Herc Flex 360 Universal Stabilizer Arm T401162Recalls & reported problems
Terumo Cardiovascular Systems Corporation · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00699753450943), received between Jun 23, 2021 and Feb 13, 2024 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Physical Resistance/Sticking4
- Break2
- Defective Component1
- Material Too Rigid or Stiff1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Insufficient Information1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 13, 2024Break; Break
NEW INFORMATION RECEIVED THAT THE EVENT OCCURRED DURING OTHER - NON CLINICAL ACTIVITY; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18694774 ↗MalfunctionNov 6, 2023Break; Break
NEW INFORMATION RECEIVED THAT INDICATES, THE EVENT OCCURRED DURING SETUP, THE PRODUCT WAS CHANGED OUT, AND THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO PATIENT EFFECT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18073553 ↗MalfunctionApr 24, 2023Defective Component; Defective Component
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR, THAT DURING CARDIOPULMONARY BYPASS, THE SLEEVE WAS DEFECTIVE. IT IS UNKNOWN WHETHER THERE WAS A DELAY IN THE PROCEDURE. DUE TO THE UNKNOWN INFORMATION FOR THIS EVENT, IT IS BEING REPORTED. TERUMO CONTINUES TO ATTEMPT TO GAIN MORE INFORMATION REGARDING THIS EVENT FROM THE USER FACILITY. NO KNOWN HEALTH CONSEQUENCES OR IMPACT TO PATIENT. PRODUCT WAS NOT CHANGED OUT. PROCEDURE COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16799085 ↗MalfunctionJul 21, 2021Physical Resistance/Sticking; Physical Resistance/Sticking
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT THE ARM LINKS STUCK. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. DUE TO THE UNKNOWN INFORMATION FOR THIS EVENT, IT IS BEING REPORTED. TERUMO CONTINUES TO ATTEMPT TO GAIN MORE INFORMATION REGARDING THIS EVENT FROM THE USER FACILITY. THE PRODUCT WAS CHANGED OUT. PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12207757 ↗MalfunctionJun 29, 2021Physical Resistance/Sticking; Physical Resistance/Sticking
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT, THE LINKS ON ARMS WERE STICKING AND BINDING TOGETHER, AND IT WAS DIFFICULT TO USE AND POSITION. IT IS UNKNOWN WHETHER THERE WAS A DELAY IN THE PROCEDURE. DUE TO THE UNKNOWN INFORMATION FOR THIS EVENT, IT IS BEING REPORTED. TERUMO CONTINUES TO ATTEMPT TO GAIN MORE INFORMATION REGARDING THIS EVENT FROM THE USER FACILITY. THE PRODUCT WAS NOT CHANGED OUT. PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12087850 ↗MalfunctionJun 28, 2021Physical Resistance/Sticking; Physical Resistance/Sticking
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT DURING CARDIOPULMONARY BYPASS, THEY NOTICED THE LINKS ON ARMS STICKING AND BINDING TOGETHER AND DIFFICULT TO USE AND POSITION. IT IS UNKNOWN IF THE ISSUE CAUSED ANY DELAY TO THE PROCEDURE. DUE TO THE UNKNOWN INFORMATION FOR THIS EVENT, IT IS BEING REPORTED. TERUMO CONTINUES TO ATTEMPT TO GAIN MORE INFORMATION REGARDING THIS EVENT FROM THE USER FACILITY. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT. PRODUCT WAS NOT CHANGED OUT. PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12078825 ↗MalfunctionJun 28, 2021Physical Resistance/Sticking; Physical Resistance/Sticking
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT DURING CARDIOPULMONARY BYPASS, THE ARM LINKS GOT STUCK AND BOUND TOGETHER AND WERE DIFFICULT TO USE AND POSITION. IT IS UNKNOWN WHETHER THERE WAS A DELAY IN THE PROCEDURE. DUE TO THE UNKNOWN INFORMATION FOR THIS EVENT, IT IS BEING REPORTED. TERUMO CONTINUES TO ATTEMPT TO GAIN MORE INFORMATION REGARDING THIS EVENT FROM THE USER FACILITY. NO CONSEQUENCES OR IMPACT TO PATIENT. THE PRODUCT WAS NOT CHANGED OUT. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12078738 ↗MalfunctionJun 23, 2021Material Too Rigid or Stiff; Material Too Rigid or Stiff
