Capiox® RX Hollow Fiber Oxygenator with Hardshell Reservoir 3CX*RX15RW30Recalls & reported problems
Terumo Cardiovascular Systems Corporation · Class 2
All models of Capiox RX Hollow Fiber Oxygenator with Hardshell Reservoir (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00699753450127), received between Aug 17, 2018 and Sep 10, 2021 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Coagulation in Device or Device Ingredient” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Coagulation in Device or Device Ingredient2
- Use of Device Problem2
- Infusion or Flow Problem1
Patient problems
- Blood Loss1
- Injury1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- No Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 10, 2021Coagulation in Device or Device Ingredient; Coagulation in Device or Device Ingredient
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING PRIME, THE OXYGENATOR RESERVOIR GOT CLOT WHILE PRIMING WITH BLOOD. DE-FOAMING CHAMBER AND SCREEN FILTER OF RESERVOIR GOT CLOT RESULTING IN DECREASE OF DYNAMIC VOLUME ( LEVEL) WHILE CIRCULATING AND AIR CAME IN THE CIRCUIT. AS PER THE CLINICAL SPECIALIST, ADDITIONAL INFORMATION FORM THE FACILITY STATES THAT, 1 UNIT OF BLOOD WERE PUT INTO THE RESERVOIR AND A BLOOD FILTER WAS USED IN THE UNITS OF BLOOD WHEN IT WAS PUT INTO THE RESERVOIR. IT IS UNKNOWN AS TO WHAT THE LEVEL OF ANTICOAGULATION WAS OF THE BLOOD INTRODUCED TO THE RESERVOIR AND HOW MUCH HEPARIN WAS PUT IN THE PUMP PRIME. BLOOD IS GOING TO CLOT WHENEVER IS CONTACTS ANY FOREIGN SURFACE IF IT IS NOT ADEQUATELY ANTICOAGULATED. *NO PATIENT INVOLVEMENT *PRODUCT WAS CHANGED OUT PRIOR TO CPB DURING BLOOD PRIMING *PROCEDURE WAS COMPLETED SUCCESSFULLY
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12452724 ↗MalfunctionSep 25, 2020Infusion or Flow Problem; Infusion or Flow Problem
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT DURING CARDIOPULMONARY, AT THE TIME OF ENTRY IN PARTIAL EXTRA-CORPORARY CIRCULATION, THE PATIENT SATURATION BEGAN TO GO DOWN, REACHING 70%. IT IS UNKNOWN IF THERE IS BLOOD LOSS. AS PER SUBSIDIARY, IT WAS NOTICED THAT THE POST OXYGENER BLOOD SIDE¿S COLORING WAS EQUAL TO THE INSATURED VENOUS RETURN. IMMEDIATELY, AN ARTERIAL GASOMETRY WAS MADE WHERE THE OXYGEN PARTIAL PRESSURE (PO2) WAS 35.2 MMHG, AND THE SATURATION OF OXYGEN WAS 62.1%. IT WAS PROCESSED TO CONNECT AN OXYGEN TANK TO THE OXYGENATOR, THINKING THAT IT COULD BE THE BLENDER OF THE HEART-LUNG MACHINE, AND THE RESULT WAS THE SAME SO THE ANESTHESIOLOGIST WAS REQUESTED TO INCREASE THE FRACTION OF INSPIRED OXYGEN (FIO2) OF THE VENTILATOR TO REMOVE THE PUMP PATIENT, AND CAN REPLACE THE OXYGENATOR TO AN RX25, AND THIS WORKS. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT. PRODUCT WAS CHANGED OUT. PROCEDURE COMPLETED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10588401 ↗MalfunctionNov 11, 2019Coagulation in Device or Device Ingredient; Coagulation in Device or Device Ingredient
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT WHITE BITS ON SURFACE OF THE BLOOD WERE OBSERVED. IT IS UNKNOWN WHEN THE EVENT OCCURRED, IF THERE WAS A DELAY IN THE PROCEDURE, IF THE PRODUCT WAS CHANGED OUT, THE RESULTS OF THE PROCEDURE OR IF THERE WAS ANY BLOOD LOSS.
