Capiox® RX Hollow Fiber Oxygenator with Hardshell Reservoir 3CX*RX15RE30Recalls & reported problems
Terumo Cardiovascular Systems Corporation · Class 2
All models of Capiox RX Hollow Fiber Oxygenator with Hardshell Reservoir (6)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00699753450134), received between Mar 11, 2019 and Oct 18, 2019 — classified as 16 malfunctions.
The most frequently reported specific device problem is “Unable to Obtain Readings” (15 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Mar 1, 2023Class IIOpen, Classified
Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the oxygenator port.
Z-1345-2023 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Unable to Obtain Readings15
- Device Damaged Prior to Receipt by User1
Patient problems
- No Known Impact Or Consequence To Patient7
- No Patient Involvement7
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 18, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING PRIME, THE DEVICE DOES NOT REGISTER AN ARTERIAL TEMPERATURE. PER USER FACILITY THE TEMPERATURE CABLE REGISTERS A TEMPERATURE WHEN USED ON THE VENOUS TEMP PROBE BUT NOT ON THE ARTERIAL. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT; THE PRODUCT WAS NOT CHANGED OUT; THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9210852 ↗MalfunctionAug 23, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT, DURING OTHER-NON CLINICAL ACTIVITY, THE ARTERIAL PROBE WAS FAULTY. NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8926943 ↗MalfunctionAug 23, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING SETUP, THE ARTERIAL TEMPERATURE PROBE WAS FAULTY. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT. PRODUCT WAS NOT CHANGED OUT. PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8926707 ↗MalfunctionAug 20, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING SETUP, THE ARTERIAL TEMPERATURE PROBE WAS FAULTY. NO PATIENT INVOLVEMENT. PRODUCT WAS CHANGED OUT. PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Patient Involvement
MAUDE report 8910501 ↗MalfunctionAug 20, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING SETUP, THE ARTERIAL PROBE WAS FAULTY. NO PATIENT INVOLVEMENT. PRODUCT WAS CHANGED OUT. PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Patient Involvement
MAUDE report 8910316 ↗MalfunctionAug 20, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING SETUP, THE ARTERIAL TEMPERATURE PROBE WAS FAULTY. NO PATIENT INVOLVEMENT. PRODUCT WAS CHANGED OUT. PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Patient Involvement
MAUDE report 8910293 ↗MalfunctionAug 15, 2019Device Damaged Prior to Receipt by User; Device Damaged Prior to Receipt by User
THE DISTRIBUTOR REPORTED TO TERUMO CARDIOVASCULAR THAT DURING OUT OF BOX, THE RECEIVED CODE OF THE DEVICE WAS DIFFERENT FROM THE ONE REQUESTED. NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8897815 ↗MalfunctionAug 9, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING SETUP, THE ARTERIAL PROBE WAS FAULTY. NO PATIENT INVOLVEMENT. PRODUCT WAS CHANGED OUT. PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Patient Involvement
MAUDE report 8880790 ↗MalfunctionJul 31, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING SETUP, THE FAULTY TEMPERATURE PROBE WAS UNABLE TO REGISTER A TEMPERATURE READING. AS PER THE USER FACILITY, THEY DID NOT CHANGE OUT THE PRODUCT; BYPASSED THE TEMPERATURE READING AND CONTINUED WITH THE SURGERY. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT. PRODUCT WAS NOT CHANGED OUT. PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8848185 ↗MalfunctionMay 20, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING SETUP, THE ARTERIAL PROBE WAS UNABLE TO REGISTER A TEMPERATURE READING. NO CONSEQUENCES OR IMPACT TO PATIENT. PRODUCT WAS NOT CHANGED OUT. PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8625490 ↗MalfunctionMay 20, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING SETUP, THE DEVICE WAS UNABLE TO REGISTER A TEMPERATURE READING. NO CONSEQUENCES OR IMPACT TO THE PATIENT. PRODUCT WAS NOT CHANGED OUT. PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8625112 ↗MalfunctionMar 19, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING SETUP, THE ARTERIAL TEMPERATURE PROBE PORTS WERE UNABLE TO RECORD TEMPERATURE. NO PATIENT INVOLVEMENT AS THIS OCCURRED DURING SETUP. PRODUCT WAS CHANGED OUT. PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Patient Involvement
MAUDE report 8433617 ↗MalfunctionMar 11, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING SETUP, THE ARTERIAL TEMPERATURE PROBE PORTS WERE UNABLE TO RECORD TEMPERATURE. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT; THE PRODUCT WAS NOT CHANGED OUT; THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8409696 ↗MalfunctionMar 11, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING SETUP, THE ARTERIAL TEMPERATURE PROBE PORTS WERE UNABLE TO RECORD TEMPERATURE. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT. THE PRODUCT WAS NOT CHANGED OUT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8409603 ↗MalfunctionMar 11, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING SETUP, THE ARTERIAL TEMPERATURE PROBE PORTS WERE UNABLE TO RECORD TEMPERATURE. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT. THE PRODUCT NOT CHANGED OUT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8409557 ↗MalfunctionMar 11, 2019Unable to Obtain Readings; Unable to Obtain Readings
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT PRIOR TO CARDIOPULMONARY BYPASS, DURING SETUP, THE ARTERIAL TEMPERATURE PROBE PORTS WERE UNABLE TO RECORD TEMPERATURE. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT. THE PRODUCT WAS NOT CHANGED OUT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8409522 ↗
FDA clearance and classification
The FDA regulates product code DTZ (oxygenator, cardiopulmonary bypass) under 21 CFR 870.4350 as a Class II device. 21 CFR 870.4350 ↗
Devices of this type normally need 510(k) clearance before sale.
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Capiox® RX Hollow Fiber Oxygenator with Hardshell Reservoir 3CX*RX15RE30
Has the Capiox® RX Hollow Fiber Oxygenator with Hardshell Reservoir 3CX*RX15RE30 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1345-2023 (Class II), began on Mar 1, 2023: Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Capiox® RX Hollow Fiber Oxygenator with Hardshell Reservoir 3CX*RX15RE30?
The FDA's MAUDE database lists 16 adverse event reports for this device, received between Mar 11, 2019 and Oct 18, 2019. In those reports, the most common specific device problems are “Unable to Obtain Readings” (15) and “Device Damaged Prior to Receipt by User” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Capiox® RX Hollow Fiber Oxygenator with Hardshell Reservoir 3CX*RX15RE30?
None of the reports are classified as deaths, 0 as injuries and 16 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Capiox® RX Hollow Fiber Oxygenator with Hardshell Reservoir 3CX*RX15RE30, and is it still sold?
The FDA's device database lists Terumo Cardiovascular Systems Corporation as the company responsible for it (brand Capiox® RX Hollow Fiber Oxygenator with Hardshell Reservoir, model 3CX*RX15RE30). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Oxygenator, Cardiopulmonary Bypass devices
- HLS Set BEQ-HLS 7050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls13 deaths87 reports
- Affinity Fusion® BB811Medtronic, Inc. · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls90 reports
- HLS Set BEQ-HLS 5050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls4 reports
- NA HLS Set Advanced 5.0, Small Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS3 recalls0 reports
- A.M.G. PMP Sterile A.M.G. Module PMP NO T.P. SterileEurosets Srl · Oxygenator, Cardiopulmonary Bypass2 deaths106 reports
- CAPIOX®FX CX*FX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass2 deaths102 reports
- CAPIOX®RX CX*RX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass1 death102 reports
- QUADROX-i HMO 50000-USA #SQUADROX-i Small Adult, no FilterMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.