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT DURING CARDIOPULMONARY BYPASS, THE ARM LINKS GOT STUCK. PER SUBSIDIARY,THE ARM WAS USED WITH 360 STABILIZER ATTACHMENT IN OFF-PUMP CABG. THE FIRST ANASTOMOSIS WAS COMPLETED SUCCESSFULLY AND WHEN THE HANDLE WAS LOOSENED TO REPOSITION THE ARM TO THE SECOND ANASTOMOSIS, THE ARM LINKS BECAME STUCK IN TWO PLACES AND WERE UNABLE TO BE BENT. THE ARM WAS ONCE REMOVED FROM THE RETRACTOR AND CLEANED WITH SALINE, BUT STILL THE STUCK PARTS WERE NOT LOOSENED AND THE STABILIZER WAS UNABLE TO BE REMOVED FROM THE DISTAL CONNECTION OF THE ARM. AFTER THAT, THE STABILIZER WAS SOMEHOW REMOVED FROM THE DISTAL CONNECTION AND THE ARM WAS ATTACHED TO THE RETRACTOR AGAIN TO BE USED THOUGH THE ARM LINKS WERE STILL STUCK. THE ARM WAS NOT REPOSITIONED WHEN THE ARM WAS TENSIONED. IT IS UNKNOWN WHETHER THERE WAS A DELAY IN THE PROCEDURE. DUE TO THE UNKNOWN INFORMATION FOR THIS EVENT, IT IS BEING REPORTED. TERUMO CONTINUES TO ATTEMPT TO GAIN MORE INFORMATION REGARDING THIS EVENT FROM THE USER FACILITY. NO KNOWN CONSEQUENCE OR IMPACT TO PATIENT. THE PRODUCT WAS NOT CHANGED OUT. THE SURGERY WAS COMPLETES SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12053384 ↗
FDA clearance and classification
The FDA regulates product code MWS (stabilizer, heart) under 21 CFR 870.4500 as a Class I device. 21 CFR 870.4500 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Herc Flex 360 Universal Stabilizer Arm T401162
Has the Herc Flex 360 Universal Stabilizer Arm T401162 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00699753450943), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Herc Flex 360 Universal Stabilizer Arm T401162?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Jun 23, 2021 and Feb 13, 2024. In those reports, the most common specific device problems are “Physical Resistance/Sticking” (4), “Break” (2) and “Defective Component” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Herc Flex 360 Universal Stabilizer Arm T401162?
None of the reports are classified as deaths, 0 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Herc Flex 360 Universal Stabilizer Arm T401162 FDA approved?
Devices of this type (product code MWS) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Terumo Cardiovascular Systems Corporation.
Who makes the Herc Flex 360 Universal Stabilizer Arm T401162, and is it still sold?
The FDA's device database lists Terumo Cardiovascular Systems Corporation as the company responsible for it (brand Herc Flex 360 Universal Stabilizer Arm, model T401162). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Stabilizer, Heart devices
- Acrobat-I Stabilizer OM-10000ZMaquet Cardiovascular LLC · Stabilizer,Heart2 recalls119 reports
- Acrobat-i Positioner XP-5000ZMaquet Cardiovascular LLC · Stabilizer,Heart2 recalls34 reports
- Acrobat-I Stabilizer OM-10000Maquet Cardiovascular LLC · Stabilizer,Heart1 recall1 death51 reports
- Acrobat SUV Stabilizer OM-9000ZMaquet Cardiovascular LLC · Stabilizer,Heart1 recall19 reports
- OCTOPUS® TS2000Medtronic, Inc. · stabilizer,heart1 recall15 reports
- Herc Flex Universal Stabilizer Arm T401163Terumo Cardiovascular Systems Corporation · Stabilizer, Heart1 recall10 reports
- Acrobat V Stabilizer OM-9100SMaquet Cardiovascular LLC · Stabilizer,Heart1 recall7 reports
- OCTOPUS® TSMICS1Medtronic, Inc. · Stabilizer, heart1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page