISTANCE WHILE ADVANCING THE DELIVERY SYSTEM WITH CRIMPED VALVE THROUGH THE 16F OPTIMA SHEATH DUE TO A KINK IN THE OPTIMA AS WELL AS A SMALL PUNCTURE FROM THE PATIENT'S TORTUOSITY OF THE RIGHT ILIAC. THE DELIVERY SYSTEM WAS ABLE TO BE PULLED OUT OF THE SHEATH. THERE WERE NO PATIENT INJURIES. THE VALVE WAS IMPLANTED SUCCESSFULLY, AND THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23742505 ↗

Other Introducer, Catheter devices

All Introducer, Catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page