Patient problems: No Information
MAUDE report 9306911 ↗InjuryNov 12, 2018Use of Device Problem; Use of Device Problem
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT DURING CARDIOPULMONARY BYPASS, THERE WAS NO COLOR CHANGE WITH FIO2 AT 65% AND SWEEP AT 2.5 ON A 50KG PT. THE FIO2 WAS INCREASED TO 100% AND SWEEP INCREASED TO 4 TO SEE IF IT WOULD HELP AND OPENED THE PURGE ON TOP OF THE OXYGENATOR TO HELP RECIRCULATE THE BLOOD. THE PO2 AT 100% WAS ONLY 41, ANESTHESIA TURNED ON THE VENT FOR THE O2 TO RISE. THE PO2 ONLY GOT TO 45. AT THAT POINT, IT WAS DECIDED TO CHANGE OUT THE OXYGENATOR. ONCE CHANGED OUT AND DEAIRED, THE PO2 WAS AT 188 WITH JUST 50%. PER USER FACILITY THE PATIENT WAS BEING VENTILATED AS THEY WENT ON BYPASS. THERE WAS A DELAY FOR 10-15 MINUTES. PRODUCT WAS CHANGED OUT. THERE WAS A BLOOD LOSS OF 150 ML. AFTER CHANGE OUT, SURGERY WAS COMPLETED SUCCESSFULLY. POST OP, PATIENT WAS SITTING UP IN BED AND DOING WELL, FOLLOWING COMMENDS,
Patient problems: Blood Loss; Injury
MAUDE report 8062603 ↗MalfunctionAug 17, 2018Use of Device Problem; Use of Device Problem
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT DURING PRIME, A PO2 OF 330 MMHG WAS NOTED DURING THE PRIMING PROCESS WHEN THE CDI 500 CELL WAS PUT IN-LINE WHICH IS NORMALLY ABOVE 600MMHG. A SAMPLE WAS DRAWN OFF THE PUMP AND IT WAS RUN IT ON THE I-STAT BLOOD GAS ANALYZER. THE 1ST PO2 CAME BACK AT 360MMHG, THE SWEEP GAS WAS TURNED UP TO 3 LITERS PER MIN AND CHECKED FOR CRACKS/LEAKS/LOOSE CONNECTIONS IN THE GAS DELIVERY SYSTEM. THE CDI PO2 VARIED BETWEEN 330-360MMHG. AN OXYGEN TANK WAS THEN BROUGHT TO THE ROOM AND HAD ANESTHESIA RUN A GAS LINE FROM THEIR VENTILATOR THROUGH OUR GAS FILTER AND INTO THE OXYGENATOR. THE VENTILATOR WAS SET AT 100% FIO2 WITH 2 LITERS PER MINUTE OF SWEEP GAS. THE HIGHEST THE CDI READ WAS 365MMHG. ANOTHER I-STAT WAS RAN OFF THE PUMP AND THE RESULTS CAME BACK AT 394 MMHG. THE OXYGENATOR WAS CHANGED OUT. THE PUMP¿S GAS DELIVERY SYSTEM WAS HOOKED TO THE NEW OXYGENATOR AND THE CDI 500 STILL HAD A LOW READING OF 380MMHG. ANOTHER PRIME SAMPLE WAS DRAWN AND CHANGED THE CDI CELL. THE NEW CELL GAVE A READING OF APPROXIMATELY 400MMHG AND THE PRIME GAS FROM THE I-STAT CAME BACK AT 535 MMHG. THE PROCEDURE WAS COMPLETED WITH THE NEW OXYGENATOR WITH NO ISSUES. NO
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7794633 ↗
Other Oxygenator, Cardiopulmonary Bypass devices
- HLS Set BEQ-HLS 7050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls13 deaths90 reports
- Affinity Fusion® BB811Medtronic, Inc. · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls90 reports
- HLS Set BEQ-HLS 5050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls4 reports
- NA HLS Set Advanced 5.0, Small Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS3 recalls0 reports
- A.M.G. PMP Sterile A.M.G. Module PMP NO T.P. SterileEurosets Srl · Oxygenator, Cardiopulmonary Bypass2 deaths106 reports
- CAPIOX®FX CX*FX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass2 deaths102 reports
- CAPIOX®RX CX*RX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass1 death102 reports
- QUADROX-i HMO 50000-USA #SQUADROX-i Small Adult, no FilterMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page